Development of a novel ArcCHECK™ insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities.

Saved in:
Bibliographic Details
Title: Development of a novel ArcCHECK™ insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities.
Authors: Fakir, H.1, Gaede, S.1, Mulligan, M.1, Chen, J. Z.1
Source: Medical Physics. Jul2012, Vol. 39 Issue 7,Part 1, p4203-4208. 6p.
Subjects: Radiotherapy, Quality assurance, Radiation doses, Stereotaxic techniques, Comparative studies, Medical radiology
Abstract: Purpose: To design a versatile, nonhomogeneous insert for the dose verification phantom ArcCHECK™ (Sun Nuclear Corp., FL) and to demonstrate its usefulness for the verification of dose distributions in inhomogeneous media. As an example, we demonstrate it can be used clinically for routine quality assurance of two volumetric modulated arc therapy (VMAT) systems for lung stereotactic body radiation therapy (SBRT): SmartArc® (Pinnacle3, Philips Radiation Oncology Systems, Fitchburg, WI) and RapidArc® (Eclipse™, Varian Medical Systems, Palo Alto, CA). Methods: The cylindrical detector array ArcCHECK™ has a retractable homogeneous acrylic insert. In this work, we designed and manufactured a customized heterogeneous insert with densities that simulate soft tissue, lung, bone, and air. The insert offers several possible heterogeneity configurations and multiple locations for point dose measurements. SmartArc® and RapidArc® plans for lung SBRT were generated and copied to ArcCHECK™ for each inhomogeneity configuration. Dose delivery was done on a Varian 2100 ix linac. The evaluation of dose distributions was based on gamma analysis of the diode measurements and point doses measurements at different positions near the inhomogeneities. Results: The insert was successfully manufactured and tested with different measurements of VMAT plans. Dose distributions measured with the homogeneous insert showed gamma passing rates similar to our clinical results (∼99%) for both treatment-planning systems. Using nonhomogeneous inserts decreased the passing rates by up to 3.6% in the examples studied. Overall, SmartArc® plans showed better gamma passing rates for nonhomogeneous measurements. The discrepancy between calculated and measured point doses was increased up to 6.5% for the nonhomogeneous insert depending on the inhomogeneity configuration and measurement location. SmartArc® and RapidArc® plans had similar plan quality but RapidArc® plans had significantly higher monitor units (up to 70%). Conclusions: A versatile, nonhomogeneous insert was developed for ArcCHECK™ for an easy and quick evaluation of dose calculations with nonhomogeneous media and for comparison of different treatment planning systems. The device was tested for SmartArc® and RapidArc® plans for lung SBRT, showing the uncertainties of dose calculations with inhomogeneities. The new insert combines the convenience of the ArcCHECK™ and the possibility of assessing dose distributions in inhomogeneous media. [ABSTRACT FROM AUTHOR]
Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Links:
  – Type: pdflink
Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 78029793
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Development of a novel ArcCHECK™ insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Fakir%2C+H%2E%22">Fakir, H.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Gaede%2C+S%2E%22">Gaede, S.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Mulligan%2C+M%2E%22">Mulligan, M.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Chen%2C+J%2E+Z%2E%22">Chen, J. Z.</searchLink><relatesTo>1</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Medical+Physics%22">Medical Physics</searchLink>. Jul2012, Vol. 39 Issue 7,Part 1, p4203-4208. 6p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Radiotherapy%22">Radiotherapy</searchLink><br /><searchLink fieldCode="DE" term="%22Quality+assurance%22">Quality assurance</searchLink><br /><searchLink fieldCode="DE" term="%22Radiation+doses%22">Radiation doses</searchLink><br /><searchLink fieldCode="DE" term="%22Stereotaxic+techniques%22">Stereotaxic techniques</searchLink><br /><searchLink fieldCode="DE" term="%22Comparative+studies%22">Comparative studies</searchLink><br /><searchLink fieldCode="DE" term="%22Medical+radiology%22">Medical radiology</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Purpose: To design a versatile, nonhomogeneous insert for the dose verification phantom ArcCHECK™ (Sun Nuclear Corp., FL) and to demonstrate its usefulness for the verification of dose distributions in inhomogeneous media. As an example, we demonstrate it can be used clinically for routine quality assurance of two volumetric modulated arc therapy (VMAT) systems for lung stereotactic body radiation therapy (SBRT): SmartArc® (Pinnacle3, Philips Radiation Oncology Systems, Fitchburg, WI) and RapidArc® (Eclipse™, Varian Medical Systems, Palo Alto, CA). Methods: The cylindrical detector array ArcCHECK™ has a retractable homogeneous acrylic insert. In this work, we designed and manufactured a customized heterogeneous insert with densities that simulate soft tissue, lung, bone, and air. The insert offers several possible heterogeneity configurations and multiple locations for point dose measurements. SmartArc® and RapidArc® plans for lung SBRT were generated and copied to ArcCHECK™ for each inhomogeneity configuration. Dose delivery was done on a Varian 2100 ix linac. The evaluation of dose distributions was based on gamma analysis of the diode measurements and point doses measurements at different positions near the inhomogeneities. Results: The insert was successfully manufactured and tested with different measurements of VMAT plans. Dose distributions measured with the homogeneous insert showed gamma passing rates similar to our clinical results (∼99%) for both treatment-planning systems. Using nonhomogeneous inserts decreased the passing rates by up to 3.6% in the examples studied. Overall, SmartArc® plans showed better gamma passing rates for nonhomogeneous measurements. The discrepancy between calculated and measured point doses was increased up to 6.5% for the nonhomogeneous insert depending on the inhomogeneity configuration and measurement location. SmartArc® and RapidArc® plans had similar plan quality but RapidArc® plans had significantly higher monitor units (up to 70%). Conclusions: A versatile, nonhomogeneous insert was developed for ArcCHECK™ for an easy and quick evaluation of dose calculations with nonhomogeneous media and for comparison of different treatment planning systems. The device was tested for SmartArc® and RapidArc® plans for lung SBRT, showing the uncertainties of dose calculations with inhomogeneities. The new insert combines the convenience of the ArcCHECK™ and the possibility of assessing dose distributions in inhomogeneous media. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=78029793
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1118/1.4728222
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 6
        StartPage: 4203
    Subjects:
      – SubjectFull: Radiotherapy
        Type: general
      – SubjectFull: Quality assurance
        Type: general
      – SubjectFull: Radiation doses
        Type: general
      – SubjectFull: Stereotaxic techniques
        Type: general
      – SubjectFull: Comparative studies
        Type: general
      – SubjectFull: Medical radiology
        Type: general
    Titles:
      – TitleFull: Development of a novel ArcCHECK™ insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Fakir, H.
      – PersonEntity:
          Name:
            NameFull: Gaede, S.
      – PersonEntity:
          Name:
            NameFull: Mulligan, M.
      – PersonEntity:
          Name:
            NameFull: Chen, J. Z.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 07
              Text: Jul2012
              Type: published
              Y: 2012
          Identifiers:
            – Type: issn-print
              Value: 00942405
          Numbering:
            – Type: volume
              Value: 39
            – Type: issue
              Value: 7,Part 1
          Titles:
            – TitleFull: Medical Physics
              Type: main
ResultId 1