Validation of a pretreatment delivery quality assurance method for the CyberKnife Synchrony system.

Saved in:
Bibliographic Details
Title: Validation of a pretreatment delivery quality assurance method for the CyberKnife Synchrony system.
Authors: Mastella, E.1, Vigorito, S.2, Rondi, E.2, Piperno, G.3, Ferrari, A.3, Strata, E.3, Rozza, D.3, Jereczek‐Fossa, B. A.4, Cattani, F.2
Source: Medical Physics. Aug2016, Vol. 43 Issue 8, p4565-4574. 10p.
Subjects: Radiation doses, Medical quality control, Drug delivery systems, Medical dosimetry, Respiratory diseases, Treatment of respiratory diseases, Patients
Abstract: Purpose: To evaluate the geometric and dosimetric accuracies of the CyberKnife Synchrony respiratory tracking system (RTS) and to validate a method for pretreatment patient-specific delivery quality assurance (DQA). Methods: An EasyCube phantom was mounted on the ExacTrac gating phantom, which can move along the superior-inferior (SI) axis of a patient to simulate a moving target. The authors compared dynamic and static measurements. For each case, a Gafchromic EBT3 film was positioned between two slabs of the EasyCube, while a PinPoint ionization chamber was placed in the appropriate space. There were three steps to their evaluation: (1) the field size, the penumbra, and the symmetry of six secondary collimators were measured along the two main orthogonal axes. Dynamic measurements with deliberately simulated errors were also taken. (2) The delivered dose distributions (from step 1) were compared with the planned ones, using the gamma analysis method. The local gamma passing rates were evaluated using three acceptance criteria: 3% local dose difference (LDD)/3 mm, 2%LDD/2 mm, and 3%LDD/1 mm. (3) The DQA plans for six clinical patients were irradiated in different dynamic conditions, to give a total of 19 cases. The measured and planned dose distributions were evaluated with the same gamma-index criteria used in step 2 and the measured chamber doses were compared with the planned mean doses in the sensitive volume of the chamber. Results: (1) A very slight enlargement of the field size and of the penumbra was observed in the SI direction (on average <1 mm), in line with the overall average CyberKnife system error for tracking treatments. (2) Comparison between the planned and the correctly delivered dose distributions confirmed the dosimetric accuracy of the RTS for simple plans. The multicriteria gamma analysis was able to detect the simulated errors, proving the robustness of their method of analysis. (3) All of the DQA clinical plans passed the tests, both in static and dynamic conditions. No statistically significant differences were found between static and dynamic cases, confirming the high degree of accuracy of the Synchrony RTS. Conclusions: The presented methods and measurements verified the mechanical and dosimetric accuracy of the Synchrony RTS. Their method confirms the fact that the RTS, if used properly, is able to treat a moving target with great precision. By combining PinPoint ion chamber, EBT3 films, and gamma evaluation of dose distributions, their DQA method robustly validated the effectiveness of CyberKnife and Synchrony system. [ABSTRACT FROM AUTHOR]
Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Links:
  – Type: pdflink
Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 117203463
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Validation of a pretreatment delivery quality assurance method for the CyberKnife Synchrony system.
– Name: Author
  Label: Authors
  Group: Au
  Data: &lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Mastella%2C+E%2E%22&quot;&gt;Mastella, E.&lt;/searchLink&gt;&lt;relatesTo&gt;1&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Vigorito%2C+S%2E%22&quot;&gt;Vigorito, S.&lt;/searchLink&gt;&lt;relatesTo&gt;2&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Rondi%2C+E%2E%22&quot;&gt;Rondi, E.&lt;/searchLink&gt;&lt;relatesTo&gt;2&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Piperno%2C+G%2E%22&quot;&gt;Piperno, G.&lt;/searchLink&gt;&lt;relatesTo&gt;3&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Ferrari%2C+A%2E%22&quot;&gt;Ferrari, A.&lt;/searchLink&gt;&lt;relatesTo&gt;3&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Strata%2C+E%2E%22&quot;&gt;Strata, E.&lt;/searchLink&gt;&lt;relatesTo&gt;3&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Rozza%2C+D%2E%22&quot;&gt;Rozza, D.&lt;/searchLink&gt;&lt;relatesTo&gt;3&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Jereczek‐Fossa%2C+B%2E+A%2E%22&quot;&gt;Jereczek‐Fossa, B. A.&lt;/searchLink&gt;&lt;relatesTo&gt;4&lt;/relatesTo&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Cattani%2C+F%2E%22&quot;&gt;Cattani, F.&lt;/searchLink&gt;&lt;relatesTo&gt;2&lt;/relatesTo&gt;
– Name: TitleSource
  Label: Source
  Group: Src
