| Title: |
Two-dimensional position-sensitive detectors for high resolution diffraction with neutrons and high energy synchrotron radiation |
| Authors: |
Marmotti, M.1, Burmester, J.1, Haese-Seiller, M.1, Kampmann, R. reinhard.kampmann@gkss.de |
| Source: |
Nuclear Instruments & Methods in Physics Research Section A. Jan2002, Vol. 477 Issue 1-3, p347. 6p. |
| Subjects: |
Neutron counters, Synchrotron radiation |
| Abstract: |
A two dimensional position-sensitive neutron detector has been developed at the Geesthacht Neutron Facility (GeNF) for the new diffractometer ARES which is dedicated to the analysis of residual stresses. The detector is a 3He and CF4 filled multiwire proportional counter with delay line readout and has a sensitive area of 300 mm×300 mm.The detector was tested at different 3He and CF4 partial pressures with a well collimated neutron beam. Excellent linearity and homogeneity of the detection probability combined with a spatial resolution of about 2 mm×2 mm were found. [Copyright &y& Elsevier] |
|
Copyright of Nuclear Instruments & Methods in Physics Research Section A is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) |