Three‐dimensional printing of patient‐specific lung phantoms for CT imaging: Emulating lung tissue with accurate attenuation profiles and textures.

Saved in:
Bibliographic Details
Title: Three‐dimensional printing of patient‐specific lung phantoms for CT imaging: Emulating lung tissue with accurate attenuation profiles and textures.
Authors: Mei, Kai1 (AUTHOR) kai.mei@pennmedicine.upenn.edu, Geagan, Michael1 (AUTHOR), Roshkovan, Leonid1 (AUTHOR), Litt, Harold I.1 (AUTHOR), Gang, Grace J.2 (AUTHOR), Shapira, Nadav1 (AUTHOR), Stayman, J. Webster2 (AUTHOR), Noël, Peter B.1,3 (AUTHOR) pbnoel@upenn.edu
Source: Medical Physics. Feb2022, Vol. 49 Issue 2, p825-835. 11p.
Subjects: Imaging phantoms, Computed tomography, Three-dimensional printing, Lungs, Radiomics, Texture analysis (Image processing), Digital communications
Abstract: Purpose: Phantoms are a basic tool for assessing and verifying performance in CT research and clinical practice. Patient‐based realistic lung phantoms accurately representing textures and densities are essential in developing and evaluating novel CT hardware and software. This study introduces PixelPrint, a 3D printing solution to create patient‐based lung phantoms with accurate attenuation profiles and textures. Methods: PixelPrint, a software tool, was developed to convert patient digital imaging and communications in medicine (DICOM) images directly into FDM printer instructions (G‐code). Density was modeled as the ratio of filament to voxel volume to emulate attenuation profiles for each voxel, with the filament ratio controlled through continuous modification of the printing speed. A calibration phantom was designed to determine the mapping between filament line width and Hounsfield units (HU) within the range of human lungs. For evaluation of PixelPrint, a phantom based on a single human lung slice was manufactured and scanned with the same CT scanner and protocol used for the patient scan. Density and geometrical accuracy between phantom and patient CT data were evaluated for various anatomical features in the lung. Results: For the calibration phantom, measured mean HU show a very high level of linear correlation with respect to the utilized filament line widths, (r > 0.999). Qualitatively, the CT image of the patient‐based phantom closely resembles the original CT image both in texture and contrast levels (from −800 to 0 HU), with clearly visible vascular and parenchymal structures. Regions of interest comparing attenuation illustrated differences below 15 HU. Manual size measurements performed by an experienced thoracic radiologist reveal a high degree of geometrical correlation of details between identical patient and phantom features, with differences smaller than the intrinsic spatial resolution of the scans. Conclusion: The present study demonstrates the feasibility of 3D‐printed patient‐based lung phantoms with accurate organ geometry, image texture, and attenuation profiles. PixelPrint will enable applications in the research and development of CT technology, including further development in radiomics. [ABSTRACT FROM AUTHOR]
Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 155147371
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Three‐dimensional printing of patient‐specific lung phantoms for CT imaging: Emulating lung tissue with accurate attenuation profiles and textures.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Mei%2C+Kai%22">Mei, Kai</searchLink><relatesTo>1</relatesTo> (AUTHOR)<i> kai.mei@pennmedicine.upenn.edu</i><br /><searchLink fieldCode="AR" term="%22Geagan%2C+Michael%22">Geagan, Michael</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Roshkovan%2C+Leonid%22">Roshkovan, Leonid</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Litt%2C+Harold+I%2E%22">Litt, Harold I.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Gang%2C+Grace+J%2E%22">Gang, Grace J.</searchLink><relatesTo>2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Shapira%2C+Nadav%22">Shapira, Nadav</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Stayman%2C+J%2E+Webster%22">Stayman, J. Webster</searchLink><relatesTo>2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Noël%2C+Peter+B%2E%22">Noël, Peter B.</searchLink><relatesTo>1,3</relatesTo> (AUTHOR)<i> pbnoel@upenn.edu</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. Feb2022, Vol. 49 Issue 2, p825-835. 