Complete Dosimetric Characterization of an In-House Manufactured SFRT Grid Collimator by 3D Printing with PLA-W Composite Filament.

Saved in:
Bibliographic Details
Title: Complete Dosimetric Characterization of an In-House Manufactured SFRT Grid Collimator by 3D Printing with PLA-W Composite Filament.
Authors: Velásquez, José1,2,3 (AUTHOR), Fuentealba, Melani1,2 (AUTHOR) melani.fuentealba@ufrontera.cl, Santibáñez, Mauricio1,2,3 (AUTHOR) mauricio.santibanez@ufrontera.cl
Source: Polymers (20734360). Jun2025, Vol. 17 Issue 11, p1496. 15p.
Subjects: Three-dimensional printing, Collimators, Reliability in engineering, Doping agents (Chemistry), Diodes
Abstract: This study presents a comprehensive dosimetric characterization and commissioning of a grid-type collimator manufactured via 3D printing using PLA-W composite filament, following an international protocol for small-field dosimetry. PLA doped with high concentrations of tungsten (>90% w/w) enables the fabrication of miniaturized collimators (<1 cm) with complex geometries, suitable for non-conventional radiotherapy applications. However, accurate assessment of spatial dose modulation is challenged by penumbra overlap between closely spaced beamlets, limiting the application of conventional instrumentation and protocols. To address this, absolute and relative dose distributions were evaluated for various radiation field configurations (number of beamlets) in both lateral and depth directions. Measurements were performed according to the IAEA TRS-483 protocol, using micro-ionization chambers and diode detectors. Additionally, long-term stability assessments were carried out to evaluate both the structural integrity and modulation performance of the printed grid over time. Point dose measurements using the same detectors were repeated after one year, and 2D surface dose distributions measured with EBT3 films were compared to SRS MapCHECK measurements two years later. The generated radiation field size of the central beamlet (FWHM) differed by less than 0.2% (15.8 mm) from the physical projection size (15.6 mm) and the lateral transmission due simultaneous beamlets resulted in FWHM variations of less than 3.8%, confirming manufacturing precision and collimator capability. Output factor measurements increased with the number of beamlets, from 0.75 for a single beamlet to 0.82 for the full beamlets configuration. No significant changes were observed in the depth of maximum dose across the different beamlets configurations (1.20 ± 0.20 cm). On the other hand, the long-term evaluations show no relevant changes in the FWHM or VPR, confirming the performance and reliability of the system. These results support the clinical feasibility and lasting performance stability of in-house manufactured grid collimators using PLA-W filaments and accessible 3D printing technology. [ABSTRACT FROM AUTHOR]
Copyright of Polymers (20734360) is the property of MDPI and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 185866802
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Complete Dosimetric Characterization of an In-House Manufactured SFRT Grid Collimator by 3D Printing with PLA-W Composite Filament.
