Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy.

Saved in:
Bibliographic Details
Title: Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy.
Authors: Foongkajornkiat, Satcha1 (AUTHOR), Sokolowski, Kamil2 (AUTHOR), Stephenson, James3,4 (AUTHOR), Lloyd, Thomas5 (AUTHOR), Hugo, Honor J.6,7 (AUTHOR), Thompson, Erik W.4,8 (AUTHOR), Momot, Konstantin I.1 (AUTHOR) k.momot@qut.edu.au
Source: Magnetic Resonance in Medicine. Jul2024, Vol. 92 Issue 1, p374-388. 15p.
Subjects: Hematoxylin & eosin staining, Bland-Altman plot, Density, Confidence intervals, Nuclear magnetic resonance
Abstract: Purpose: Single‐sided portable NMR (pNMR) has previously been demonstrated to be suitable for quantification of mammographic density (MD) in excised breast tissue samples. Here we investigate the precision and accuracy of pNMR measurements of MD ex vivo as compared with the gold standards. Methods: Forty‐five breast‐tissue explants from 9 prophylactic mastectomy patients were measured. The relative tissue water content was taken as the MD‐equivalent quantity. In each sample, the water content was measured using some combination of three pNMR techniques (apparent T2, diffusion, and T1 measurements) and two gold‐standard techniques (computed microtomography [μCT] and hematoxylin and eosin [H&E] histology). Pairwise correlation plots and Bland–Altman analysis were used to quantify the degree of agreement between pNMR techniques and the gold standards. Results: Relative water content measured from both apparent T2 relaxation spectra, and diffusion decays exhibited strong correlation with the H&E and μCT results. Bland–Altman analysis yielded average bias values of −0.4, −2.6, 2.6, and 2.8 water percentage points (pp) and 95% confidence intervals of 13.1, 7.5, 11.2, and 11.8 pp for the H&E – T2, μCT – T2, H&E – diffusion, and μCT – diffusion comparison pairs, respectively. T1‐based measurements were found to be less reliable, with the Bland–Altman confidence intervals of 27.7 and 33.0 pp when compared with H&E and μCT, respectively. Conclusion: Apparent T2‐based and diffusion‐based pNMR measurements enable quantification of MD in breast‐tissue explants with the precision of approximately 10 pp and accuracy of approximately 3 pp or better, making pNMR a promising measurement modality for radiation‐free quantification of MD. [ABSTRACT FROM AUTHOR]
Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 176868680
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Foongkajornkiat%2C+Satcha%22">Foongkajornkiat, Satcha</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Sokolowski%2C+Kamil%22">Sokolowski, Kamil</searchLink><relatesTo>2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Stephenson%2C+James%22">Stephenson, James</searchLink><relatesTo>3,4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Lloyd%2C+Thomas%22">Lloyd, Thomas</searchLink><relatesTo>5</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Hugo%2C+Honor+J%2E%22">Hugo, Honor J.</searchLink><relatesTo>6,7</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Thompson%2C+Erik+W%2E%22">Thompson, Erik W.</searchLink><relatesTo>4,8</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Momot%2C+Konstantin+I%2E%22">Momot, Konstantin I.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<i> k.momot@qut.edu.au</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+in+Medicine%22">Magnetic Resonance in Medicine</searchLink>. Jul2024, Vol. 92 Issue 1, p374-388. 15p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Hematoxylin+%26+eosin+staining%22">Hematoxylin & eosin staining</searchLink><br /><searchLink fieldCode="DE" term="%22Bland-Altman+plot%22">Bland-Altman plot</searchLink><br /><searchLink fieldCode="DE" term="%22Density%22">Density</searchLink><br /><searchLink fieldCode="DE" term="%22Confidence+intervals%22">Confidence intervals</searchLink><br /><searchLink fieldCode="DE" term="%22Nuclear+magnetic+resonance%22">Nuclear magnetic resonance</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: Single‐sided portable NMR (pNMR) has previously been demonstrated to be suitable for quantification of mammographic density (MD) in excised breast tissue samples. Here we investigate the precision and accuracy of pNMR measurements of MD ex vivo as compared with the gold standards. Methods: Forty‐five breast‐tissue explants from 9 prophylactic mastectomy patients were measured. The relative tissue water content was taken as the MD‐equivalent quantity. In each sample, the water content was measured using some combination of three pNMR techniques (apparent T2, diffusion, and T1 measurements) and two gold‐standard techniques (computed microtomography [μCT] and hematoxylin and eosin [H&E] histology). Pairwise correlation plots and Bland–Altman analysis were used to quantify the degree of agreement between pNMR techniques and the gold standards. Results: Relative water content measured from both apparent T2 relaxation spectra, and diffusion decays exhibited strong correlation with the H&E and μCT results. Bland–Altman analysis yielded average bias values of −0.4, −2.6, 2.6, and 2.8 water percentage points (pp) and 95% confidence intervals of 13.1, 7.5, 11.2, and 11.8 pp for the H&E – T2, μCT – T2, H&E – diffusion, and μCT – diffusion comparison pairs, respectively. T1‐based measurements were found to be less reliable, with the Bland–Altman confidence intervals of 27.7 and 33.0 pp when compared with H&E and μCT, respectively. Conclusion: Apparent T2‐based and diffusion‐based pNMR measurements enable quantification of MD in breast‐tissue explants with the precision of approximately 10 pp and accuracy of approximately 3 pp or better, making pNMR a promising measurement modality for radiation‐free quantification of MD. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=176868680
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1002/mrm.30040
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 15
        StartPage: 374
    Subjects:
      – SubjectFull: Hematoxylin & eosin staining
        Type: general
      – SubjectFull: Bland-Altman plot
        Type: general
      – SubjectFull: Density
        Type: general
      – SubjectFull: Confidence intervals
        Type: general
      – SubjectFull: Nuclear magnetic resonance
        Type: general
    Titles:
      – TitleFull: Quantitative measurement of mammographic density in breast‐tissue explants using portable NMR: Precision and accuracy.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Foongkajornkiat, Satcha
      – PersonEntity:
          Name:
            NameFull: Sokolowski, Kamil
      – PersonEntity:
          Name:
            NameFull: Stephenson, James
      – PersonEntity:
          Name:
            NameFull: Lloyd, Thomas
      – PersonEntity:
          Name:
            NameFull: Hugo, Honor J.
      – PersonEntity:
          Name:
            NameFull: Thompson, Erik W.
      – PersonEntity:
          Name:
            NameFull: Momot, Konstantin I.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 07
              Text: Jul2024
              Type: published
              Y: 2024
          Identifiers:
            – Type: issn-print
              Value: 07403194
          Numbering:
            – Type: volume
              Value: 92
            – Type: issue
              Value: 1
          Titles:
            – TitleFull: Magnetic Resonance in Medicine
              Type: main
ResultId 1