On the RF safety of titanium mesh head implants in 7 T MRI systems: an investigation.

Saved in:
Bibliographic Details
Title: On the RF safety of titanium mesh head implants in 7 T MRI systems: an investigation.
Authors: Mustafa, Mazin1 (AUTHOR) musta132@umn.edu, Zulkarnain, Nur Izzati Huda1 (AUTHOR), Sadeghi‐Tarakameh, Alireza1 (AUTHOR), Grant, Andrea1 (AUTHOR), Darrow, David2 (AUTHOR), Ozutemiz, Can3 (AUTHOR), Eryaman, Yigitcan1 (AUTHOR)
Source: Magnetic Resonance in Medicine. Jul2025, Vol. 94 Issue 1, p414-423. 10p.
Subjects: Facial muscles, Simulation methods & models, Human anatomical models, Skull surgery, Temperature measurements
Abstract: Purpose: Patients undergoing craniofacial surgery for skull reconstruction may have titanium mesh implants. The safety risks related to 7 T MRI with these patients are not well understood. This study investigates the RF heating of titanium mesh head implants at 7 T. Methods: A simulation model for a 7 T birdcage head coil was developed and validated against B1+$$ \left|{B}_1^{+}\right| $$, 1 g‐averaged specific absorption rate (SAR), and temperature measurements in the presence of a titanium mesh. Various mesh sizes and shapes at different angular positions were simulated to determine the worst‐case scenario in a spherical phantom in addition to the effect of rounding the mesh edges. Full‐wave electromagnetic and bioheat thermal simulations were conducted on anatomical human models. Results: Preliminary results indicate an increase in the local SAR near the meshes depending on the shape, size, and location. The maximum absolute temperatures in the head were, on average, around 38.2°C after 15 min of RF power exposure, corresponding to 3.2 W/kg whole‐head SAR without a titanium mesh implant. The maximum absolute temperatures did not significantly change after introducing the titanium mesh implants, and the highest temperature was 38.4°C, observed near the cerebellum and the facial muscles. The maximum local increase in temperature was observed at the vicinity of the mesh as 2.8°C. Finally, it was shown that large mesh implants can negatively impact B1+$$ \left|{B}_1^{+}\right| $$ field. Conclusions: Small rounded titanium mesh head implants can be generally safe for 7 T MRI scans under the standard guidelines. Avoiding sharp corners and edges may reduce the chances of RF safety risks. [ABSTRACT FROM AUTHOR]
Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 184713984
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: On the RF safety of titanium mesh head implants in 7 T MRI systems: an investigation.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Mustafa%2C+Mazin%22">Mustafa, Mazin</searchLink><relatesTo>1</relatesTo> (AUTHOR)<i> musta132@umn.edu</i><br /><searchLink fieldCode="AR" term="%22Zulkarnain%2C+Nur+Izzati+Huda%22">Zulkarnain, Nur Izzati Huda</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Sadeghi‐Tarakameh%2C+Alireza%22">Sadeghi‐Tarakameh, Alireza</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Grant%2C+Andrea%22">Grant, Andrea</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Darrow%2C+David%22">Darrow, David</searchLink><relatesTo>2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Ozutemiz%2C+Can%22">Ozutemiz, Can</searchLink><relatesTo>3</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Eryaman%2C+Yigitcan%22">Eryaman, Yigitcan</searchLink><relatesTo>1</relatesTo> (AUTHOR)
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+in+Medicine%22">Magnetic Resonance in Medicine</searchLink>. Jul2025, Vol. 94 Issue 1, p414-423. 10p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Facial+muscles%22">Facial muscles</searchLink><br /><searchLink fieldCode="DE" term="%22Simulation+methods+%26+models%22">Simulation methods & models</searchLink><br /><searchLink fieldCode="DE" term="%22Human+anatomical+models%22">Human anatomical models</searchLink><br /><searchLink fieldCode="DE" term="%22Skull+surgery%22">Skull surgery</searchLink><br /><searchLink fieldCode="DE" term="%22Temperature+measurements%22">Temperature measurements</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: Patients undergoing craniofacial surgery for skull reconstruction may have titanium mesh implants. The safety risks related to 7 T MRI with these patients are not well understood. This study investigates the RF heating of titanium mesh head implants at 7 T. Methods: A simulation model for a 7 T birdcage head coil was developed and validated against B1+$$ \left|{B}_1^{+}\right| $$, 1 g‐averaged specific absorption rate (SAR), and temperature measurements in the presence of a titanium mesh. Various mesh sizes and shapes at different angular positions were simulated to determine the worst‐case scenario in a spherical phantom in addition to the effect of rounding the mesh edges. Full‐wave electromagnetic and bioheat thermal simulations were conducted on anatomical human models. Results: Preliminary results indicate an increase in the local SAR near the meshes depending on the shape, size, and location. The maximum absolute temperatures in the head were, on average, around 38.2°C after 15 min of RF power exposure, corresponding to 3.2 W/kg whole‐head SAR without a titanium mesh implant. The maximum absolute temperatures did not significantly change after introducing the titanium mesh implants, and the highest temperature was 38.4°C, observed near the cerebellum and the facial muscles. The maximum local increase in temperature was observed at the vicinity of the mesh as 2.8°C. Finally, it was shown that large mesh implants can negatively impact B1+$$ \left|{B}_1^{+}\right| $$ field. Conclusions: Small rounded titanium mesh head implants can be generally safe for 7 T MRI scans under the standard guidelines. Avoiding sharp corners and edges may reduce the chances of RF safety risks. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=184713984
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1002/mrm.30477
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 10
        StartPage: 414
    Subjects:
      – SubjectFull: Facial muscles
        Type: general
      – SubjectFull: Simulation methods & models
        Type: general
      – SubjectFull: Human anatomical models
        Type: general
      – SubjectFull: Skull surgery
        Type: general
      – SubjectFull: Temperature measurements
        Type: general
    Titles:
      – TitleFull: On the RF safety of titanium mesh head implants in 7 T MRI systems: an investigation.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Mustafa, Mazin
      – PersonEntity:
          Name:
            NameFull: Zulkarnain, Nur Izzati Huda
      – PersonEntity:
          Name:
            NameFull: Sadeghi‐Tarakameh, Alireza
      – PersonEntity:
          Name:
            NameFull: Grant, Andrea
      – PersonEntity:
          Name:
            NameFull: Darrow, David
      – PersonEntity:
          Name:
            NameFull: Ozutemiz, Can
      – PersonEntity:
          Name:
            NameFull: Eryaman, Yigitcan
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 07
              Text: Jul2025
              Type: published
              Y: 2025
          Identifiers:
            – Type: issn-print
              Value: 07403194
          Numbering:
            – Type: volume
              Value: 94
            – Type: issue
              Value: 1
          Titles:
            – TitleFull: Magnetic Resonance in Medicine
              Type: main
ResultId 1