A ring‐based compensator IMRT system optimized for low‐ and middle‐income countries: Design and treatment planning study.

Saved in:
Bibliographic Details
Title: A ring‐based compensator IMRT system optimized for low‐ and middle‐income countries: Design and treatment planning study.
Authors: Van Schelt, Jonathon1,2 jonathon_vanschelt@rush.edu, Smith, Daniel L.1, Fong, Nicholas1, Toomeh, Dolla1, Sponseller, Patricia A.1, Brown, Derek W.3, Macomber, Meghan W.1, Mayr, Nina A.1, Patel, Shilpen4, Shulman, Adam5, Subrahmanyam, G. V.6, Govindarajan, K. N.7, Ford, Eric C.1 eford@uw.edu
Source: Medical Physics. Jul2018, Vol. 45 Issue 7, p3275-3286. 12p.
Subjects: Synchronous capacitors, Medical protocols, Intensity modulated radiotherapy, Retrofitting, Monte Carlo method, Numerical analysis
Abstract: Purpose: We propose a novel compensator‐based IMRT system designed to provide a simple, reliable, and cost‐effective adjunct technology, with the goal of expanding global access to advanced radiotherapy techniques. The system would employ easily reusable tungsten bead compensators that operate independent of a gantry (e.g., mounted in a ring around the patient). Thereby the system can be retrofitted to existing linac and cobalt teletherapy units. This study explores the quality of treatment plans from the proposed system and the dependence on associated design parameters. Methods: We considered 60Co‐based plans as the most challenging scenario for dosimetry and benchmarked them against clinical MLC‐based plans delivered on a linac. Treatment planning was performed in the Pinnacle treatment planning system with commissioning based on Monte Carlo simulations of compensated beams. 60Co‐compensator IMRT plans were generated for five patients with head‐and‐neck cancer and five with gynecological cancer and compared to respective IMRT plans using a 6 MV linac beam with an MLC. The dependence of dosimetric endpoints on compensator resolution, thickness, position, and number of beams was assessed. Dosimetric accuracy was validated by Monte Carlo simulations of dose distribution in a water phantom from beams with the IMRT plan compensators. Results: The 60Co‐compensator plans had on average equivalent PTV coverage and somewhat inferior OAR sparing compared to the 6 MV‐MLC plans, but the differences in dosimetric endpoints were clinically acceptable. Calculated treatment times for head‐and‐neck plans were 7.6 ± 2.0 min vs 3.9 ± 0.8 min (6 MV‐MLC vs 60Co‐compensator) and for gynecological plans were 8.7 ± 3.1 min vs 4.3 ± 0.4 min. Plan quality was insensitive to most design parameters over much of the ranges studied, with no degradation found when the compensator resolution was finer than 6 mm, maximum thickness at least 2 tenth‐value‐layers, and more than five beams were used. Source‐to‐compensator distances of 53 and 63 cm resulted in very similar plan quality. Monte Carlo simulations suggest no increase in surface dose for the geometries considered here. Simulated dosimetric validation tests had median gamma pass rates of 97.6% for criteria of 3% (global)/3 mm with a 10% threshold. Conclusions: The novel ring‐compensator IMRT system can produce plans of comparable quality to standard 6 MV‐MLC systems. Even when 60Co beams are used the plan quality is acceptable and treatment times are substantially reduced. 60Co‐compensator IMRT plans are adequately modeled in an existing commercial treatment planning system. These results motivate further development of this low‐cost adaptable technology with translation through clinical trials and deployment to expand the reach of IMRT in low‐ and middle‐income countries. [ABSTRACT FROM AUTHOR]
Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Links:
  – Type: pdflink
Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 130628571
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: A ring‐based compensator IMRT system optimized for low‐ and middle‐income countries: Design and treatment planning study.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Van+Schelt%2C+Jonathon%22">Van Schelt, Jonathon</searchLink><relatesTo>1,2</relatesTo><i> jonathon_vanschelt@rush.edu</i><br /><searchLink fieldCode="AR" term="%22Smith%2C+Daniel+L%2E%22">Smith, Daniel L.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Fong%2C+Nicholas%22">Fong, Nicholas</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Toomeh%2C+Dolla%22">Toomeh, Dolla</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Sponseller%2C+Patricia+A%2E%22">Sponseller, Patricia A.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Brown%2C+Derek+W%2E%22">Brown, Derek W.</searchLink><relatesTo>3</relatesTo><br /><searchLink fieldCode="AR" term="%22Macomber%2C+Meghan+W%2E%22">Macomber, Meghan W.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Mayr%2C+Nina+A%2E%22">Mayr, Nina A.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Patel%2C+Shilpen%22">Patel, Shilpen</searchLink><relatesTo>4</relatesTo><br /><searchLink fieldCode="AR" term="%22Shulman%2C+Adam%22">Shulman, Adam</searchLink><relatesTo>5</relatesTo><br /><searchLink fieldCode="AR" term="%22Subrahmanyam%2C+G%2E+V%2E%22">Subrahmanyam, G. V.</searchLink><relatesTo>6</relatesTo><br /><searchLink fieldCode="AR" term="%22Govindarajan%2C+K%2E+N%2E%22">Govindarajan, K. N.</searchLink><relatesTo>7</relatesTo><br /><searchLink fieldCode="AR" term="%22Ford%2C+Eric+C%2E%22">Ford, Eric C.</searchLink><relatesTo>1</relatesTo><i> eford@uw.edu</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. Jul2018, Vol. 45 Issue 7, p3275-3286. 