In-Situ Systematic Error Correction for Digital Volume Correlation Using a Reference Sample.
Saved in:
| Title: | |
|---|---|
| Authors: | Wang, B.1, Pan, B.1 panb@buaa.edu.cn, Lubineau, G.2 gilles.lubineau@kaust.edu.sa |
| Source: | Experimental Mechanics. Mar2018, Vol. 58 Issue 3, p427-436. 10p. |
| Subjects: | Computed tomography, Samples (Commerce), Compression fractures, Digital image processing, Polynomials |
| Abstract: | The self-heating effect of a laboratory X-ray computed tomography (CT) scanner causes slight change in its imaging geometry, which induces translation and dilatation (i.e., artificial displacement and strain) in reconstructed volume images recorded at different times. To realize high-accuracy internal full-field deformation measurements using digital volume correlation (DVC), these artificial displacements and strains associated with unstable CT imaging must be eliminated. In this work, an effective and easily implemented reference sample compensation (RSC) method is proposed for |
| Copyright of Experimental Mechanics is the property of Springer Nature and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
|
Full text is not displayed to guests.
Login for full access.
|
|
Be the first to leave a comment!