Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study.

Saved in:
Bibliographic Details
Title: Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study.
Authors: Basto, Maria Luís1 (AUTHOR), Messias, Ana1,2 (AUTHOR), Neto, Maria Augusta2,3 (AUTHOR), Krauser, Jack T.3,4 (AUTHOR), Guerra, Fernando1,4 (AUTHOR), Amaro, Ana Martins2 (AUTHOR) ana.amaro@dem.uc.pt
Source: Polymers (20734360). Jun2026, Vol. 18 Issue 12, p1457. 15p.
Subjects: Polyether ether ketone, Digital image correlation, Dental implants, Prosthetics, Biomechanics, Dead loads (Mechanics), Polymethylmethacrylate, Polymers
Abstract: Implant-Supported Fixed Prostheses (ISFPs) have become a common option for the rehabilitation of fully edentulous arches and have traditionally incorporated metallic substructures with ceramic or acrylic veneering. The rapid expansion of CAD/CAM technologies has introduced not only a range of polymer-based materials as alternatives to conventional metallic frameworks but also the possibility of the fabrication of monolithic rehabilitations. However, the evidence regarding the mechanical behavior of monolithic polymer-based full-arch rehabilitations remains limited. This study aimed to evaluate and compare the mechanical performance of monolithic polymer-based complete prostheses under static loading using Digital Image Correlation (DIC). A total of 12 specimens (3 per group) simulating an FP3 maxillary full-arch ISFP supported by four implants were milled from four materials: poly(ether ether ketone) (G1-PEEK), poly(ether ketone ketone) (G2-PEKK), poly(methyl methacrylate) (G3-PMMA), and fiber-reinforced composite (G4-FRC). All specimens were subjected to static loading up to 200 N at the incisors region, corresponding to the anterior unsupported span, and at the occlusal surface of the molars, corresponding to the most distal portion of the cantilever, using a universal testing machine. Full-field vertical displacement and strain distributions (principal tensile, compressive, and von Mises) were acquired through a stereo DIC system and analyzed using a Linear Mixed-Effects Model with Tukey's HSD post hoc comparisons (α = 0.05). All prostheses withstood the applied load without macroscopic failure. G3-PMMA exhibited the highest vertical displacement, exceeding 1000 µm in the anterior span and 1500 µm in the cantilever region, along with the greatest strain concentrations, particularly at the interproximal embrasures distal to the terminal abutment. G1-PEEK provided the lowest displacement in the anterior span. G4-FRC presented displacements similar to G1-PEEK and G2-PEKK at the distal cantilever, but the lowest tensile strains and the most homogeneous strain dissipation in both loading at the anterior unsupported span and distal cantilever. This indicated that the biomechanical performance of full-arch ISFPs is highly influenced by the polymer used. PEEK, PEKK, and FRC appear as promising alternatives to PMMA for monolithic full-arch rehabilitations. [ABSTRACT FROM AUTHOR]
Copyright of Polymers (20734360) is the property of MDPI and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
FullText Links:
  – Type: pdflink
Text:
  Availability: 1
Header DbId: egs
DbLabel: Engineering Source
An: 194908326
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Basto%2C+Maria+Luís%22">Basto, Maria Luís</searchLink><relatesTo>1</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Messias%2C+Ana%22">Messias, Ana</searchLink><relatesTo>1,2</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Neto%2C+Maria+Augusta%22">Neto, Maria Augusta</searchLink><relatesTo>2,3</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Krauser%2C+Jack+T%2E%22">Krauser, Jack T.</searchLink><relatesTo>3,4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Guerra%2C+Fernando%22">Guerra, Fernando</searchLink><relatesTo>1,4</relatesTo> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Amaro%2C+Ana+Martins%22">Amaro, Ana Martins</searchLink><relatesTo>2</relatesTo> (AUTHOR)<i> ana.amaro@dem.uc.pt</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Polymers+%2820734360%29%22">Polymers (20734360)</searchLink>. Jun2026, Vol. 18 Issue 12, p1457. 