Activity measurements of 18F and 90Y with commercial radionuclide calibrators for nuclear medicine in Switzerland

Saved in:
Bibliographic Details
Title: Activity measurements of 18F and 90Y with commercial radionuclide calibrators for nuclear medicine in Switzerland
Authors: Caffari, Yvan Yvan.Caffari@chuv.ch, Spring, Philippe1, Bailat, Claude1, Nedjadi, Youcef1, Bochud, François1
Source: Applied Radiation & Isotopes. Jul2010, Vol. 68 Issue 7/8, p1388-1391. 4p.
Subjects: Radioactivity measurements, Radioisotopes, Nuclear medicine, Radiopharmaceuticals, Gamma ray sources, Equilibrium
Geographic Terms: Switzerland
Abstract: Abstract: The activity of radiopharmaceuticals in nuclear medicine is measured before patient injection with radionuclide calibrators. In Switzerland, the general requirements for quality controls are defined in a federal ordinance and a directive of the Federal Office of Metrology (METAS) which require each instrument to be verified. A set of three gamma sources (Co-57, Cs-137 and Co-60) is used to verify the response of radionuclide calibrators in the gamma energy range of their use. A beta source, a mixture of 90Sr and 90Y in secular equilibrium, is used as well. Manufacturers are responsible for the calibration factors. The main goal of the study was to monitor the validity of the calibration factors by using two sources: a 90Sr/90Y source and a 18F source. The three types of commercial radionuclide calibrators tested do not have a calibration factor for the mixture but only for 90Y. Activity measurements of a 90Sr/90Y source with the 90Y calibration factor are performed in order to correct for the extra-contribution of 90Sr. The value of the correction factor was found to be 1.113 whereas Monte Carlo simulations of the radionuclide calibrators estimate the correction factor to be 1.117. Measurements with 18F sources in a specific geometry are also performed. Since this radionuclide is widely used in Swiss hospitals equipped with PET and PET-CT, the metrology of the 18F is very important. The 18F response normalized to the 137Cs response shows that the difference with a reference value does not exceed 3% for the three types of radionuclide calibrators. [Copyright &y& Elsevier]
Copyright of Applied Radiation & Isotopes is the property of Pergamon Press - An Imprint of Elsevier Science and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 51131083
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Activity measurements of <superscript>18</superscript>F and <superscript>90</superscript>Y with commercial radionuclide calibrators for nuclear medicine in Switzerland
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Caffari%2C+Yvan%22">Caffari, Yvan</searchLink><i> Yvan.Caffari@chuv.ch</i><br /><searchLink fieldCode="AR" term="%22Spring%2C+Philippe%22">Spring, Philippe</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Bailat%2C+Claude%22">Bailat, Claude</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Nedjadi%2C+Youcef%22">Nedjadi, Youcef</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Bochud%2C+François%22">Bochud, François</searchLink><relatesTo>1</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Applied+Radiation+%26+Isotopes%22">Applied Radiation & Isotopes</searchLink>. Jul2010, Vol. 68 Issue 7/8, p1388-1391. 4p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Radioactivity+measurements%22">Radioactivity measurements</searchLink><br /><searchLink fieldCode="DE" term="%22Radioisotopes%22">Radioisotopes</searchLink><br /><searchLink fieldCode="DE" term="%22Nuclear+medicine%22">Nuclear medicine</searchLink><br /><searchLink fieldCode="DE" term="%22Radiopharmaceuticals%22">Radiopharmaceuticals</searchLink><br /><searchLink fieldCode="DE" term="%22Gamma+ray+sources%22">Gamma ray sources</searchLink><br /><searchLink fieldCode="DE" term="%22Equilibrium%22">Equilibrium</searchLink>
– Name: SubjectGeographic
  Label: Geographic Terms
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Switzerland%22">Switzerland</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Abstract: The activity of radiopharmaceuticals in nuclear medicine is measured before patient injection with radionuclide calibrators. In Switzerland, the general requirements for quality controls are defined in a federal ordinance and a directive of the Federal Office of Metrology (METAS) which require each instrument to be verified. A set of three gamma sources (Co-57, Cs-137 and Co-60) is used to verify the response of radionuclide calibrators in the gamma energy range of their use. A beta source, a mixture of 90Sr and 90Y in secular equilibrium, is used as well. Manufacturers are responsible for the calibration factors. The main goal of the study was to monitor the validity of the calibration factors by using two sources: a 90Sr/90Y source and a 18F source. The three types of commercial radionuclide calibrators tested do not have a calibration factor for the mixture but only for 90Y. Activity measurements of a 90Sr/90Y source with the 90Y calibration factor are performed in order to correct for the extra-contribution of 90Sr. The value of the correction factor was found to be 1.113 whereas Monte Carlo simulations of the radionuclide calibrators estimate the correction factor to be 1.117. Measurements with 18F sources in a specific geometry are also performed. Since this radionuclide is widely used in Swiss hospitals equipped with PET and PET-CT, the metrology of the 18F is very important. The 18F response normalized to the 137Cs response shows that the difference with a reference value does not exceed 3% for the three types of radionuclide calibrators. [Copyright &y& Elsevier]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Applied Radiation & Isotopes is the property of Pergamon Press - An Imprint of Elsevier Science and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=51131083
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1016/j.apradiso.2009.11.015
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 4
        StartPage: 1388
    Subjects:
      – SubjectFull: Radioactivity measurements
        Type: general
      – SubjectFull: Radioisotopes
        Type: general
      – SubjectFull: Nuclear medicine
        Type: general
      – SubjectFull: Radiopharmaceuticals
        Type: general
      – SubjectFull: Gamma ray sources
        Type: general
      – SubjectFull: Equilibrium
        Type: general
      – SubjectFull: Switzerland
        Type: general
    Titles:
      – TitleFull: Activity measurements of 18F and 90Y with commercial radionuclide calibrators for nuclear medicine in Switzerland
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Caffari, Yvan
      – PersonEntity:
          Name:
            NameFull: Spring, Philippe
      – PersonEntity:
          Name:
            NameFull: Bailat, Claude
      – PersonEntity:
          Name:
            NameFull: Nedjadi, Youcef
      – PersonEntity:
          Name:
            NameFull: Bochud, François
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 07
              Text: Jul2010
              Type: published
              Y: 2010
          Identifiers:
            – Type: issn-print
              Value: 09698043
          Numbering:
            – Type: volume
              Value: 68
            – Type: issue
              Value: 7/8
          Titles:
            – TitleFull: Applied Radiation & Isotopes
              Type: main
ResultId 1