Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter – Part II: System modeling, scatter correction, and optimization.

Saved in:
Bibliographic Details
Title: Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter – Part II: System modeling, scatter correction, and optimization.
Authors: Wang, Adam1 adam.wang@varian.com, Maslowski, Alexander1, Messmer, Philippe1, Lehmann, Mathias1, Strzelecki, Adam1, Yu, Elaine1, Paysan, Pascal1, Brehm, Marcus1, Munro, Peter1, Star‐Lack, Josh1, Seghers, Dieter1
Source: Medical Physics. May2018, Vol. 45 Issue 5, p1914-1925. 12p.
Subjects: Boltzmann's equation, Cone beam computed tomography, Mathematical optimization, Imaging phantoms, Image reconstruction
Abstract: Purpose: To correct for scatter in kV cone‐beam CT (CBCT) projection data on a clinical system using a new tool, Acuros® CTS, that estimates scatter images rapidly and accurately by deterministically solving the linear Boltzmann transport equation. Methods: Phantom and patient CBCT scans were acquired on TrueBeam® radiotherapy machines. A first‐pass reconstruction was used to create water and bone density maps of the imaged object, which was updated to include a more accurate representation of the patient couch. The imaging system model accounted for the TrueBeam x‐ray source (polychromatic spectrum, beam filtration, bowtie filter, and collimation hardware) and x‐ray detection system (antiscatter grid, flat‐panel imager). Acuros CTS then used the system and object models to estimate the scatter component of each projection image, which was subtracted from the measured projections. The corrected projections were then reconstructed to produce the final result. We examined the tradeoff between run time and accuracy using a Pareto optimization of key parameters, including the voxel size of the down‐sampled object model, the number of pixels in the down‐sampled detector, and the number of scatter images (angular down‐sampling). All computations and reconstructions were performed on a research workstation containing two graphics processing units (GPUs). In addition, we established a method for selecting a subset of projections for which scatter images were calculated. The projections were selected to minimize interpolation errors in the remaining projections. Image quality improvement was assessed by measuring the accuracy of the reconstructed phantom and patient images. Results: The Pareto optimization yielded a set of parameters with an average run time of 26 seconds for scatter correction while maintaining high accuracy of scatter estimation. This was achieved in part by means of optimizing the projection angles that were processed, thus favoring the use of more angles in the lateral (i.e., horizontal) direction and fewer angles in the AP direction. In a 40 cm solid water phantom reconstruction, nonuniformities were decreased from 217 HU without scatter correction to 51 HU with conventional (kernel‐based) scatter correction to 17 HU with Acuros CTS‐based scatter correction. In clinical pelvis scans, nonuniformities in the bladder were reduced from 85 HU with conventional scatter correction to 14 HU with Acuros CTS. Conclusions: Acuros CTS is a promising new tool for fast and accurate scatter correction for CBCT imaging. By carefully modeling the imaging chain and optimizing several parameters, we achieved high correction accuracies with computation times compatible with the clinical workflow. The improvement in image quality enables better soft‐tissue visualization and potentially enables applications such as adaptive radiotherapy. [ABSTRACT FROM AUTHOR]
Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 129573863
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter – Part II: System modeling, scatter correction, and optimization.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Wang%2C+Adam%22">Wang, Adam</searchLink><relatesTo>1</relatesTo><i> adam.wang@varian.com</i><br /><searchLink fieldCode="AR" term="%22Maslowski%2C+Alexander%22">Maslowski, Alexander</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Messmer%2C+Philippe%22">Messmer, Philippe</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Lehmann%2C+Mathias%22">Lehmann, Mathias</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Strzelecki%2C+Adam%22">Strzelecki, Adam</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Yu%2C+Elaine%22">Yu, Elaine</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Paysan%2C+Pascal%22">Paysan, Pascal</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Brehm%2C+Marcus%22">Brehm, Marcus</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Munro%2C+Peter%22">Munro, Peter</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Star‐Lack%2C+Josh%22">Star‐Lack, Josh</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Seghers%2C+Dieter%22">Seghers, Dieter</searchLink><relatesTo>1</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. May2018, Vol. 45 Issue 5, p1914-1925. 