  Data: &lt;searchLink fieldCode=&quot;JN&quot; term=&quot;%22Medical+Physics%22&quot;&gt;Medical Physics&lt;/searchLink&gt;. Aug2016, Vol. 43 Issue 8, p4565-4574. 10p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: &lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Radiation+doses%22&quot;&gt;Radiation doses&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Medical+quality+control%22&quot;&gt;Medical quality control&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Drug+delivery+systems%22&quot;&gt;Drug delivery systems&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Medical+dosimetry%22&quot;&gt;Medical dosimetry&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Respiratory+diseases%22&quot;&gt;Respiratory diseases&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Treatment+of+respiratory+diseases%22&quot;&gt;Treatment of respiratory diseases&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Patients%22&quot;&gt;Patients&lt;/searchLink&gt;
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: To evaluate the geometric and dosimetric accuracies of the CyberKnife Synchrony respiratory tracking system (RTS) and to validate a method for pretreatment patient-specific delivery quality assurance (DQA). Methods: An EasyCube phantom was mounted on the ExacTrac gating phantom, which can move along the superior-inferior (SI) axis of a patient to simulate a moving target. The authors compared dynamic and static measurements. For each case, a Gafchromic EBT3 film was positioned between two slabs of the EasyCube, while a PinPoint ionization chamber was placed in the appropriate space. There were three steps to their evaluation: (1) the field size, the penumbra, and the symmetry of six secondary collimators were measured along the two main orthogonal axes. Dynamic measurements with deliberately simulated errors were also taken. (2) The delivered dose distributions (from step 1) were compared with the planned ones, using the gamma analysis method. The local gamma passing rates were evaluated using three acceptance criteria: 3% local dose difference (LDD)/3 mm, 2%LDD/2 mm, and 3%LDD/1 mm. (3) The DQA plans for six clinical patients were irradiated in different dynamic conditions, to give a total of 19 cases. The measured and planned dose distributions were evaluated with the same gamma-index criteria used in step 2 and the measured chamber doses were compared with the planned mean doses in the sensitive volume of the chamber. Results: (1) A very slight enlargement of the field size and of the penumbra was observed in the SI direction (on average &lt;1 mm), in line with the overall average CyberKnife system error for tracking treatments. (2) Comparison between the planned and the correctly delivered dose distributions confirmed the dosimetric accuracy of the RTS for simple plans. The multicriteria gamma analysis was able to detect the simulated errors, proving the robustness of their method of analysis. (3) All of the DQA clinical plans passed the tests, both in static and dynamic conditions. No statistically significant differences were found between static and dynamic cases, confirming the high degree of accuracy of the Synchrony RTS. Conclusions: The presented methods and measurements verified the mechanical and dosimetric accuracy of the Synchrony RTS. Their method confirms the fact that the RTS, if used properly, is able to treat a moving target with great precision. By combining PinPoint ion chamber, EBT3 films, and gamma evaluation of dose distributions, their DQA method robustly validated the effectiveness of CyberKnife and Synchrony system. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: &lt;i&gt;Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder&#39;s express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.&lt;/i&gt; (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=117203463
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1118/1.4955437
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 10
        StartPage: 4565
    Subjects:
      – SubjectFull: Radiation doses
        Type: general
      – SubjectFull: Medical quality control
        Type: general
      – SubjectFull: Drug delivery systems
        Type: general
      – SubjectFull: Medical dosimetry
        Type: general
      – SubjectFull: Respiratory diseases
        Type: general
      – SubjectFull: Treatment of respiratory diseases
        Type: general
      – SubjectFull: Patients
        Type: general
    Titles:
      – TitleFull: Validation of a pretreatment delivery quality assurance method for the CyberKnife Synchrony system.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Mastella, E.
      – PersonEntity:
          Name:
            NameFull: Vigorito, S.
      – PersonEntity:
          Name:
            NameFull: Rondi, E.
      – PersonEntity:
          Name:
            NameFull: Piperno, G.
      – PersonEntity:
          Name:
            NameFull: Ferrari, A.
      – PersonEntity:
          Name:
            NameFull: Strata, E.
      – PersonEntity:
          Name:
            NameFull: Rozza, D.
      – PersonEntity:
          Name:
            NameFull: Jereczek‐Fossa, B. A.
      – PersonEntity:
          Name:
            NameFull: Cattani, F.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 08
              Text: Aug2016
              Type: published
              Y: 2016
          Identifiers:
            – Type: issn-print
              Value: 00942405
          Numbering:
            – Type: volume
              Value: 43
            – Type: issue
              Value: 8
          Titles:
            – TitleFull: Medical Physics
              Type: main
ResultId 1