Position tracking of moving liver lesion based on real-time registration between 2D ultrasound and 3D preoperative images.
Saved in:
| Title: | Position tracking of moving liver lesion based on real-time registration between 2D ultrasound and 3D preoperative images. |
|---|---|
| Authors: | Weon, Chijun1, Hyun Nam, Woo1, Lee, Duhgoon1, Lee, Jae Young2, Ra, Jong Beom1 |
| Source: | Medical Physics. Jan2015, Vol. 42 Issue 1, p335-347. 13p. |
| Subjects: | Liver injuries, Position tracking (Virtual reality), Image registration, Ultrasonic imaging, Magnetic resonance imaging, Image-guided radiation therapy |
| Abstract: | Purpose: Registration between 2D ultrasound (US) and 3D preoperative magnetic resonance (MR) (or computed tomography, CT) images has been studied recently for US-guided intervention. However, the existing techniques have some limits, either in the registration speed or the performance. The purpose of this work is to develop a real-time and fully automatic registration system between two intermodal images of the liver, and subsequently an indirect lesion positioning/tracking algorithm based on the registration result, for image-guided interventions. Methods: The proposed position tracking system consists of three stages. In the preoperative stage, the authors acquire several 3D preoperative MR (or CT) images at different respiratory phases. Based on the transformations obtained from nonrigid registration of the acquired 3D images, they then generate a 4D preoperative image along the respiratory phase. In the intraoperative preparatory stage, they properly attach a 3D US transducer to the patient's body and fix its pose using a holding mechanism. They then acquire a couple of respiratory-controlled 3D US images. Via the rigid registration of these US images to the 3D preoperative images in the 4D image, the pose information of the fixed-pose 3D US transducer is determined with respect to the preoperative image coordinates. As feature(s) to use for the rigid registration, they may choose either internal liver vessels or the inferior vena cava. Since the latter is especially useful in patients with a diffuse liver disease, the authors newly propose using it. In the intraoperative real-time stage, they acquire 2D US images in real-time from the fixed-pose transducer. For each US image, they select candidates for its corresponding 2D preoperative slice from the 4D preoperative MR (or CT) image, based on the predetermined pose information of the transducer. The correct corresponding image is then found among those candidates via real-time 2D registration based on a gradient-based similarity measure. Finally, if needed, they obtain the position information of the liver lesion using the 3D preoperative image to which the registered 2D preoperative slice belongs. Results: The proposed method was applied to 23 clinical datasets and quantitative evaluations were conducted. With the exception of one clinical dataset that included US images of extremely low quality, 22 datasets of various liver status were successfully applied in the evaluation. Experimental results showed that the registration error between the anatomical features of US and preoperative MR images is less than 3 mm on average. The lesion tracking error was also found to be less than 5 mm at maximum. Conclusions: A new system has been proposed for real-time registration between 2D US and successive multiple 3D preoperative MR/CT images of the liver and was applied for indirect lesion tracking for image-guided intervention. The system is fully automatic and robust even with images that had low quality due to patient status. Through visual examinations and quantitative evaluations, it was verified that the proposed system can provide high lesion tracking accuracy as well as high registration accuracy, at performance levels which were acceptable for various clinical applications. [ABSTRACT FROM AUTHOR] |
| Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
|
Full text is not displayed to guests.
Login for full access.