– Name: Author
  Label: Authors
  Group: Au
  Data: &lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Vel&#225;squez%2C+Jos&#233;%22&quot;&gt;Vel&#225;squez, Jos&#233;&lt;/searchLink&gt;&lt;relatesTo&gt;1,2,3&lt;/relatesTo&gt; (AUTHOR)&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Fuentealba%2C+Melani%22&quot;&gt;Fuentealba, Melani&lt;/searchLink&gt;&lt;relatesTo&gt;1,2&lt;/relatesTo&gt; (AUTHOR)&lt;i&gt; melani.fuentealba@ufrontera.cl&lt;/i&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;AR&quot; term=&quot;%22Santib&#225;&#241;ez%2C+Mauricio%22&quot;&gt;Santib&#225;&#241;ez, Mauricio&lt;/searchLink&gt;&lt;relatesTo&gt;1,2,3&lt;/relatesTo&gt; (AUTHOR)&lt;i&gt; mauricio.santibanez@ufrontera.cl&lt;/i&gt;
– Name: TitleSource
  Label: Source
  Group: Src
  Data: &lt;searchLink fieldCode=&quot;JN&quot; term=&quot;%22Polymers+%2820734360%29%22&quot;&gt;Polymers (20734360)&lt;/searchLink&gt;. Jun2025, Vol. 17 Issue 11, p1496. 15p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: &lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Three-dimensional+printing%22&quot;&gt;Three-dimensional printing&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Collimators%22&quot;&gt;Collimators&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Reliability+in+engineering%22&quot;&gt;Reliability in engineering&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Doping+agents+%28Chemistry%29%22&quot;&gt;Doping agents (Chemistry)&lt;/searchLink&gt;&lt;br /&gt;&lt;searchLink fieldCode=&quot;DE&quot; term=&quot;%22Diodes%22&quot;&gt;Diodes&lt;/searchLink&gt;
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: This study presents a comprehensive dosimetric characterization and commissioning of a grid-type collimator manufactured via 3D printing using PLA-W composite filament, following an international protocol for small-field dosimetry. PLA doped with high concentrations of tungsten (&gt;90% w/w) enables the fabrication of miniaturized collimators (&lt;1 cm) with complex geometries, suitable for non-conventional radiotherapy applications. However, accurate assessment of spatial dose modulation is challenged by penumbra overlap between closely spaced beamlets, limiting the application of conventional instrumentation and protocols. To address this, absolute and relative dose distributions were evaluated for various radiation field configurations (number of beamlets) in both lateral and depth directions. Measurements were performed according to the IAEA TRS-483 protocol, using micro-ionization chambers and diode detectors. Additionally, long-term stability assessments were carried out to evaluate both the structural integrity and modulation performance of the printed grid over time. Point dose measurements using the same detectors were repeated after one year, and 2D surface dose distributions measured with EBT3 films were compared to SRS MapCHECK measurements two years later. The generated radiation field size of the central beamlet (FWHM) differed by less than 0.2% (15.8 mm) from the physical projection size (15.6 mm) and the lateral transmission due simultaneous beamlets resulted in FWHM variations of less than 3.8%, confirming manufacturing precision and collimator capability. Output factor measurements increased with the number of beamlets, from 0.75 for a single beamlet to 0.82 for the full beamlets configuration. No significant changes were observed in the depth of maximum dose across the different beamlets configurations (1.20 &#177; 0.20 cm). On the other hand, the long-term evaluations show no relevant changes in the FWHM or VPR, confirming the performance and reliability of the system. These results support the clinical feasibility and lasting performance stability of in-house manufactured grid collimators using PLA-W filaments and accessible 3D printing technology. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: &lt;i&gt;Copyright of Polymers (20734360) is the property of MDPI and its content may not be copied or emailed to multiple sites without the copyright holder&#39;s express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.&lt;/i&gt; (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=185866802
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.3390/polym17111496
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 15
        StartPage: 1496
    Subjects:
      – SubjectFull: Three-dimensional printing
        Type: general
      – SubjectFull: Collimators
        Type: general
      – SubjectFull: Reliability in engineering
        Type: general
      – SubjectFull: Doping agents (Chemistry)
        Type: general
      – SubjectFull: Diodes
        Type: general
    Titles:
      – TitleFull: Complete Dosimetric Characterization of an In-House Manufactured SFRT Grid Collimator by 3D Printing with PLA-W Composite Filament.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Velásquez, José
      – PersonEntity:
          Name:
            NameFull: Fuentealba, Melani
      – PersonEntity:
          Name:
            NameFull: Santibáñez, Mauricio
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 06
              Text: Jun2025
              Type: published
              Y: 2025
          Identifiers:
            – Type: issn-print
              Value: 20734360
          Numbering:
            – Type: volume
              Value: 17
            – Type: issue
              Value: 11
          Titles:
            – TitleFull: Polymers (20734360)
              Type: main
ResultId 1