Technical Note: Impact of impurities on Yttrium‐90 glass microsphere activity quantitation.

Saved in:
Bibliographic Details
Title: Technical Note: Impact of impurities on Yttrium‐90 glass microsphere activity quantitation.
Authors: Beattie, Bradley J.1 (AUTHOR), Kirov, Assen S.1 (AUTHOR), Kesner, Adam L.1 (AUTHOR) kesnera@mskcc.org
Source: Medical Physics. Mar2024, Vol. 51 Issue 3, p2306-2310. 5p.
Subjects: Radioactivity measurements, Ionization chambers, Technical specifications, Absorbed dose, Glass
Abstract: Background: Glass 90Y microspheres are produced with known radionuclide impurities. These impurities are not independently monitored. Clinical instruments, including ionization chamber dose calibrators and positron emmission tomography (PET) cameras, can be much more sensitive in detecting signals from these impurities than to signals from 90Y itself. Purpose: The "typical" levels of 90Y impurities have been studied to assess their impact on dosimetry during internal implantation, and for the management of waste. However, unaccounted‐for decay spectra of impurities can also have an impact on dose calibrator and PET readings. Thus, even what might be considered negligibly small impurity fractions, can in principle cause substantial overestimates of the amount of 90Y activity present in a sample. To our knowledge, quantitative effects of radionuclide impurities in glass microspheres on activity measurements have not been documented in the field. As activity quantitation for dosimetry and its correlations with outcome becomes more prevalent, the effects of impurities on measurements may remain unaccounted for in dosimetry studies. Methods: In this letter, we review theoretical and physical considerations that will result in asymmetric errors in quantitation from 90Y impurities and estimate their typical and potential impact on clinical utilization. Among the common impurities 88Y is of particular concern for its impact on 90Y dose measurements because of its decay characteristics, along with other isotopes 91Y and 46Sc which can also impact measurements. Results: The typical level of 88Y impurities reported by the manufacturer should only cause small errors in dose calibrator and PET measurements made within the 12‐day label‐specified use‐by period, up to 2.0% and 1.6%, respectively. However, the product specification max allowable impurity levels, specified by the manufacturer, leave open the potential for much greater bias from within the 12‐day use‐by period, potentially as high as 13.2% for dose calibrator measurements and 10.6% for PET from the 88Y impurities. Conclusions: While typical levels of impurities appear to have acceptable impact on patient absorbed dose, it should be noted that they can have adverse effects on 90Y radioactivity measurements. Furthermore, there is currently minimal independent verification and/or monitoring of impurity levels within the field. [ABSTRACT FROM AUTHOR]
Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 175989212
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Technical Note: Impact of impurities on Yttrium‐90 glass microsphere activity quantitation.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Beattie%2C+Bradley+J%2E%22">Beattie, Bradley J.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Kirov%2C+Assen+S%2E%22">Kirov, Assen S.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Kesner%2C+Adam+L%2E%22">Kesner, Adam L.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<i> kesnera@mskcc.org</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. Mar2024, Vol. 51 Issue 3, p2306-2310. 5p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Radioactivity+measurements%22">Radioactivity measurements</searchLink><br /><searchLink fieldCode="DE" term="%22Ionization+chambers%22">Ionization chambers</searchLink><br /><searchLink fieldCode="DE" term="%22Technical+specifications%22">Technical specifications</searchLink><br /><searchLink fieldCode="DE" term="%22Absorbed+dose%22">Absorbed dose</searchLink><br /><searchLink fieldCode="DE" term="%22Glass%22">Glass</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Background: Glass 90Y microspheres are produced with known radionuclide impurities. These impurities are not independently monitored. Clinical instruments, including ionization chamber dose calibrators and positron emmission tomography (PET) cameras, can be much more sensitive in detecting signals from these impurities than to signals from 90Y itself. Purpose: The "typical" levels of 90Y impurities have been studied to assess their impact on dosimetry during internal implantation, and for the management of waste. However, unaccounted‐for decay spectra of impurities can also have an impact on dose calibrator and PET readings. Thus, even what might be considered negligibly small impurity fractions, can in principle cause substantial overestimates of the amount of 90Y activity present in a sample. To our knowledge, quantitative effects of radionuclide impurities in glass microspheres on activity measurements have not been documented in the field. As activity quantitation for dosimetry and its correlations with outcome becomes more prevalent, the effects of impurities on measurements may remain unaccounted for in dosimetry studies. Methods: In this letter, we review theoretical and physical considerations that will result in asymmetric errors in quantitation from 90Y impurities and estimate their typical and potential impact on clinical utilization. Among the common impurities 88Y is of particular concern for its impact on 90Y dose measurements because of its decay characteristics, along with other isotopes 91Y and 46Sc which can also impact measurements. Results: The typical level of 88Y impurities reported by the manufacturer should only cause small errors in dose calibrator and PET measurements made within the 12‐day label‐specified use‐by period, up to 2.0% and 1.6%, respectively. However, the product specification max allowable impurity levels, specified by the manufacturer, leave open the potential for much greater bias from within the 12‐day use‐by period, potentially as high as 13.2% for dose calibrator measurements and 10.6% for PET from the 88Y impurities. Conclusions: While typical levels of impurities appear to have acceptable impact on patient absorbed dose, it should be noted that they can have adverse effects on 90Y radioactivity measurements. Furthermore, there is currently minimal independent verification and/or monitoring of impurity levels within the field. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=175989212
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1002/mp.16828
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 5
        StartPage: 2306
    Subjects:
      – SubjectFull: Radioactivity measurements
        Type: general
      – SubjectFull: Ionization chambers
        Type: general
      – SubjectFull: Technical specifications
        Type: general
      – SubjectFull: Absorbed dose
        Type: general
      – SubjectFull: Glass
        Type: general
    Titles:
      – TitleFull: Technical Note: Impact of impurities on Yttrium‐90 glass microsphere activity quantitation.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Beattie, Bradley J.
      – PersonEntity:
          Name:
            NameFull: Kirov, Assen S.
      – PersonEntity:
          Name:
            NameFull: Kesner, Adam L.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 03
              Text: Mar2024
              Type: published
              Y: 2024
          Identifiers:
            – Type: issn-print
              Value: 00942405
          Numbering:
            – Type: volume
              Value: 51
            – Type: issue
              Value: 3
          Titles:
            – TitleFull: Medical Physics
              Type: main
ResultId 1