11p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Imaging+phantoms%22">Imaging phantoms</searchLink><br /><searchLink fieldCode="DE" term="%22Computed+tomography%22">Computed tomography</searchLink><br /><searchLink fieldCode="DE" term="%22Three-dimensional+printing%22">Three-dimensional printing</searchLink><br /><searchLink fieldCode="DE" term="%22Lungs%22">Lungs</searchLink><br /><searchLink fieldCode="DE" term="%22Radiomics%22">Radiomics</searchLink><br /><searchLink fieldCode="DE" term="%22Texture+analysis+%28Image+processing%29%22">Texture analysis (Image processing)</searchLink><br /><searchLink fieldCode="DE" term="%22Digital+communications%22">Digital communications</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: Phantoms are a basic tool for assessing and verifying performance in CT research and clinical practice. Patient‐based realistic lung phantoms accurately representing textures and densities are essential in developing and evaluating novel CT hardware and software. This study introduces PixelPrint, a 3D printing solution to create patient‐based lung phantoms with accurate attenuation profiles and textures. Methods: PixelPrint, a software tool, was developed to convert patient digital imaging and communications in medicine (DICOM) images directly into FDM printer instructions (G‐code). Density was modeled as the ratio of filament to voxel volume to emulate attenuation profiles for each voxel, with the filament ratio controlled through continuous modification of the printing speed. A calibration phantom was designed to determine the mapping between filament line width and Hounsfield units (HU) within the range of human lungs. For evaluation of PixelPrint, a phantom based on a single human lung slice was manufactured and scanned with the same CT scanner and protocol used for the patient scan. Density and geometrical accuracy between phantom and patient CT data were evaluated for various anatomical features in the lung. Results: For the calibration phantom, measured mean HU show a very high level of linear correlation with respect to the utilized filament line widths, (r > 0.999). Qualitatively, the CT image of the patient‐based phantom closely resembles the original CT image both in texture and contrast levels (from −800 to 0 HU), with clearly visible vascular and parenchymal structures. Regions of interest comparing attenuation illustrated differences below 15 HU. Manual size measurements performed by an experienced thoracic radiologist reveal a high degree of geometrical correlation of details between identical patient and phantom features, with differences smaller than the intrinsic spatial resolution of the scans. Conclusion: The present study demonstrates the feasibility of 3D‐printed patient‐based lung phantoms with accurate organ geometry, image texture, and attenuation profiles. PixelPrint will enable applications in the research and development of CT technology, including further development in radiomics. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=155147371
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1002/mp.15407
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 11
        StartPage: 825
    Subjects:
      – SubjectFull: Imaging phantoms
        Type: general
      – SubjectFull: Computed tomography
        Type: general
      – SubjectFull: Three-dimensional printing
        Type: general
      – SubjectFull: Lungs
        Type: general
      – SubjectFull: Radiomics
        Type: general
      – SubjectFull: Texture analysis (Image processing)
        Type: general
      – SubjectFull: Digital communications
        Type: general
    Titles:
      – TitleFull: Three‐dimensional printing of patient‐specific lung phantoms for CT imaging: Emulating lung tissue with accurate attenuation profiles and textures.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Mei, Kai
      – PersonEntity:
          Name:
            NameFull: Geagan, Michael
      – PersonEntity:
          Name:
            NameFull: Roshkovan, Leonid
      – PersonEntity:
          Name:
            NameFull: Litt, Harold I.
      – PersonEntity:
          Name:
            NameFull: Gang, Grace J.
      – PersonEntity:
          Name:
            NameFull: Shapira, Nadav
      – PersonEntity:
          Name:
            NameFull: Stayman, J. Webster
      – PersonEntity:
          Name:
            NameFull: Noël, Peter B.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 02
              Text: Feb2022
              Type: published
              Y: 2022
          Identifiers:
            – Type: issn-print
              Value: 00942405
          Numbering:
            – Type: volume
              Value: 49
            – Type: issue
              Value: 2
          Titles:
            – TitleFull: Medical Physics
              Type: main
ResultId 1