Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T
Saved in:
| Title: | Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T |
|---|---|
| Authors: | Shellock, Frank G. frank.shellock@gte.net, Valencerina, Samuel |
| Source: | Magnetic Resonance Imaging (0730725X). Dec2005, Vol. 23 Issue 10, p1021-1025. 5p. |
| Subjects: | Magnetic resonance imaging, Medical imaging systems, Intellectual property, Magnetic resonance |
| Abstract: | Abstract: Specialized implants are used for transcatheter closure of septal defects, including atrial and ventricular septal defects, and patent foramen ovale. These metallic devices may pose a risk to patients undergoing magnetic resonance imaging (MRI) procedures especially if performed at 3 T. Therefore, this investigation evaluated MRI safety at 3 T for septal repair implants (CardioSEAL Septal Repair Implant and STARFlex Septal Repair Implant, NMT Medical, Boston, MA, USA) by characterizing magnetic field interactions, heating and artifacts. These implants exhibited minor magnetic field interactions; heating was not excessive (+0.5°C); and artifacts will only create a problem if the area of interest is in the same area as or near these devices. Thus, the findings indicated that it would be safe for a patient with these implants to undergo MRI at 3 T or lower. Importantly, because of the minor magnetic field interactions, MRI may be performed immediately after implantation. [Copyright &y& Elsevier] |
| Copyright of Magnetic Resonance Imaging (0730725X) is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
| FullText | Text: Availability: 0 |
|---|---|
| Header | DbId: egs DbLabel: Engineering Source An: 19305888 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Shellock%2C+Frank+G%2E%22">Shellock, Frank G.</searchLink><i> frank.shellock@gte.net</i><br /><searchLink fieldCode="AR" term="%22Valencerina%2C+Samuel%22">Valencerina, Samuel</searchLink> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+Imaging+%280730725X%29%22">Magnetic Resonance Imaging (0730725X)</searchLink>. Dec2005, Vol. 23 Issue 10, p1021-1025. 5p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22Magnetic+resonance+imaging%22">Magnetic resonance imaging</searchLink><br /><searchLink fieldCode="DE" term="%22Medical+imaging+systems%22">Medical imaging systems</searchLink><br /><searchLink fieldCode="DE" term="%22Intellectual+property%22">Intellectual property</searchLink><br /><searchLink fieldCode="DE" term="%22Magnetic+resonance%22">Magnetic resonance</searchLink> – Name: Abstract Label: Abstract Group: Ab Data: Abstract: Specialized implants are used for transcatheter closure of septal defects, including atrial and ventricular septal defects, and patent foramen ovale. These metallic devices may pose a risk to patients undergoing magnetic resonance imaging (MRI) procedures especially if performed at 3 T. Therefore, this investigation evaluated MRI safety at 3 T for septal repair implants (CardioSEAL Septal Repair Implant and STARFlex Septal Repair Implant, NMT Medical, Boston, MA, USA) by characterizing magnetic field interactions, heating and artifacts. These implants exhibited minor magnetic field interactions; heating was not excessive (+0.5°C); and artifacts will only create a problem if the area of interest is in the same area as or near these devices. Thus, the findings indicated that it would be safe for a patient with these implants to undergo MRI at 3 T or lower. Importantly, because of the minor magnetic field interactions, MRI may be performed immediately after implantation. [Copyright &y& Elsevier] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Magnetic Resonance Imaging (0730725X) is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=19305888 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.1016/j.mri.2005.10.010 Languages: – Code: eng Text: English PhysicalDescription: Pagination: PageCount: 5 StartPage: 1021 Subjects: – SubjectFull: Magnetic resonance imaging Type: general – SubjectFull: Medical imaging systems Type: general – SubjectFull: Intellectual property Type: general – SubjectFull: Magnetic resonance Type: general Titles: – TitleFull: Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Shellock, Frank G. – PersonEntity: Name: NameFull: Valencerina, Samuel IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 12 Text: Dec2005 Type: published Y: 2005 Identifiers: – Type: issn-print Value: 0730725X Numbering: – Type: volume Value: 23 – Type: issue Value: 10 Titles: – TitleFull: Magnetic Resonance Imaging (0730725X) Type: main |
| ResultId | 1 |