Analysis of artefacts and detail resolution of lung MRI with breath-hold T1-weighted gradient-echo and T2-weighted fast spin-echo sequences with respiratory triggering.

Saved in:
Bibliographic Details
Title: Analysis of artefacts and detail resolution of lung MRI with breath-hold T1-weighted gradient-echo and T2-weighted fast spin-echo sequences with respiratory triggering.
Authors: Biederer, J.1 juergen.biederer@rad.uni-kiel.de, Reuter, M.1, Both, M.1, Muhle, C.2, Grimm, J.1, Graessner, J.3, Heller, M.1
Source: European Radiology. Feb2002, Vol. 12 Issue 2, p378-384. 7p.
Subjects: Lungs, Respiration, Mediastinum, Bronchi, Blood vessels, Chest (Anatomy)
Abstract: The aim of this study was to evaluate feasibility and limitations of two MR sequences for imaging of the lung using a semi-quantitative rating scale. Ten healthy volunteers were assessed with a breath-hold T1-weighted gradient-recalled-echo (TR/TE=129/2.2 ms, matrix 173 x 256) and a T2-weighted turbo spin-echo (TSE) sequence with respiratory triggering (TR/TE=3000-5000/120 ms, matrix 270 x 512) in axial 6-mm slices. The T1-weighted GRE protocol included a pre-saturation pulse over the mediastinal structures. Artefacts and resolution of vessel/airway structures in each lung segment were evaluated by two observers (10 volunteers, 180 segments). Cardiac and vessel pulsation artefacts predominated on T1-weighted GRE, respiration artefacts on T2-weighted TSE (lingula and middle lobe). Pre-saturation of the mediastinum reduced pulsation artefacts on T1-weighted GRE. T1-weighted GRE images were improved by bright flow signal of vessels, whereas image quality of T2-weighted TSE was reduced by black-blood effects in central parts of the lung. Delineation of lung periphery and the mediastinum was superior with T2-weighted TSE. Segmental/sub-segmental vessels (up to fourth/fifth order) and bronchi (up to third order) were identified. All 180 lung segments were imaged in diagnostic quality with at least one of the two sequences (T1-weighted GRE not diagnostic in 9 of 180, T2-weighted TSE in 4 of 180). Both sequences were found to be complementary: superior identification of gross lung anatomy with T1-weighted GRE and higher detail resolution in the periphery and the mediastinum with T2--weighted TSE. [ABSTRACT FROM AUTHOR]
Copyright of European Radiology is the property of Springer Nature and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Links:
  – Type: pdflink
Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 15654681
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Analysis of artefacts and detail resolution of lung MRI with breath-hold T1-weighted gradient-echo and T2-weighted fast spin-echo sequences with respiratory triggering.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Biederer%2C+J%2E%22">Biederer, J.</searchLink><relatesTo>1</relatesTo><i> juergen.biederer@rad.uni-kiel.de</i><br /><searchLink fieldCode="AR" term="%22Reuter%2C+M%2E%22">Reuter, M.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Both%2C+M%2E%22">Both, M.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Muhle%2C+C%2E%22">Muhle, C.</searchLink><relatesTo>2</relatesTo><br /><searchLink fieldCode="AR" term="%22Grimm%2C+J%2E%22">Grimm, J.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Graessner%2C+J%2E%22">Graessner, J.</searchLink><relatesTo>3</relatesTo><br /><searchLink fieldCode="AR" term="%22Heller%2C+M%2E%22">Heller, M.</searchLink><relatesTo>1</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22European+Radiology%22">European Radiology</searchLink>. Feb2002, Vol. 12 Issue 2, p378-384. 7p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Lungs%22">Lungs</searchLink><br /><searchLink fieldCode="DE" term="%22Respiration%22">Respiration</searchLink><br /><searchLink fieldCode="DE" term="%22Mediastinum%22">Mediastinum</searchLink><br /><searchLink fieldCode="DE" term="%22Bronchi%22">Bronchi</searchLink><br /><searchLink fieldCode="DE" term="%22Blood+vessels%22">Blood vessels</searchLink><br /><searchLink fieldCode="DE" term="%22Chest+%28Anatomy%29%22">Chest (Anatomy)</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: The aim of this study was to evaluate feasibility and limitations of two MR sequences for imaging of the lung using a semi-quantitative rating scale. Ten healthy volunteers were assessed with a breath-hold T1-weighted gradient-recalled-echo (TR/TE=129/2.2 ms, matrix 173 x 256) and a T2-weighted turbo spin-echo (TSE) sequence with respiratory triggering (TR/TE=3000-5000/120 ms, matrix 270 x 512) in axial 6-mm slices. The T1-weighted GRE protocol included a pre-saturation pulse over the mediastinal structures. Artefacts and resolution of vessel/airway structures in each lung segment were evaluated by two observers (10 volunteers, 180 segments). Cardiac and vessel pulsation artefacts predominated on T1-weighted GRE, respiration artefacts on T2-weighted TSE (lingula and middle lobe). Pre-saturation of the mediastinum reduced pulsation artefacts on T1-weighted GRE. T1-weighted GRE images were improved by bright flow signal of vessels, whereas image quality of T2-weighted TSE was reduced by black-blood effects in central parts of the lung. Delineation of lung periphery and the mediastinum was superior with T2-weighted TSE. Segmental/sub-segmental vessels (up to fourth/fifth order) and bronchi (up to third order) were identified. All 180 lung segments were imaged in diagnostic quality with at least one of the two sequences (T1-weighted GRE not diagnostic in 9 of 180, T2-weighted TSE in 4 of 180). Both sequences were found to be complementary: superior identification of gross lung anatomy with T1-weighted GRE and higher detail resolution in the periphery and the mediastinum with T2--weighted TSE. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of European Radiology is the property of Springer Nature and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=15654681
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1007/s00330-001-1147-7
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 7
        StartPage: 378
    Subjects:
      – SubjectFull: Lungs
        Type: general
      – SubjectFull: Respiration
        Type: general
      – SubjectFull: Mediastinum
        Type: general
      – SubjectFull: Bronchi
        Type: general
      – SubjectFull: Blood vessels
        Type: general
      – SubjectFull: Chest (Anatomy)
        Type: general
    Titles:
      – TitleFull: Analysis of artefacts and detail resolution of lung MRI with breath-hold T1-weighted gradient-echo and T2-weighted fast spin-echo sequences with respiratory triggering.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Biederer, J.
      – PersonEntity:
          Name:
            NameFull: Reuter, M.
      – PersonEntity:
          Name:
            NameFull: Both, M.
      – PersonEntity:
          Name:
            NameFull: Muhle, C.
      – PersonEntity:
          Name:
            NameFull: Grimm, J.
      – PersonEntity:
          Name:
            NameFull: Graessner, J.
      – PersonEntity:
          Name:
            NameFull: Heller, M.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 02
              Text: Feb2002
              Type: published
              Y: 2002
          Identifiers:
            – Type: issn-print
              Value: 09387994
          Numbering:
            – Type: volume
              Value: 12
            – Type: issue
              Value: 2
          Titles:
            – TitleFull: European Radiology
              Type: main
ResultId 1