A Pilot Trial Testing a Novel 2D Antiscatter Grid and Quantitative CBCT Pipeline: An Interim Quantitative Image Quality Analysis.

Saved in:
Bibliographic Details
Title: A Pilot Trial Testing a Novel 2D Antiscatter Grid and Quantitative CBCT Pipeline: An Interim Quantitative Image Quality Analysis.
Authors: Bayat, F.1 (AUTHOR), Pyakurel, U.1 (AUTHOR), Sabounchi, R.2 (AUTHOR), Bliley, R.C.1 (AUTHOR), Fisher, C.M.3 (AUTHOR), Rabinovitch, R.A.3 (AUTHOR), Binder, D.C.4 (AUTHOR), Kavanagh, B.D.5 (AUTHOR), Lanning, R.M.5 (AUTHOR), Robin, T.P.5 (AUTHOR), Altunbas, C.6 (AUTHOR)
Source: International Journal of Radiation Oncology, Biology, Physics. 2024 Supplement, Vol. 120 Issue 2, pS224-S224. 1p.
Subjects: Cone beam computed tomography, Image quality analysis, Convex sets, Rear-screen projection, Diagnostic imaging
Abstract: Cone beam computerized tomography (CBCT) image quality plays an important role in target localization, response assessment, and treatment plan adaptation. However, standard CBCT images have lower quality than planning CT (pCT) scans, in part due to scattered X-rays, limiting their utility. To address this, we developed a novel quantitative CBCT (qCBCT) pipeline, featuring a 2-dimensional anti-scatter grid (2D ASG) and associated data processing methods. This interim study hypothesizes that qCBCT demonstrates improved image quality compared to standard CBCT, as measured by established image quality metrics. To date, 17 patients with cancers of the abdomen and pelvis were enrolled into this IRB-approved trial. Each participant underwent CBCT imaging twice within 30 minutes (pelvis CBCT protocol in the CBCT-linac as well as qCBCT protocol employing a prototype 2D ASG developed by our group). CBCT scan parameters, patient immobilization, and imaging dose were the same in both scans. Standard clinical CBCT images were reconstructed using both standard and advanced Iterative Reconstruction (IR) options in the CBCT-linac. The qCBCT raw projections were first processed offline using a data correction pipeline with in-house developed software and reconstructed by employing filtered back projection and Projection onto Convex Sets (POCS) based IR algorithms. Image quality metrics - structural similarity index (SSIM) and CT number accuracy with respect to the gold-standard pCT images, CT number nonuniformity, and Contrast to Noise Ratio (CNR) - were calculated for all CBCT images. In addition, trends in image quality metrics as a function of patients' average separation in AP and lateral directions were evaluated. Clinical CBCT and qCBCT images exhibited an average SSIM of 0.77±0.1 and 0.79±0.09, respectively, whereas average CT number accuracy of clinical CBCT and qCBCT was 39±43 HU and 18±17 HU, respectively. Additionally, CT nonuniformity in clinical CBCT scans and qCBCT was 35±41HU and 18±17 HU, respectively. Soft tissue CNR in both CBCT methods were comparable, 7±4 and 7±3. While image quality metrics in clinical CBCT and qCBCT images were more comparable in patients with lateral separations less than 35 cm, clinical CBCT image quality metrics degraded faster for separations larger than 35 cm. While structural similarity and CNR were comparable in clinical CBCT and qCBCT images, both CT number accuracy and nonuniformity were improved substantially with the qCBCT method. The use of a 2D anti-scatter grid with qCBCT did not have any adverse impact on soft tissue contrast and appearance. Lastly, patients with larger AP and lateral separations may benefit more from qCBCT's improved quality when compared to the current clinical CBCT. [ABSTRACT FROM AUTHOR]
Copyright of International Journal of Radiation Oncology, Biology, Physics is the property of Pergamon Press - An Imprint of Elsevier Science and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 179877218
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: A Pilot Trial Testing a Novel 2D Antiscatter Grid and Quantitative CBCT Pipeline: An Interim Quantitative Image Quality Analysis.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Bayat%2C+F%2E%22">Bayat, F.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Pyakurel%2C+U%2E%22">Pyakurel, U.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Sabounchi%2C+R%2E%22">Sabounchi, R.</searchLink><relatesTo>2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Bliley%2C+R%2EC%2E%22">Bliley, R.C.</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Fisher%2C+C%2EM%2E%22">Fisher, C.M.</searchLink><relatesTo>3</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Rabinovitch%2C+R%2EA%2E%22">Rabinovitch, R.A.</searchLink><relatesTo>3</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Binder%2C+D%2EC%2E%22">Binder, D.C.</searchLink><relatesTo>4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Kavanagh%2C+B%2ED%2E%22">Kavanagh, B.D.</searchLink><relatesTo>5</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Lanning%2C+R%2EM%2E%22">Lanning, R.M.</searchLink><relatesTo>5</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Robin%2C+T%2EP%2E%22">Robin, T.P.</searchLink><relatesTo>5</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Altunbas%2C+C%2E%22">Altunbas, C.</searchLink><relatesTo>6</relatesTo> (AUTHOR)
