Characterization of a shielded beam current transformer for ultra‐high dose rate (FLASH) electron beam monitoring and dose reporting.
Saved in:
| Title: | Characterization of a shielded beam current transformer for ultra‐high dose rate (FLASH) electron beam monitoring and dose reporting. |
|---|---|
| Authors: | Bernelin, Thibault1,2 (AUTHOR), Muir, Bryan3 (AUTHOR), Renaud, James3 (AUTHOR), Zerouali, Karim4 (AUTHOR), Guillet, Dominique4 (AUTHOR), Archambault, Louis1 (AUTHOR), Lalonde, Arthur2,4,5 (AUTHOR) arthur.lalonde@umontreal.ca |
| Source: | Medical Physics. Jul2025, Vol. 52 Issue 7, p1-11. 11p. |
| Subjects: | Electron beams, Current transformers (Instrument transformer), Linear accelerators, Real-time computing, Radiation dosimetry, Radiotherapy, Radiation doses |
| Abstract: | Background: Real‐time beam monitoring and accurate dose reporting is challenging in ultra‐high dose rate (UHDR) electron beams. Although beam current transformers (BCTs) can effectively track parameters such as pulse width (PW) and repetition frequency for UHDR electron beams, recent work has highlighted their sensitivity to electric fields induced by transient charge buildup in irradiated media under UHDR conditions. Purpose: This study evaluates the performance of a novel electrostatically shielded BCT for real‐time, high‐accuracy dose monitoring in UHDR electron beams. Methods: Irradiations were conducted using the Mobetron linear accelerator configured for UHDR electron beams with energies of 6 and 9 MeV. A shielded BCT was implemented to monitor beam delivery, with dose calibration established using alanine dosimeters in solid water phantoms. Dose stability was assessed over short (7‐day) and long (16‐week) periods. The BCT's response to variations in PW, pulse number, and pulse repetition frequency was also evaluated to determine its robustness across beam configurations. Results: The BCT showed high reproducibility and accuracy, with standard deviations of the difference between BCT‐predicted and alanine‐measured doses within 0.21% over short‐term measurements and 0.57% over long‐term measurements, even when subject to large (10%) machine output adjustments. When varying beam parameters, the BCT maintained accurate dose prediction within 1.0% and 1.4% of alanine measurements for 6 and 9 MeV, respectively, with high linearity (R2≥${\rm R}^2\ge$ 0.9997) across total doses. Conclusion: Shielded BCTs provide a stable and accurate solution for real‐time dose monitoring in FLASH radiotherapy, demonstrating robustness against output fluctuations and beam parameter variations. While further calibration standardization is required, this study supports the feasibility of using shielded BCTs for reliable UHDR dose monitoring, facilitating safe and precise implementation of FLASH radiotherapy in preclinical and clinical settings. [ABSTRACT FROM AUTHOR] |
| Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
|
Full text is not displayed to guests.
Login for full access.