15p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Polyether+ether+ketone%22">Polyether ether ketone</searchLink><br /><searchLink fieldCode="DE" term="%22Digital+image+correlation%22">Digital image correlation</searchLink><br /><searchLink fieldCode="DE" term="%22Dental+implants%22">Dental implants</searchLink><br /><searchLink fieldCode="DE" term="%22Prosthetics%22">Prosthetics</searchLink><br /><searchLink fieldCode="DE" term="%22Biomechanics%22">Biomechanics</searchLink><br /><searchLink fieldCode="DE" term="%22Dead+loads+%28Mechanics%29%22">Dead loads (Mechanics)</searchLink><br /><searchLink fieldCode="DE" term="%22Polymethylmethacrylate%22">Polymethylmethacrylate</searchLink><br /><searchLink fieldCode="DE" term="%22Polymers%22">Polymers</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Implant-Supported Fixed Prostheses (ISFPs) have become a common option for the rehabilitation of fully edentulous arches and have traditionally incorporated metallic substructures with ceramic or acrylic veneering. The rapid expansion of CAD/CAM technologies has introduced not only a range of polymer-based materials as alternatives to conventional metallic frameworks but also the possibility of the fabrication of monolithic rehabilitations. However, the evidence regarding the mechanical behavior of monolithic polymer-based full-arch rehabilitations remains limited. This study aimed to evaluate and compare the mechanical performance of monolithic polymer-based complete prostheses under static loading using Digital Image Correlation (DIC). A total of 12 specimens (3 per group) simulating an FP3 maxillary full-arch ISFP supported by four implants were milled from four materials: poly(ether ether ketone) (G1-PEEK), poly(ether ketone ketone) (G2-PEKK), poly(methyl methacrylate) (G3-PMMA), and fiber-reinforced composite (G4-FRC). All specimens were subjected to static loading up to 200 N at the incisors region, corresponding to the anterior unsupported span, and at the occlusal surface of the molars, corresponding to the most distal portion of the cantilever, using a universal testing machine. Full-field vertical displacement and strain distributions (principal tensile, compressive, and von Mises) were acquired through a stereo DIC system and analyzed using a Linear Mixed-Effects Model with Tukey's HSD post hoc comparisons (α = 0.05). All prostheses withstood the applied load without macroscopic failure. G3-PMMA exhibited the highest vertical displacement, exceeding 1000 µm in the anterior span and 1500 µm in the cantilever region, along with the greatest strain concentrations, particularly at the interproximal embrasures distal to the terminal abutment. G1-PEEK provided the lowest displacement in the anterior span. G4-FRC presented displacements similar to G1-PEEK and G2-PEKK at the distal cantilever, but the lowest tensile strains and the most homogeneous strain dissipation in both loading at the anterior unsupported span and distal cantilever. This indicated that the biomechanical performance of full-arch ISFPs is highly influenced by the polymer used. PEEK, PEKK, and FRC appear as promising alternatives to PMMA for monolithic full-arch rehabilitations. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Polymers (20734360) is the property of MDPI and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=194908326
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.3390/polym18121457
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 15
        StartPage: 1457
    Subjects:
      – SubjectFull: Polyether ether ketone
        Type: general
      – SubjectFull: Digital image correlation
        Type: general
      – SubjectFull: Dental implants
        Type: general
      – SubjectFull: Prosthetics
        Type: general
      – SubjectFull: Biomechanics
        Type: general
      – SubjectFull: Dead loads (Mechanics)
        Type: general
      – SubjectFull: Polymethylmethacrylate
        Type: general
      – SubjectFull: Polymers
        Type: general
    Titles:
      – TitleFull: Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Basto, Maria Luís
      – PersonEntity:
          Name:
            NameFull: Messias, Ana
      – PersonEntity:
          Name:
            NameFull: Neto, Maria Augusta
      – PersonEntity:
          Name:
            NameFull: Krauser, Jack T.
      – PersonEntity:
          Name:
            NameFull: Guerra, Fernando
      – PersonEntity:
          Name:
            NameFull: Amaro, Ana Martins
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 15
              M: 06
              Text: Jun2026
              Type: published
              Y: 2026
          Identifiers:
            – Type: issn-print
              Value: 20734360
          Numbering:
            – Type: volume
              Value: 18
            – Type: issue
              Value: 12
          Titles:
            – TitleFull: Polymers (20734360)
              Type: main
ResultId 1