Replication of the bSTAR sequence and open‐source implementation.

Saved in:
Bibliographic Details
Title: Replication of the bSTAR sequence and open‐source implementation.
Authors: Lee, Nam G.1,2 (AUTHOR) namgyunl@usc.edu, Bauman, Grzegorz3,4 (AUTHOR), Bieri, Oliver3,4 (AUTHOR), Nayak, Krishna S.1,2 (AUTHOR)
Source: Magnetic Resonance in Medicine. Apr2024, Vol. 91 Issue 4, p1464-1477. 14p.
Subjects: Image reconstruction, Spatial resolution, Rapid prototyping, Technical reports, Research methodology
Abstract: Purpose: The reproducibility of scientific reports is crucial to advancing human knowledge. This paper is a summary of our experience in replicating a balanced SSFP half‐radial dual‐echo imaging technique (bSTAR) using open‐source frameworks as a response to the 2023 ISMRM "repeat it with me" Challenge. Methods: We replicated the bSTAR technique for thoracic imaging at 0.55T. The bSTAR pulse sequence is implemented in Pulseq, a vendor neutral open‐source rapid sequence prototyping environment. Image reconstruction is performed with the open‐source Berkeley Advanced Reconstruction Toolbox (BART). The replication of bSTAR, termed open‐source bSTAR, is tested by replicating several figures from the published literature. Original bSTAR, using the pulse sequence and image reconstruction developed by the original authors, and open‐source bSTAR, with pulse sequence and image reconstruction developed in this work, were performed in healthy volunteers. Results: Both echo images obtained from open‐source bSTAR contain no visible artifacts and show identical spatial resolution and image quality to those in the published literature. A direct head‐to‐head comparison between open‐source bSTAR and original bSTAR on a healthy volunteer indicates that open‐source bSTAR provides adequate SNR, spatial resolution, level of artifacts, and conspicuity of pulmonary vessels comparable to original bSTAR. Conclusion: We have successfully replicated bSTAR lung imaging at 0.55T using two open‐source frameworks. Full replication of a research method solely relying on information on a research paper is unfortunately rare in research, but our success gives greater confidence that a research methodology can be indeed replicated as described. [ABSTRACT FROM AUTHOR]
Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 175196980
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Replication of the bSTAR sequence and open‐source implementation.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Lee%2C+Nam+G%2E%22">Lee, Nam G.</searchLink><relatesTo>1,2</relatesTo> (AUTHOR)<i> namgyunl@usc.edu</i><br /><searchLink fieldCode="AR" term="%22Bauman%2C+Grzegorz%22">Bauman, Grzegorz</searchLink><relatesTo>3,4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Bieri%2C+Oliver%22">Bieri, Oliver</searchLink><relatesTo>3,4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Nayak%2C+Krishna+S%2E%22">Nayak, Krishna S.</searchLink><relatesTo>1,2</relatesTo> (AUTHOR)
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+in+Medicine%22">Magnetic Resonance in Medicine</searchLink>. Apr2024, Vol. 91 Issue 4, p1464-1477. 14p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Image+reconstruction%22">Image reconstruction</searchLink><br /><searchLink fieldCode="DE" term="%22Spatial+resolution%22">Spatial resolution</searchLink><br /><searchLink fieldCode="DE" term="%22Rapid+prototyping%22">Rapid prototyping</searchLink><br /><searchLink fieldCode="DE" term="%22Technical+reports%22">Technical reports</searchLink><br /><searchLink fieldCode="DE" term="%22Research+methodology%22">Research methodology</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: The reproducibility of scientific reports is crucial to advancing human knowledge. This paper is a summary of our experience in replicating a balanced SSFP half‐radial dual‐echo imaging technique (bSTAR) using open‐source frameworks as a response to the 2023 ISMRM "repeat it with me" Challenge. Methods: We replicated the bSTAR technique for thoracic imaging at 0.55T. The bSTAR pulse sequence is implemented in Pulseq, a vendor neutral open‐source rapid sequence prototyping environment. Image reconstruction is performed with the open‐source Berkeley Advanced Reconstruction Toolbox (BART). The replication of bSTAR, termed open‐source bSTAR, is tested by replicating several figures from the published literature. Original bSTAR, using the pulse sequence and image reconstruction developed by the original authors, and open‐source bSTAR, with pulse sequence and image reconstruction developed in this work, were performed in healthy volunteers. Results: Both echo images obtained from open‐source bSTAR contain no visible artifacts and show identical spatial resolution and image quality to those in the published literature. A direct head‐to‐head comparison between open‐source bSTAR and original bSTAR on a healthy volunteer indicates that open‐source bSTAR provides adequate SNR, spatial resolution, level of artifacts, and conspicuity of pulmonary vessels comparable to original bSTAR. Conclusion: We have successfully replicated bSTAR lung imaging at 0.55T using two open‐source frameworks. Full replication of a research method solely relying on information on a research paper is unfortunately rare in research, but our success gives greater confidence that a research methodology can be indeed replicated as described. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=175196980
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1002/mrm.29947
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 14
        StartPage: 1464
    Subjects:
      – SubjectFull: Image reconstruction
        Type: general
      – SubjectFull: Spatial resolution
        Type: general
      – SubjectFull: Rapid prototyping
        Type: general
      – SubjectFull: Technical reports
        Type: general
      – SubjectFull: Research methodology
        Type: general
    Titles:
      – TitleFull: Replication of the bSTAR sequence and open‐source implementation.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Lee, Nam G.
      – PersonEntity:
          Name:
            NameFull: Bauman, Grzegorz
      – PersonEntity:
          Name:
            NameFull: Bieri, Oliver
      – PersonEntity:
          Name:
            NameFull: Nayak, Krishna S.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 04
              Text: Apr2024
              Type: published
              Y: 2024
          Identifiers:
            – Type: issn-print
              Value: 07403194
          Numbering:
            – Type: volume
              Value: 91
            – Type: issue
              Value: 4
          Titles:
            – TitleFull: Magnetic Resonance in Medicine
              Type: main
ResultId 1