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22International+Journal+of+Radiation+Oncology%2C+Biology%2C+Physics%22">International Journal of Radiation Oncology, Biology, Physics</searchLink>. 2024 Supplement, Vol. 120 Issue 2, pS224-S224. 1p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Cone+beam+computed+tomography%22">Cone beam computed tomography</searchLink><br /><searchLink fieldCode="DE" term="%22Image+quality+analysis%22">Image quality analysis</searchLink><br /><searchLink fieldCode="DE" term="%22Convex+sets%22">Convex sets</searchLink><br /><searchLink fieldCode="DE" term="%22Rear-screen+projection%22">Rear-screen projection</searchLink><br /><searchLink fieldCode="DE" term="%22Diagnostic+imaging%22">Diagnostic imaging</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Cone beam computerized tomography (CBCT) image quality plays an important role in target localization, response assessment, and treatment plan adaptation. However, standard CBCT images have lower quality than planning CT (pCT) scans, in part due to scattered X-rays, limiting their utility. To address this, we developed a novel quantitative CBCT (qCBCT) pipeline, featuring a 2-dimensional anti-scatter grid (2D ASG) and associated data processing methods. This interim study hypothesizes that qCBCT demonstrates improved image quality compared to standard CBCT, as measured by established image quality metrics. To date, 17 patients with cancers of the abdomen and pelvis were enrolled into this IRB-approved trial. Each participant underwent CBCT imaging twice within 30 minutes (pelvis CBCT protocol in the CBCT-linac as well as qCBCT protocol employing a prototype 2D ASG developed by our group). CBCT scan parameters, patient immobilization, and imaging dose were the same in both scans. Standard clinical CBCT images were reconstructed using both standard and advanced Iterative Reconstruction (IR) options in the CBCT-linac. The qCBCT raw projections were first processed offline using a data correction pipeline with in-house developed software and reconstructed by employing filtered back projection and Projection onto Convex Sets (POCS) based IR algorithms. Image quality metrics - structural similarity index (SSIM) and CT number accuracy with respect to the gold-standard pCT images, CT number nonuniformity, and Contrast to Noise Ratio (CNR) - were calculated for all CBCT images. In addition, trends in image quality metrics as a function of patients' average separation in AP and lateral directions were evaluated. Clinical CBCT and qCBCT images exhibited an average SSIM of 0.77±0.1 and 0.79±0.09, respectively, whereas average CT number accuracy of clinical CBCT and qCBCT was 39±43 HU and 18±17 HU, respectively. Additionally, CT nonuniformity in clinical CBCT scans and qCBCT was 35±41HU and 18±17 HU, respectively. Soft tissue CNR in both CBCT methods were comparable, 7±4 and 7±3. While image quality metrics in clinical CBCT and qCBCT images were more comparable in patients with lateral separations less than 35 cm, clinical CBCT image quality metrics degraded faster for separations larger than 35 cm. While structural similarity and CNR were comparable in clinical CBCT and qCBCT images, both CT number accuracy and nonuniformity were improved substantially with the qCBCT method. The use of a 2D anti-scatter grid with qCBCT did not have any adverse impact on soft tissue contrast and appearance. Lastly, patients with larger AP and lateral separations may benefit more from qCBCT's improved quality when compared to the current clinical CBCT. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of International Journal of Radiation Oncology, Biology, Physics is the property of Pergamon Press - An Imprint of Elsevier Science and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=179877218
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1016/j.ijrobp.2024.07.2311
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 1
        StartPage: S224
    Subjects:
      – SubjectFull: Cone beam computed tomography
        Type: general
      – SubjectFull: Image quality analysis
        Type: general
      – SubjectFull: Convex sets
        Type: general
      – SubjectFull: Rear-screen projection
        Type: general
      – SubjectFull: Diagnostic imaging
        Type: general
    Titles:
      – TitleFull: A Pilot Trial Testing a Novel 2D Antiscatter Grid and Quantitative CBCT Pipeline: An Interim Quantitative Image Quality Analysis.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Bayat, F.
      – PersonEntity:
          Name:
            NameFull: Pyakurel, U.
      – PersonEntity:
          Name:
            NameFull: Sabounchi, R.
      – PersonEntity:
          Name:
            NameFull: Bliley, R.C.
      – PersonEntity:
          Name:
            NameFull: Fisher, C.M.
      – PersonEntity:
          Name:
            NameFull: Rabinovitch, R.A.
      – PersonEntity:
          Name:
            NameFull: Binder, D.C.
      – PersonEntity:
          Name:
            NameFull: Kavanagh, B.D.
      – PersonEntity:
          Name:
            NameFull: Lanning, R.M.
      – PersonEntity:
          Name:
            NameFull: Robin, T.P.
      – PersonEntity:
          Name:
            NameFull: Altunbas, C.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 02
              M: 10
              Text: 2024 Supplement
              Type: published
              Y: 2024
          Identifiers:
            – Type: issn-print
              Value: 03603016
          Numbering:
            – Type: volume
              Value: 120
            – Type: issue
              Value: 2
          Titles:
            – TitleFull: International Journal of Radiation Oncology, Biology, Physics
              Type: main
ResultId 1