A novel respiratory tracking system for smart-gated PET acquisition.

Saved in:
Bibliographic Details
Title: A novel respiratory tracking system for smart-gated PET acquisition.
Authors: Nehmeh, S. A.1, Haj-Ali, A. A.2, Qing, C.1, Stearns, C.3, Kalaigian, H.1, Kohlmyer, S.3, Schoder, H.4, Ho, A. Y.5, Larson, S. M.4, Humm, J. L.1
Source: Medical Physics. Jan2011, Vol. 38 Issue 1, p531-538. 8p.
Subjects: Medical equipment, Breathing apparatus, Positron emission tomography, Acquisition of data, Medical imaging systems, Radiosurgery, Longitudinal method, Imaging phantoms
Abstract: Purpose: In this study, the authors validated a novel respiratory tracking device, the multidimensional respiratory tracking (MDRT) system, that was designed to assist in correcting for respiratory motion in PET/CT images. The authors also investigated a novel PET acquisition technique, smart gating (SG), that enables to acquire motion-free PET data prospectively, with minimum user interference and with no additional postprocessing of the PET data. Methods: MDRT uses visual tracking techniques to track simultaneously the two-dimensional (in the vertical plane) motion of multiple fiducial markers using a standard video camera. A threshold window is set at the breathing amplitude of interest using the MDRT GUI. A trigger is generated at a rate of 250 Hz as long as the breathing signal is within the threshold window. The triggers are fed into the PET scanner to initialize one single bin of a gated acquisition every 4 ms. No triggers are delivered as the breathing signal drifts outside the threshold window. Consequently, PET data are acquired only whenever the breathing signal is confined within the amplitude threshold window, thus resulting into a motion-free image set. The accuracy of MDRT in tracking the breathing signal was assessed (1) by comparing the period of an oscillating phantom, as measured by MDRT, to that measured with a photogate timer and (2) by comparing the MDRT output to that of the real-time position management (RPM) in ten patients. The SG PET/CT acquisition was validated in phantoms and in two stereotactic body radiosurgery (SBRS) lung DIBH-PET/CT patients. Results: MDRT was in agreement with the photogate timer in determining the period of motion to less than 2%. The percent errors between MDRT and RPM in the positions of the peaks and troughs of the ten patients' breathing signals were within 10%. In phantoms, SG technique enables to correct for motion-induced artifacts in the PET images and improve the accuracy of PET quantitation. For the SBRS application, in one patient, the patient's CT lesion was not detected in the corresponding clinical PET images, while it exhibited an SUV of 1.8 in the DIBH image set. In the second patient, DIBH-PET images showed an improved PET-to-CT spatial matching and a 52% increase in the lesion SUV. Conclusions: MDRT has been shown to be accurate in tracking breathing motion and assisted in implementing a smart-gating PET acquisition technique that allowed to acquire prospectively motion-free PET images. [ABSTRACT FROM AUTHOR]
Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Links:
  – Type: pdflink
Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 56911477
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: A novel respiratory tracking system for smart-gated PET acquisition.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Nehmeh%2C+S%2E+A%2E%22">Nehmeh, S. A.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Haj-Ali%2C+A%2E+A%2E%22">Haj-Ali, A. A.</searchLink><relatesTo>2</relatesTo><br /><searchLink fieldCode="AR" term="%22Qing%2C+C%2E%22">Qing, C.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Stearns%2C+C%2E%22">Stearns, C.</searchLink><relatesTo>3</relatesTo><br /><searchLink fieldCode="AR" term="%22Kalaigian%2C+H%2E%22">Kalaigian, H.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Kohlmyer%2C+S%2E%22">Kohlmyer, S.</searchLink><relatesTo>3</relatesTo><br /><searchLink fieldCode="AR" term="%22Schoder%2C+H%2E%22">Schoder, H.</searchLink><relatesTo>4</relatesTo><br /><searchLink fieldCode="AR" term="%22Ho%2C+A%2E+Y%2E%22">Ho, A. Y.</searchLink><relatesTo>5</relatesTo><br /><searchLink fieldCode="AR" term="%22Larson%2C+S%2E+M%2E%22">Larson, S. M.</searchLink><relatesTo>4</relatesTo><br /><searchLink fieldCode="AR" term="%22Humm%2C+J%2E+L%2E%22">Humm, J. L.</searchLink><relatesTo>1</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. Jan2011, Vol. 38 Issue 1, p531-538. 