|
|
| FullText | Links: – Type: pdflink Text: Availability: 1 |
|---|---|
| Header | DbId: egs DbLabel: Engineering Source An: 100334352 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Position tracking of moving liver lesion based on real-time registration between 2D ultrasound and 3D preoperative images. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Weon%2C+Chijun%22">Weon, Chijun</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Hyun+Nam%2C+Woo%22">Hyun Nam, Woo</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Lee%2C+Duhgoon%22">Lee, Duhgoon</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Lee%2C+Jae+Young%22">Lee, Jae Young</searchLink><relatesTo>2</relatesTo><br /><searchLink fieldCode="AR" term="%22Ra%2C+Jong+Beom%22">Ra, Jong Beom</searchLink><relatesTo>1</relatesTo> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. Jan2015, Vol. 42 Issue 1, p335-347. 13p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22Liver+injuries%22">Liver injuries</searchLink><br /><searchLink fieldCode="DE" term="%22Position+tracking+%28Virtual+reality%29%22">Position tracking (Virtual reality)</searchLink><br /><searchLink fieldCode="DE" term="%22Image+registration%22">Image registration</searchLink><br /><searchLink fieldCode="DE" term="%22Ultrasonic+imaging%22">Ultrasonic imaging</searchLink><br /><searchLink fieldCode="DE" term="%22Magnetic+resonance+imaging%22">Magnetic resonance imaging</searchLink><br /><searchLink fieldCode="DE" term="%22Image-guided+radiation+therapy%22">Image-guided radiation therapy</searchLink> – Name: Abstract Label: Abstract Group: Ab Data: Purpose: Registration between 2D ultrasound (US) and 3D preoperative magnetic resonance (MR) (or computed tomography, CT) images has been studied recently for US-guided intervention. However, the existing techniques have some limits, either in the registration speed or the performance. The purpose of this work is to develop a real-time and fully automatic registration system between two intermodal images of the liver, and subsequently an indirect lesion positioning/tracking algorithm based on the registration result, for image-guided interventions. Methods: The proposed position tracking system consists of three stages. In the preoperative stage, the authors acquire several 3D preoperative MR (or CT) images at different respiratory phases. Based on the transformations obtained from nonrigid registration of the acquired 3D images, they then generate a 4D preoperative image along the respiratory phase. In the intraoperative preparatory stage, they properly attach a 3D US transducer to the patient's body and fix its pose using a holding mechanism. They then acquire a couple of respiratory-controlled 3D US images. Via the rigid registration of these US images to the 3D preoperative images in the 4D image, the pose information of the fixed-pose 3D US transducer is determined with respect to the preoperative image coordinates. As feature(s) to use for the rigid registration, they may choose either internal liver vessels or the inferior vena cava. Since the latter is especially useful in patients with a diffuse liver disease, the authors newly propose using it. In the intraoperative real-time stage, they acquire 2D US images in real-time from the fixed-pose transducer. For each US image, they select candidates for its corresponding 2D preoperative slice from the 4D preoperative MR (or CT) image, based on the predetermined pose information of the transducer. The correct corresponding image is then found among those candidates via real-time 2D registration based on a gradient-based similarity measure. Finally, if needed, they obtain the position information of the liver lesion using the 3D preoperative image to which the registered 2D preoperative slice belongs. Results: The proposed method was applied to 23 clinical datasets and quantitative evaluations were conducted. With the exception of one clinical dataset that included US images of extremely low quality, 22 datasets of various liver status were successfully applied in the evaluation. Experimental results showed that the registration error between the anatomical features of US and preoperative MR images is less than 3 mm on average. The lesion tracking error was also found to be less than 5 mm at maximum. Conclusions: A new system has been proposed for real-time registration between 2D US and successive multiple 3D preoperative MR/CT images of the liver and was applied for indirect lesion tracking for image-guided intervention. The system is fully automatic and robust even with images that had low quality due to patient status. Through visual examinations and quantitative evaluations, it was verified that the proposed system can provide high lesion tracking accuracy as well as high registration accuracy, at performance levels which were acceptable for various clinical applications. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=100334352 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.1118/1.4903945 Languages: – Code: eng Text: English PhysicalDescription: Pagination: PageCount: 13 StartPage: 335 Subjects: – SubjectFull: Liver injuries Type: general – SubjectFull: Position tracking (Virtual reality) Type: general – SubjectFull: Image registration Type: general – SubjectFull: Ultrasonic imaging Type: general – SubjectFull: Magnetic resonance imaging Type: general – SubjectFull: Image-guided radiation therapy Type: general Titles: – TitleFull: Position tracking of moving liver lesion based on real-time registration between 2D ultrasound and 3D preoperative images. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Weon, Chijun – PersonEntity: Name: NameFull: Hyun Nam, Woo – PersonEntity: Name: NameFull: Lee, Duhgoon – PersonEntity: Name: NameFull: Lee, Jae Young – PersonEntity: Name: NameFull: Ra, Jong Beom IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 01 Text: Jan2015 Type: published Y: 2015 Identifiers: – Type: issn-print Value: 00942405 Numbering: – Type: volume Value: 42 – Type: issue Value: 1 Titles: – TitleFull: Medical Physics Type: main |
| ResultId | 1 |