Evaluation of MRI artifacts at 3 Tesla for 38 commonly used cosmetics

Saved in:
Bibliographic Details
Title: Evaluation of MRI artifacts at 3 Tesla for 38 commonly used cosmetics
Authors: Escher, Kirin1, Shellock, Frank G.2 frank.shellock@mrisafety.com
Source: Magnetic Resonance Imaging (0730725X). Jun2013, Vol. 31 Issue 5, p778-782. 5p.
Subjects: Magnetic resonance imaging, Cosmetics, Imaging phantoms, Contrast media, Baldness, Eye care
Abstract: Abstract: Purpose: To evaluate MRI artifacts at 3-Tesla for 38 commonly used cosmetics. Materials and Methods: Thirty-eight cosmetics (16, nail polishes; 5, eyeliners; 3, mascaras; 10, eye shadows; 1, lip gloss; 1, body lotion; 1, body glitter, and 1, hair loss concealer) underwent evaluation for MRI artifacts at 3-Tesla. The cosmetics were applied a copper-sulfate-filled, phantom and initially assessed using a “screening” gradient echo (GRE) pulse sequence. Of the 38 different cosmetics, 14 (37%) exhibited artifacts. For these 14 cosmetics, additional characterization of artifacts was performed using a GRE pulse sequence. A qualitative scale was applied to characterize the artifact size. Results: Artifacts were observed, as follows: 2, nail polishes; 5, eyeliners; 3, mascaras; 3, eye shadows; 1, hair loss concealer. Artifact size ranged from small (eye shadow) to very large (hair loss concealer) and tended to be associated with the presence of iron oxide or other metal-based ingredient. Conclusions: Commonly used cosmetics caused artifacts that may create issues if the area of interest is the same as where the cosmetic was applied or if its presence was unknown, thus, potentially causing it to be construed as pathology. Therefore, these findings have important implications for patients referred for MRI examinations. [Copyright &y& Elsevier]
Copyright of Magnetic Resonance Imaging (0730725X) is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 87503500
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Evaluation of MRI artifacts at 3 Tesla for 38 commonly used cosmetics
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Escher%2C+Kirin%22">Escher, Kirin</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Shellock%2C+Frank+G%2E%22">Shellock, Frank G.</searchLink><relatesTo>2</relatesTo><i> frank.shellock@mrisafety.com</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+Imaging+%280730725X%29%22">Magnetic Resonance Imaging (0730725X)</searchLink>. Jun2013, Vol. 31 Issue 5, p778-782. 5p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Magnetic+resonance+imaging%22">Magnetic resonance imaging</searchLink><br /><searchLink fieldCode="DE" term="%22Cosmetics%22">Cosmetics</searchLink><br /><searchLink fieldCode="DE" term="%22Imaging+phantoms%22">Imaging phantoms</searchLink><br /><searchLink fieldCode="DE" term="%22Contrast+media%22">Contrast media</searchLink><br /><searchLink fieldCode="DE" term="%22Baldness%22">Baldness</searchLink><br /><searchLink fieldCode="DE" term="%22Eye+care%22">Eye care</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Abstract: Purpose: To evaluate MRI artifacts at 3-Tesla for 38 commonly used cosmetics. Materials and Methods: Thirty-eight cosmetics (16, nail polishes; 5, eyeliners; 3, mascaras; 10, eye shadows; 1, lip gloss; 1, body lotion; 1, body glitter, and 1, hair loss concealer) underwent evaluation for MRI artifacts at 3-Tesla. The cosmetics were applied a copper-sulfate-filled, phantom and initially assessed using a “screening” gradient echo (GRE) pulse sequence. Of the 38 different cosmetics, 14 (37%) exhibited artifacts. For these 14 cosmetics, additional characterization of artifacts was performed using a GRE pulse sequence. A qualitative scale was applied to characterize the artifact size. Results: Artifacts were observed, as follows: 2, nail polishes; 5, eyeliners; 3, mascaras; 3, eye shadows; 1, hair loss concealer. Artifact size ranged from small (eye shadow) to very large (hair loss concealer) and tended to be associated with the presence of iron oxide or other metal-based ingredient. Conclusions: Commonly used cosmetics caused artifacts that may create issues if the area of interest is the same as where the cosmetic was applied or if its presence was unknown, thus, potentially causing it to be construed as pathology. Therefore, these findings have important implications for patients referred for MRI examinations. [Copyright &y& Elsevier]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Magnetic Resonance Imaging (0730725X) is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=87503500
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1016/j.mri.2012.11.002
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 5
        StartPage: 778
    Subjects:
      – SubjectFull: Magnetic resonance imaging
        Type: general
      – SubjectFull: Cosmetics
        Type: general
      – SubjectFull: Imaging phantoms
        Type: general
      – SubjectFull: Contrast media
        Type: general
      – SubjectFull: Baldness
        Type: general
      – SubjectFull: Eye care
        Type: general
    Titles:
      – TitleFull: Evaluation of MRI artifacts at 3 Tesla for 38 commonly used cosmetics
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Escher, Kirin
      – PersonEntity:
          Name:
            NameFull: Shellock, Frank G.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 06
              Text: Jun2013
              Type: published
              Y: 2013
          Identifiers:
            – Type: issn-print
              Value: 0730725X
          Numbering:
            – Type: volume
              Value: 31
            – Type: issue
              Value: 5
          Titles:
            – TitleFull: Magnetic Resonance Imaging (0730725X)
              Type: main
ResultId 1