|
|
| FullText | Links: – Type: pdflink Text: Availability: 1 |
|---|---|
| Header | DbId: egs DbLabel: Engineering Source An: 186809944 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Characterization of a shielded beam current transformer for ultra‐high dose rate (FLASH) electron beam monitoring and dose reporting. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Bernelin%2C+Thibault%22">Bernelin, Thibault</searchLink><relatesTo>1,2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Muir%2C+Bryan%22">Muir, Bryan</searchLink><relatesTo>3</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Renaud%2C+James%22">Renaud, James</searchLink><relatesTo>3</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Zerouali%2C+Karim%22">Zerouali, Karim</searchLink><relatesTo>4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Guillet%2C+Dominique%22">Guillet, Dominique</searchLink><relatesTo>4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Archambault%2C+Louis%22">Archambault, Louis</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Lalonde%2C+Arthur%22">Lalonde, Arthur</searchLink><relatesTo>2,4,5</relatesTo> (AUTHOR)<i> arthur.lalonde@umontreal.ca</i> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. Jul2025, Vol. 52 Issue 7, p1-11. 11p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22Electron+beams%22">Electron beams</searchLink><br /><searchLink fieldCode="DE" term="%22Current+transformers+%28Instrument+transformer%29%22">Current transformers (Instrument transformer)</searchLink><br /><searchLink fieldCode="DE" term="%22Linear+accelerators%22">Linear accelerators</searchLink><br /><searchLink fieldCode="DE" term="%22Real-time+computing%22">Real-time computing</searchLink><br /><searchLink fieldCode="DE" term="%22Radiation+dosimetry%22">Radiation dosimetry</searchLink><br /><searchLink fieldCode="DE" term="%22Radiotherapy%22">Radiotherapy</searchLink><br /><searchLink fieldCode="DE" term="%22Radiation+doses%22">Radiation doses</searchLink> – Name: Abstract Label: Abstract Group: Ab Data: Background: Real‐time beam monitoring and accurate dose reporting is challenging in ultra‐high dose rate (UHDR) electron beams. Although beam current transformers (BCTs) can effectively track parameters such as pulse width (PW) and repetition frequency for UHDR electron beams, recent work has highlighted their sensitivity to electric fields induced by transient charge buildup in irradiated media under UHDR conditions. Purpose: This study evaluates the performance of a novel electrostatically shielded BCT for real‐time, high‐accuracy dose monitoring in UHDR electron beams. Methods: Irradiations were conducted using the Mobetron linear accelerator configured for UHDR electron beams with energies of 6 and 9 MeV. A shielded BCT was implemented to monitor beam delivery, with dose calibration established using alanine dosimeters in solid water phantoms. Dose stability was assessed over short (7‐day) and long (16‐week) periods. The BCT's response to variations in PW, pulse number, and pulse repetition frequency was also evaluated to determine its robustness across beam configurations. Results: The BCT showed high reproducibility and accuracy, with standard deviations of the difference between BCT‐predicted and alanine‐measured doses within 0.21% over short‐term measurements and 0.57% over long‐term measurements, even when subject to large (10%) machine output adjustments. When varying beam parameters, the BCT maintained accurate dose prediction within 1.0% and 1.4% of alanine measurements for 6 and 9 MeV, respectively, with high linearity (R2≥${\rm R}^2\ge$ 0.9997) across total doses. Conclusion: Shielded BCTs provide a stable and accurate solution for real‐time dose monitoring in FLASH radiotherapy, demonstrating robustness against output fluctuations and beam parameter variations. While further calibration standardization is required, this study supports the feasibility of using shielded BCTs for reliable UHDR dose monitoring, facilitating safe and precise implementation of FLASH radiotherapy in preclinical and clinical settings. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=186809944 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.1002/mp.17927 Languages: – Code: eng Text: English PhysicalDescription: Pagination: PageCount: 11 StartPage: 1 Subjects: – SubjectFull: Electron beams Type: general – SubjectFull: Current transformers (Instrument transformer) Type: general – SubjectFull: Linear accelerators Type: general – SubjectFull: Real-time computing Type: general – SubjectFull: Radiation dosimetry Type: general – SubjectFull: Radiotherapy Type: general – SubjectFull: Radiation doses Type: general Titles: – TitleFull: Characterization of a shielded beam current transformer for ultra‐high dose rate (FLASH) electron beam monitoring and dose reporting. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Bernelin, Thibault – PersonEntity: Name: NameFull: Muir, Bryan – PersonEntity: Name: NameFull: Renaud, James – PersonEntity: Name: NameFull: Zerouali, Karim – PersonEntity: Name: NameFull: Guillet, Dominique – PersonEntity: Name: NameFull: Archambault, Louis – PersonEntity: Name: NameFull: Lalonde, Arthur IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 07 Text: Jul2025 Type: published Y: 2025 Identifiers: – Type: issn-print Value: 00942405 Numbering: – Type: volume Value: 52 – Type: issue Value: 7 Titles: – TitleFull: Medical Physics Type: main |
| ResultId | 1 |