8p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Medical+equipment%22">Medical equipment</searchLink><br /><searchLink fieldCode="DE" term="%22Breathing+apparatus%22">Breathing apparatus</searchLink><br /><searchLink fieldCode="DE" term="%22Positron+emission+tomography%22">Positron emission tomography</searchLink><br /><searchLink fieldCode="DE" term="%22Acquisition+of+data%22">Acquisition of data</searchLink><br /><searchLink fieldCode="DE" term="%22Medical+imaging+systems%22">Medical imaging systems</searchLink><br /><searchLink fieldCode="DE" term="%22Radiosurgery%22">Radiosurgery</searchLink><br /><searchLink fieldCode="DE" term="%22Longitudinal+method%22">Longitudinal method</searchLink><br /><searchLink fieldCode="DE" term="%22Imaging+phantoms%22">Imaging phantoms</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: In this study, the authors validated a novel respiratory tracking device, the multidimensional respiratory tracking (MDRT) system, that was designed to assist in correcting for respiratory motion in PET/CT images. The authors also investigated a novel PET acquisition technique, smart gating (SG), that enables to acquire motion-free PET data prospectively, with minimum user interference and with no additional postprocessing of the PET data. Methods: MDRT uses visual tracking techniques to track simultaneously the two-dimensional (in the vertical plane) motion of multiple fiducial markers using a standard video camera. A threshold window is set at the breathing amplitude of interest using the MDRT GUI. A trigger is generated at a rate of 250 Hz as long as the breathing signal is within the threshold window. The triggers are fed into the PET scanner to initialize one single bin of a gated acquisition every 4 ms. No triggers are delivered as the breathing signal drifts outside the threshold window. Consequently, PET data are acquired only whenever the breathing signal is confined within the amplitude threshold window, thus resulting into a motion-free image set. The accuracy of MDRT in tracking the breathing signal was assessed (1) by comparing the period of an oscillating phantom, as measured by MDRT, to that measured with a photogate timer and (2) by comparing the MDRT output to that of the real-time position management (RPM) in ten patients. The SG PET/CT acquisition was validated in phantoms and in two stereotactic body radiosurgery (SBRS) lung DIBH-PET/CT patients. Results: MDRT was in agreement with the photogate timer in determining the period of motion to less than 2%. The percent errors between MDRT and RPM in the positions of the peaks and troughs of the ten patients' breathing signals were within 10%. In phantoms, SG technique enables to correct for motion-induced artifacts in the PET images and improve the accuracy of PET quantitation. For the SBRS application, in one patient, the patient's CT lesion was not detected in the corresponding clinical PET images, while it exhibited an SUV of 1.8 in the DIBH image set. In the second patient, DIBH-PET images showed an improved PET-to-CT spatial matching and a 52% increase in the lesion SUV. Conclusions: MDRT has been shown to be accurate in tracking breathing motion and assisted in implementing a smart-gating PET acquisition technique that allowed to acquire prospectively motion-free PET images. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=56911477
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1118/1.3523100
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 8
        StartPage: 531
    Subjects:
      – SubjectFull: Medical equipment
        Type: general
      – SubjectFull: Breathing apparatus
        Type: general
      – SubjectFull: Positron emission tomography
        Type: general
      – SubjectFull: Acquisition of data
        Type: general
      – SubjectFull: Medical imaging systems
        Type: general
      – SubjectFull: Radiosurgery
        Type: general
      – SubjectFull: Longitudinal method
        Type: general
      – SubjectFull: Imaging phantoms
        Type: general
    Titles:
      – TitleFull: A novel respiratory tracking system for smart-gated PET acquisition.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Nehmeh, S. A.
      – PersonEntity:
          Name:
            NameFull: Haj-Ali, A. A.
      – PersonEntity:
          Name:
            NameFull: Qing, C.
      – PersonEntity:
          Name:
            NameFull: Stearns, C.
      – PersonEntity:
          Name:
            NameFull: Kalaigian, H.
      – PersonEntity:
          Name:
            NameFull: Kohlmyer, S.
      – PersonEntity:
          Name:
            NameFull: Schoder, H.
      – PersonEntity:
          Name:
            NameFull: Ho, A. Y.
      – PersonEntity:
          Name:
            NameFull: Larson, S. M.
      – PersonEntity:
          Name:
            NameFull: Humm, J. L.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 01
              Text: Jan2011
              Type: published
              Y: 2011
          Identifiers:
            – Type: issn-print
              Value: 00942405
          Numbering:
            – Type: volume
              Value: 38
            – Type: issue
              Value: 1
          Titles:
            – TitleFull: Medical Physics
              Type: main
ResultId 1