12p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Synchronous+capacitors%22">Synchronous capacitors</searchLink><br /><searchLink fieldCode="DE" term="%22Medical+protocols%22">Medical protocols</searchLink><br /><searchLink fieldCode="DE" term="%22Intensity+modulated+radiotherapy%22">Intensity modulated radiotherapy</searchLink><br /><searchLink fieldCode="DE" term="%22Retrofitting%22">Retrofitting</searchLink><br /><searchLink fieldCode="DE" term="%22Monte+Carlo+method%22">Monte Carlo method</searchLink><br /><searchLink fieldCode="DE" term="%22Numerical+analysis%22">Numerical analysis</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: We propose a novel compensator‐based IMRT system designed to provide a simple, reliable, and cost‐effective adjunct technology, with the goal of expanding global access to advanced radiotherapy techniques. The system would employ easily reusable tungsten bead compensators that operate independent of a gantry (e.g., mounted in a ring around the patient). Thereby the system can be retrofitted to existing linac and cobalt teletherapy units. This study explores the quality of treatment plans from the proposed system and the dependence on associated design parameters. Methods: We considered 60Co‐based plans as the most challenging scenario for dosimetry and benchmarked them against clinical MLC‐based plans delivered on a linac. Treatment planning was performed in the Pinnacle treatment planning system with commissioning based on Monte Carlo simulations of compensated beams. 60Co‐compensator IMRT plans were generated for five patients with head‐and‐neck cancer and five with gynecological cancer and compared to respective IMRT plans using a 6 MV linac beam with an MLC. The dependence of dosimetric endpoints on compensator resolution, thickness, position, and number of beams was assessed. Dosimetric accuracy was validated by Monte Carlo simulations of dose distribution in a water phantom from beams with the IMRT plan compensators. Results: The 60Co‐compensator plans had on average equivalent PTV coverage and somewhat inferior OAR sparing compared to the 6 MV‐MLC plans, but the differences in dosimetric endpoints were clinically acceptable. Calculated treatment times for head‐and‐neck plans were 7.6 ± 2.0 min vs 3.9 ± 0.8 min (6 MV‐MLC vs 60Co‐compensator) and for gynecological plans were 8.7 ± 3.1 min vs 4.3 ± 0.4 min. Plan quality was insensitive to most design parameters over much of the ranges studied, with no degradation found when the compensator resolution was finer than 6 mm, maximum thickness at least 2 tenth‐value‐layers, and more than five beams were used. Source‐to‐compensator distances of 53 and 63 cm resulted in very similar plan quality. Monte Carlo simulations suggest no increase in surface dose for the geometries considered here. Simulated dosimetric validation tests had median gamma pass rates of 97.6% for criteria of 3% (global)/3 mm with a 10% threshold. Conclusions: The novel ring‐compensator IMRT system can produce plans of comparable quality to standard 6 MV‐MLC systems. Even when 60Co beams are used the plan quality is acceptable and treatment times are substantially reduced. 60Co‐compensator IMRT plans are adequately modeled in an existing commercial treatment planning system. These results motivate further development of this low‐cost adaptable technology with translation through clinical trials and deployment to expand the reach of IMRT in low‐ and middle‐income countries. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=130628571
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1002/mp.12985
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 12
        StartPage: 3275
    Subjects:
      – SubjectFull: Synchronous capacitors
        Type: general
      – SubjectFull: Medical protocols
        Type: general
      – SubjectFull: Intensity modulated radiotherapy
        Type: general
      – SubjectFull: Retrofitting
        Type: general
      – SubjectFull: Monte Carlo method
        Type: general
      – SubjectFull: Numerical analysis
        Type: general
    Titles:
      – TitleFull: A ring‐based compensator IMRT system optimized for low‐ and middle‐income countries: Design and treatment planning study.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Van Schelt, Jonathon
      – PersonEntity:
          Name:
            NameFull: Smith, Daniel L.
      – PersonEntity:
          Name:
            NameFull: Fong, Nicholas
      – PersonEntity:
          Name:
            NameFull: Toomeh, Dolla
      – PersonEntity:
          Name:
            NameFull: Sponseller, Patricia A.
      – PersonEntity:
          Name:
            NameFull: Brown, Derek W.
      – PersonEntity:
          Name:
            NameFull: Macomber, Meghan W.
      – PersonEntity:
          Name:
            NameFull: Mayr, Nina A.
      – PersonEntity:
          Name:
            NameFull: Patel, Shilpen
      – PersonEntity:
          Name:
            NameFull: Shulman, Adam
      – PersonEntity:
          Name:
            NameFull: Subrahmanyam, G. V.
      – PersonEntity:
          Name:
            NameFull: Govindarajan, K. N.
      – PersonEntity:
          Name:
            NameFull: Ford, Eric C.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 07
              Text: Jul2018
              Type: published
              Y: 2018
          Identifiers:
            – Type: issn-print
              Value: 00942405
          Numbering:
            – Type: volume
              Value: 45
            – Type: issue
              Value: 7
          Titles:
            – TitleFull: Medical Physics
              Type: main
ResultId 1