From Offline to Inline Without Pain: A Practical Framework for Translating Offline MR Reconstructions to Inline Deployment Using the Gadgetron Platform. |
| Authors: |
Ning, Zihan1 (AUTHOR) zihan.1.ning@kcl.ac.uk, Brackenier, Yannick1 (AUTHOR), McElroy, Sarah1,2 (AUTHOR), Silva, Sara Neves1 (AUTHOR), Cordero‐Grande, Lucilio3 (AUTHOR), Rot, Samuel4 (AUTHOR), Canas, Liane S.1 (AUTHOR), Thornley, Rebecca E.1,5 (AUTHOR), Leitão, David1 (AUTHOR), Poccecai, Davide1 (AUTHOR), Cantell, Andrew1 (AUTHOR), Kerosi, Rene1 (AUTHOR), Price, Anthony N.1,5 (AUTHOR), Cleary, Jon1,5 (AUTHOR), Tournier, Donald J.1 (AUTHOR), Hutter, Jana1,5 (AUTHOR), Bridgen, Philippa5,6 (AUTHOR), Di Cio, Pierluigi5,6 (AUTHOR), Cleri, Michela5,6 (AUTHOR), Granlund, Inka5,6 (AUTHOR) |
| Source: |
Magnetic Resonance in Medicine. Jul2026, Vol. 96 Issue 1, p448-459. 12p. |
| Subjects: |
Imaging system software, Siemens AG, Signal processing, Open source software, Magnetic resonance imaging, Scheduling |
| Abstract: |
Purpose: To develop and validate a practical, open‐source framework to overcome common issues in inline deployment of established offline MR reconstruction, including (1) scan disruption from lengthy reconstructions, (2) limited support for multi‐scan input reconstructions, (3) needs to adapt scripts for different raw‐data formats, and (4) limited guidance and experience in retaining scanner reconstructions and applying scanner‐based post‐processing to custom‐reconstructed images. Methods: The framework builds upon the Gadgetron platform as implemented on Siemens scanners and includes: (1) a general input converter to convert Gadgetron‐used ISMRMRD format raw into a Siemens format raw structure, facilitating reuse of code; (2) an asynchronous trigger‐and‐retrieve mechanism enabling long custom reconstructions without delaying scanner processes; (3) resource‐aware scheduling for parallel execution of reconstructions; (4) integrated file management to support multi‐scan inputs; and (5) preservation of scanner‐based reconstructions and post‐processing. The framework was validated on 2 Siemens scanners for SENSE, AlignedSENSE, and NUFFT reconstructions, and in a large‐cohort study. Results: Minimum code modification for inline deployment was demonstrated, and all reconstructions were successfully executed inline without disrupting scanner workflows. Images were retrieved automatically via retrieval scans or manually via retro‐reconstruction, with scanner‐based post‐processing applied to custom outputs. Multi‐sequence reconstructions were executed using GPU‐aware scheduling, confirming feasibility for large‐scale applications. In 480 examinations, inline reconstructions were retrieved in 99% of cases without disruptions. Conclusion: The framework lowers the technical barrier to inline deployment of offline reconstructions, enabling robust, scalable, and post‐processing‐compatible integration. It is openly available with documentation and demonstration cases to support reproducibility and community adoption. [ABSTRACT FROM AUTHOR] |
|
Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) |
| Database: |
Engineering Source |