Performance characteristics of a novel megavoltage cone-beam-computed tomography device.

Saved in:
Bibliographic Details
Title: Performance characteristics of a novel megavoltage cone-beam-computed tomography device.
Authors: Fast, M. F.1 m.fast@dkfz.de, Koenig, T.1, Oelfke, U.1, Nill, S.1 s.nill@dkfz.de
Source: Physics in Medicine & Biology. Feb2012, Vol. 57 Issue 3, pN15-N24. 10p.
Subjects: Tomography, Image quality analysis, Image-guided radiation therapy, Signal-to-noise ratio, Electron distribution, Imaging phantoms
Abstract: In this work, the image quality of a novel megavoltage cone-beam-computed tomography (CBCT) scanner is compared to three other image-guided radiation therapy devices by analysing images of different-sized quality assurance phantoms. The following devices are compared in terms of image uniformity, signal-to-noise ratio, contrast-to-noise ratio (CNR), electron density to HU conversion, presampling modulation transfer function (MTFpre) and combined spatial resolution and noise (Q-factor): (i) the Siemens Artiste kilovoltage (kV) (121 kV) CBCT device, (ii) the Artiste treatment beam line (TBL), 6 MV, (iii) the Tomotherapy (3.5 MV) fan-beam CT and (iv) Siemens' novel approach using a carbon target for a dedicated imaging beam line (IBL), 4.2 MV. Machine settings were selected to produce the same imaging dose for all devices. For a head phantom, IBL scans display CNR values 2.6±0.3 times higher than for the TBL at the same dose level (for a CT-number range of -200 to-60 HU). kV CBCT, on the other hand, displays CNR values 7.9±0.3 times higher than the IBL. There was no significant deviation in spatial resolution between IBL, TBL and Tomotherapy in terms of 50% and 10% MTFpre. For kV CBCT, the MTFpre was significantly higher than those for other devices. In our Q-factor analysis, the IBL (14.6) scores higher than the TBL (7.9) and Tomotherapy (9.7) due to its lower noise level. The linearity of electron density to HU conversion is demonstrated for different-sized phantoms. Employing the IBL instead of the TBL significantly reduces the imaging dose by up to a factor of 5 at a constant image quality level, providing an immediate benefit for the patient. [ABSTRACT FROM AUTHOR]
© 2025 IOP Publishing Ltd. All rights, including for text and data mining, AI training, and similar technologies, are reserved (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 73919524
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Performance characteristics of a novel megavoltage cone-beam-computed tomography device.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Fast%2C+M%2E+F%2E%22">Fast, M. F.</searchLink><relatesTo>1</relatesTo><i> m.fast@dkfz.de</i><br /><searchLink fieldCode="AR" term="%22Koenig%2C+T%2E%22">Koenig, T.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Oelfke%2C+U%2E%22">Oelfke, U.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Nill%2C+S%2E%22">Nill, S.</searchLink><relatesTo>1</relatesTo><i> s.nill@dkfz.de</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Physics+in+Medicine+%26+Biology%22">Physics in Medicine & Biology</searchLink>. Feb2012, Vol. 57 Issue 3, pN15-N24. 10p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Tomography%22">Tomography</searchLink><br /><searchLink fieldCode="DE" term="%22Image+quality+analysis%22">Image quality analysis</searchLink><br /><searchLink fieldCode="DE" term="%22Image-guided+radiation+therapy%22">Image-guided radiation therapy</searchLink><br /><searchLink fieldCode="DE" term="%22Signal-to-noise+ratio%22">Signal-to-noise ratio</searchLink><br /><searchLink fieldCode="DE" term="%22Electron+distribution%22">Electron distribution</searchLink><br /><searchLink fieldCode="DE" term="%22Imaging+phantoms%22">Imaging phantoms</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: In this work, the image quality of a novel megavoltage cone-beam-computed tomography (CBCT) scanner is compared to three other image-guided radiation therapy devices by analysing images of different-sized quality assurance phantoms. The following devices are compared in terms of image uniformity, signal-to-noise ratio, contrast-to-noise ratio (CNR), electron density to HU conversion, presampling modulation transfer function (MTFpre) and combined spatial resolution and noise (Q-factor): (i) the Siemens Artiste kilovoltage (kV) (121 kV) CBCT device, (ii) the Artiste treatment beam line (TBL), 6 MV, (iii) the Tomotherapy (3.5 MV) fan-beam CT and (iv) Siemens' novel approach using a carbon target for a dedicated imaging beam line (IBL), 4.2 MV. Machine settings were selected to produce the same imaging dose for all devices. For a head phantom, IBL scans display CNR values 2.6±0.3 times higher than for the TBL at the same dose level (for a CT-number range of -200 to-60 HU). kV CBCT, on the other hand, displays CNR values 7.9±0.3 times higher than the IBL. There was no significant deviation in spatial resolution between IBL, TBL and Tomotherapy in terms of 50% and 10% MTFpre. For kV CBCT, the MTFpre was significantly higher than those for other devices. In our Q-factor analysis, the IBL (14.6) scores higher than the TBL (7.9) and Tomotherapy (9.7) due to its lower noise level. The linearity of electron density to HU conversion is demonstrated for different-sized phantoms. Employing the IBL instead of the TBL significantly reduces the imaging dose by up to a factor of 5 at a constant image quality level, providing an immediate benefit for the patient. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>© 2025 IOP Publishing Ltd. All rights, including for text and data mining, AI training, and similar technologies, are reserved</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=73919524
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1088/0031-9155/57/3/N15
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 10
        StartPage: N15
    Subjects:
      – SubjectFull: Tomography
        Type: general
      – SubjectFull: Image quality analysis
        Type: general
      – SubjectFull: Image-guided radiation therapy
        Type: general
      – SubjectFull: Signal-to-noise ratio
        Type: general
      – SubjectFull: Electron distribution
        Type: general
      – SubjectFull: Imaging phantoms
        Type: general
    Titles:
      – TitleFull: Performance characteristics of a novel megavoltage cone-beam-computed tomography device.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Fast, M. F.
      – PersonEntity:
          Name:
            NameFull: Koenig, T.
      – PersonEntity:
          Name:
            NameFull: Oelfke, U.
      – PersonEntity:
          Name:
            NameFull: Nill, S.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 07
              M: 02
              Text: Feb2012
              Type: published
              Y: 2012
          Identifiers:
            – Type: issn-print
              Value: 00319155
          Numbering:
            – Type: volume
              Value: 57
            – Type: issue
              Value: 3
          Titles:
            – TitleFull: Physics in Medicine & Biology
              Type: main
ResultId 1