12p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Boltzmann's+equation%22">Boltzmann's equation</searchLink><br /><searchLink fieldCode="DE" term="%22Cone+beam+computed+tomography%22">Cone beam computed tomography</searchLink><br /><searchLink fieldCode="DE" term="%22Mathematical+optimization%22">Mathematical optimization</searchLink><br /><searchLink fieldCode="DE" term="%22Imaging+phantoms%22">Imaging phantoms</searchLink><br /><searchLink fieldCode="DE" term="%22Image+reconstruction%22">Image reconstruction</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: To correct for scatter in kV cone‐beam CT (CBCT) projection data on a clinical system using a new tool, Acuros® CTS, that estimates scatter images rapidly and accurately by deterministically solving the linear Boltzmann transport equation. Methods: Phantom and patient CBCT scans were acquired on TrueBeam® radiotherapy machines. A first‐pass reconstruction was used to create water and bone density maps of the imaged object, which was updated to include a more accurate representation of the patient couch. The imaging system model accounted for the TrueBeam x‐ray source (polychromatic spectrum, beam filtration, bowtie filter, and collimation hardware) and x‐ray detection system (antiscatter grid, flat‐panel imager). Acuros CTS then used the system and object models to estimate the scatter component of each projection image, which was subtracted from the measured projections. The corrected projections were then reconstructed to produce the final result. We examined the tradeoff between run time and accuracy using a Pareto optimization of key parameters, including the voxel size of the down‐sampled object model, the number of pixels in the down‐sampled detector, and the number of scatter images (angular down‐sampling). All computations and reconstructions were performed on a research workstation containing two graphics processing units (GPUs). In addition, we established a method for selecting a subset of projections for which scatter images were calculated. The projections were selected to minimize interpolation errors in the remaining projections. Image quality improvement was assessed by measuring the accuracy of the reconstructed phantom and patient images. Results: The Pareto optimization yielded a set of parameters with an average run time of 26 seconds for scatter correction while maintaining high accuracy of scatter estimation. This was achieved in part by means of optimizing the projection angles that were processed, thus favoring the use of more angles in the lateral (i.e., horizontal) direction and fewer angles in the AP direction. In a 40 cm solid water phantom reconstruction, nonuniformities were decreased from 217 HU without scatter correction to 51 HU with conventional (kernel‐based) scatter correction to 17 HU with Acuros CTS‐based scatter correction. In clinical pelvis scans, nonuniformities in the bladder were reduced from 85 HU with conventional scatter correction to 14 HU with Acuros CTS. Conclusions: Acuros CTS is a promising new tool for fast and accurate scatter correction for CBCT imaging. By carefully modeling the imaging chain and optimizing several parameters, we achieved high correction accuracies with computation times compatible with the clinical workflow. The improvement in image quality enables better soft‐tissue visualization and potentially enables applications such as adaptive radiotherapy. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=129573863
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1002/mp.12849
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 12
        StartPage: 1914
    Subjects:
      – SubjectFull: Boltzmann's equation
        Type: general
      – SubjectFull: Cone beam computed tomography
        Type: general
      – SubjectFull: Mathematical optimization
        Type: general
      – SubjectFull: Imaging phantoms
        Type: general
      – SubjectFull: Image reconstruction
        Type: general
    Titles:
      – TitleFull: Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter – Part II: System modeling, scatter correction, and optimization.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Wang, Adam
      – PersonEntity:
          Name:
            NameFull: Maslowski, Alexander
      – PersonEntity:
          Name:
            NameFull: Messmer, Philippe
      – PersonEntity:
          Name:
            NameFull: Lehmann, Mathias
      – PersonEntity:
          Name:
            NameFull: Strzelecki, Adam
      – PersonEntity:
          Name:
            NameFull: Yu, Elaine
      – PersonEntity:
          Name:
            NameFull: Paysan, Pascal
      – PersonEntity:
          Name:
            NameFull: Brehm, Marcus
      – PersonEntity:
          Name:
            NameFull: Munro, Peter
      – PersonEntity:
          Name:
            NameFull: Star‐Lack, Josh
      – PersonEntity:
          Name:
            NameFull: Seghers, Dieter
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 05
              Text: May2018
              Type: published
              Y: 2018
          Identifiers:
            – Type: issn-print
              Value: 00942405
          Numbering:
            – Type: volume
              Value: 45
            – Type: issue
              Value: 5
          Titles:
            – TitleFull: Medical Physics
              Type: main
ResultId 1