Hybrid active and passive local shimming (HAPLS) for two‐region MRI.
Saved in:
| Title: | Hybrid active and passive local shimming (HAPLS) for two‐region MRI. |
|---|---|
| Authors: | Ren, Zhi Hua1 (AUTHOR), Stockmann, Jason2,3 (AUTHOR), Dewdney, Andrew4 (AUTHOR), Lee, Ray F.1 (AUTHOR) rl2946@columbia.edu |
| Source: | Magnetic Resonance in Medicine. Apr2023, Vol. 89 Issue 4, p1660-1673. 14p. |
| Subjects: | Magnetic resonance imaging, Current distribution, Magnets, Scanning systems, Magnetic susceptibility |
| Abstract: | Purpose: An MRI scanner is equipped with global shim systems for shimming one region of interest (ROI) only. However, it often fails to reach state‐of‐the‐art when shimming two isolated regions of interest simultaneously, even though the two‐area shimming can be essential in scan scenarios, such as bilateral breasts or dyadic brains. To address these challenges, a hybrid active and passive local shimming technique is proposed to simultaneously shim two isolated region‐of‐interest areas within the whole FOV. Methods: A local passive shimming system is constructed by optimized bilateral ferromagnetic chip arrays to compensate for the magnet's significant high‐order B0 inhomogeneities at the boundary of the manufacturer's specified homogeneous volume, thus locally improving the available FOV. The local active shimming consists of 40‐channel DC loops powered by 64‐channel current amplifiers. With the optimized current distribution, active shimming can correct the residual low‐order B0 inhomogeneities and subject‐specific field inhomogeneities. In addition, active shimming is used to homogenize the center frequencies of the two regions. Results: With the implementation of the hybrid active and passive local shimming, the 95% peak‐to‐peak was reduced from 1.92 to 1.12 ppm by 41.7%, and RMS decreased from 0.473 to 0.255 ppm by 46.1% in a two‐phantom experiment. The volume ratio containing MR voxels within a 0.5‐ppm frequency span increased from 64.3% to 81.3% by 26.3%. Conclusion: The proposed hybrid active and passive local shimming technique uses both passive and active local shimming, and it can efficiently shim two areas simultaneously, which is an unmet need for a commercial MRI scanner. [ABSTRACT FROM AUTHOR] |
| Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
|
Full text is not displayed to guests.
Login for full access.
|
|
| FullText | Links: – Type: pdflink Text: Availability: 1 |
|---|---|
| Header | DbId: egs DbLabel: Engineering Source An: 161587725 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Hybrid active and passive local shimming (HAPLS) for two‐region MRI. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Ren%2C+Zhi+Hua%22">Ren, Zhi Hua</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Stockmann%2C+Jason%22">Stockmann, Jason</searchLink><relatesTo>2,3</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Dewdney%2C+Andrew%22">Dewdney, Andrew</searchLink><relatesTo>4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Lee%2C+Ray+F%2E%22">Lee, Ray F.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<i> rl2946@columbia.edu</i> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+in+Medicine%22">Magnetic Resonance in Medicine</searchLink>. Apr2023, Vol. 89 Issue 4, p1660-1673. 14p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22Magnetic+resonance+imaging%22">Magnetic resonance imaging</searchLink><br /><searchLink fieldCode="DE" term="%22Current+distribution%22">Current distribution</searchLink><br /><searchLink fieldCode="DE" term="%22Magnets%22">Magnets</searchLink><br /><searchLink fieldCode="DE" term="%22Scanning+systems%22">Scanning systems</searchLink><br /><searchLink fieldCode="DE" term="%22Magnetic+susceptibility%22">Magnetic susceptibility</searchLink> – Name: Abstract Label: Abstract Group: Ab Data: Purpose: An MRI scanner is equipped with global shim systems for shimming one region of interest (ROI) only. However, it often fails to reach state‐of‐the‐art when shimming two isolated regions of interest simultaneously, even though the two‐area shimming can be essential in scan scenarios, such as bilateral breasts or dyadic brains. To address these challenges, a hybrid active and passive local shimming technique is proposed to simultaneously shim two isolated region‐of‐interest areas within the whole FOV. Methods: A local passive shimming system is constructed by optimized bilateral ferromagnetic chip arrays to compensate for the magnet's significant high‐order B0 inhomogeneities at the boundary of the manufacturer's specified homogeneous volume, thus locally improving the available FOV. The local active shimming consists of 40‐channel DC loops powered by 64‐channel current amplifiers. With the optimized current distribution, active shimming can correct the residual low‐order B0 inhomogeneities and subject‐specific field inhomogeneities. In addition, active shimming is used to homogenize the center frequencies of the two regions. Results: With the implementation of the hybrid active and passive local shimming, the 95% peak‐to‐peak was reduced from 1.92 to 1.12 ppm by 41.7%, and RMS decreased from 0.473 to 0.255 ppm by 46.1% in a two‐phantom experiment. The volume ratio containing MR voxels within a 0.5‐ppm frequency span increased from 64.3% to 81.3% by 26.3%. Conclusion: The proposed hybrid active and passive local shimming technique uses both passive and active local shimming, and it can efficiently shim two areas simultaneously, which is an unmet need for a commercial MRI scanner. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=161587725 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.1002/mrm.29542 Languages: – Code: eng Text: English PhysicalDescription: Pagination: PageCount: 14 StartPage: 1660 Subjects: – SubjectFull: Magnetic resonance imaging Type: general – SubjectFull: Current distribution Type: general – SubjectFull: Magnets Type: general – SubjectFull: Scanning systems Type: general – SubjectFull: Magnetic susceptibility Type: general Titles: – TitleFull: Hybrid active and passive local shimming (HAPLS) for two‐region MRI. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Ren, Zhi Hua – PersonEntity: Name: NameFull: Stockmann, Jason – PersonEntity: Name: NameFull: Dewdney, Andrew – PersonEntity: Name: NameFull: Lee, Ray F. IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 04 Text: Apr2023 Type: published Y: 2023 Identifiers: – Type: issn-print Value: 07403194 Numbering: – Type: volume Value: 89 – Type: issue Value: 4 Titles: – TitleFull: Magnetic Resonance in Medicine Type: main |
| ResultId | 1 |