Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy.
Saved in:
| Title: | Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy. |
|---|---|
| Authors: | Foongkajornkiat, Satcha1 (AUTHOR), Sokolowski, Kamil2 (AUTHOR), Stephenson, James3,4 (AUTHOR), Lloyd, Thomas5 (AUTHOR), Hugo, Honor J.6,7 (AUTHOR), Thompson, Erik W.4,8 (AUTHOR), Momot, Konstantin I.1 (AUTHOR) k.momot@qut.edu.au |
| Source: | Magnetic Resonance in Medicine. Jul2024, Vol. 92 Issue 1, p374-388. 15p. |
| Subjects: | Hematoxylin & eosin staining, Bland-Altman plot, Density, Confidence intervals, Nuclear magnetic resonance |
| Abstract: | Purpose: Single‐sided portable NMR (pNMR) has previously been demonstrated to be suitable for quantification of mammographic density (MD) in excised breast tissue samples. Here we investigate the precision and accuracy of pNMR measurements of MD ex vivo as compared with the gold standards. Methods: Forty‐five breast‐tissue explants from 9 prophylactic mastectomy patients were measured. The relative tissue water content was taken as the MD‐equivalent quantity. In each sample, the water content was measured using some combination of three pNMR techniques (apparent T2, diffusion, and T1 measurements) and two gold‐standard techniques (computed microtomography [μCT] and hematoxylin and eosin [H&E] histology). Pairwise correlation plots and Bland–Altman analysis were used to quantify the degree of agreement between pNMR techniques and the gold standards. Results: Relative water content measured from both apparent T2 relaxation spectra, and diffusion decays exhibited strong correlation with the H&E and μCT results. Bland–Altman analysis yielded average bias values of −0.4, −2.6, 2.6, and 2.8 water percentage points (pp) and 95% confidence intervals of 13.1, 7.5, 11.2, and 11.8 pp for the H&E – T2, μCT – T2, H&E – diffusion, and μCT – diffusion comparison pairs, respectively. T1‐based measurements were found to be less reliable, with the Bland–Altman confidence intervals of 27.7 and 33.0 pp when compared with H&E and μCT, respectively. Conclusion: Apparent T2‐based and diffusion‐based pNMR measurements enable quantification of MD in breast‐tissue explants with the precision of approximately 10 pp and accuracy of approximately 3 pp or better, making pNMR a promising measurement modality for radiation‐free quantification of MD. [ABSTRACT FROM AUTHOR] |
| Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
|
Full text is not displayed to guests.
Login for full access.
|
|
| FullText | Links: – Type: pdflink Text: Availability: 1 |
|---|---|
| Header | DbId: egs DbLabel: Engineering Source An: 176868680 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Foongkajornkiat%2C+Satcha%22">Foongkajornkiat, Satcha</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Sokolowski%2C+Kamil%22">Sokolowski, Kamil</searchLink><relatesTo>2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Stephenson%2C+James%22">Stephenson, James</searchLink><relatesTo>3,4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Lloyd%2C+Thomas%22">Lloyd, Thomas</searchLink><relatesTo>5</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Hugo%2C+Honor+J%2E%22">Hugo, Honor J.</searchLink><relatesTo>6,7</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Thompson%2C+Erik+W%2E%22">Thompson, Erik W.</searchLink><relatesTo>4,8</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Momot%2C+Konstantin+I%2E%22">Momot, Konstantin I.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<i> k.momot@qut.edu.au</i> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+in+Medicine%22">Magnetic Resonance in Medicine</searchLink>. Jul2024, Vol. 92 Issue 1, p374-388. 15p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22Hematoxylin+%26+eosin+staining%22">Hematoxylin & eosin staining</searchLink><br /><searchLink fieldCode="DE" term="%22Bland-Altman+plot%22">Bland-Altman plot</searchLink><br /><searchLink fieldCode="DE" term="%22Density%22">Density</searchLink><br /><searchLink fieldCode="DE" term="%22Confidence+intervals%22">Confidence intervals</searchLink><br /><searchLink fieldCode="DE" term="%22Nuclear+magnetic+resonance%22">Nuclear magnetic resonance</searchLink> – Name: Abstract Label: Abstract Group: Ab Data: Purpose: Single‐sided portable NMR (pNMR) has previously been demonstrated to be suitable for quantification of mammographic density (MD) in excised breast tissue samples. Here we investigate the precision and accuracy of pNMR measurements of MD ex vivo as compared with the gold standards. Methods: Forty‐five breast‐tissue explants from 9 prophylactic mastectomy patients were measured. The relative tissue water content was taken as the MD‐equivalent quantity. In each sample, the water content was measured using some combination of three pNMR techniques (apparent T2, diffusion, and T1 measurements) and two gold‐standard techniques (computed microtomography [μCT] and hematoxylin and eosin [H&E] histology). Pairwise correlation plots and Bland–Altman analysis were used to quantify the degree of agreement between pNMR techniques and the gold standards. Results: Relative water content measured from both apparent T2 relaxation spectra, and diffusion decays exhibited strong correlation with the H&E and μCT results. Bland–Altman analysis yielded average bias values of −0.4, −2.6, 2.6, and 2.8 water percentage points (pp) and 95% confidence intervals of 13.1, 7.5, 11.2, and 11.8 pp for the H&E – T2, μCT – T2, H&E – diffusion, and μCT – diffusion comparison pairs, respectively. T1‐based measurements were found to be less reliable, with the Bland–Altman confidence intervals of 27.7 and 33.0 pp when compared with H&E and μCT, respectively. Conclusion: Apparent T2‐based and diffusion‐based pNMR measurements enable quantification of MD in breast‐tissue explants with the precision of approximately 10 pp and accuracy of approximately 3 pp or better, making pNMR a promising measurement modality for radiation‐free quantification of MD. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=176868680 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.1002/mrm.30040 Languages: – Code: eng Text: English PhysicalDescription: Pagination: PageCount: 15 StartPage: 374 Subjects: – SubjectFull: Hematoxylin & eosin staining Type: general – SubjectFull: Bland-Altman plot Type: general – SubjectFull: Density Type: general – SubjectFull: Confidence intervals Type: general – SubjectFull: Nuclear magnetic resonance Type: general Titles: – TitleFull: Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Foongkajornkiat, Satcha – PersonEntity: Name: NameFull: Sokolowski, Kamil – PersonEntity: Name: NameFull: Stephenson, James – PersonEntity: Name: NameFull: Lloyd, Thomas – PersonEntity: Name: NameFull: Hugo, Honor J. – PersonEntity: Name: NameFull: Thompson, Erik W. – PersonEntity: Name: NameFull: Momot, Konstantin I. IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 07 Text: Jul2024 Type: published Y: 2024 Identifiers: – Type: issn-print Value: 07403194 Numbering: – Type: volume Value: 92 – Type: issue Value: 1 Titles: – TitleFull: Magnetic Resonance in Medicine Type: main |
| ResultId | 1 |