Comparison between absorbed dose to water standards established by water calorimetry at the LNE-LNHB and by application of international air-kerma based protocols for kilovoltage medium energy x-rays.

Saved in:
Bibliographic Details
Title: Comparison between absorbed dose to water standards established by water calorimetry at the LNE-LNHB and by application of international air-kerma based protocols for kilovoltage medium energy x-rays.
Authors: Perichon, N.1, Rapp, B.1 benjamin.rapp@cea.fr, Denoziere, M.1, Daures, J.1, Ostrowsky, A.1, Bordy, J.-M.1
Source: Physics in Medicine & Biology. 2013, Vol. 58 Issue 9, p2787-2806. 20p.
Subjects: Calorimetry, Palliative treatment, Ionization (Atomic physics), Coefficients (Statistics), X-rays
Abstract: Nowadays, the absorbed dose to water for kilovoltage x-ray beams is determined from standards in terms of air-kerma by application of international dosimetry protocols. New standards in terms of absorbed dose to water has just been established for these beams at the LNE-LNHB, using water calorimetry, at a depth of 2 cm in water in accordance with protocols. The aim of this study is to compare these new standards in terms of absorbed dose to water, to the dose values calculated from the application of four international protocols based on air-kerma standards (IAEA TRS-277, AAPM TG-61, IPEMB and NCS- 10). The acceleration potentials of the six beams studied are between 80 and 300 kV with half-value layers between 3.01 mm of aluminum and 3.40 mm of copper. A difference between the two methods smaller than 2.1% was reported. The standard uncertainty of water calorimetry being below 0.8%, and the one associated with the values from protocols being around 2.5%, the results are in good agreement. The calibration coefficients of some ionization chambers in terms of absorbed dose to water, established by application of calorimetry and air-kerma based dosimetry protocols, were also compared. The best agreement with the calibration coefficients established by water calorimetry was found for those established with the AAPM TG-61 protocol. [ABSTRACT FROM AUTHOR]
© 2025 IOP Publishing Ltd. All rights, including for text and data mining, AI training, and similar technologies, are reserved (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 90049793
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Comparison between absorbed dose to water standards established by water calorimetry at the LNE-LNHB and by application of international air-kerma based protocols for kilovoltage medium energy x-rays.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Perichon%2C+N%2E%22">Perichon, N.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Rapp%2C+B%2E%22">Rapp, B.</searchLink><relatesTo>1</relatesTo><i> benjamin.rapp@cea.fr</i><br /><searchLink fieldCode="AR" term="%22Denoziere%2C+M%2E%22">Denoziere, M.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Daures%2C+J%2E%22">Daures, J.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Ostrowsky%2C+A%2E%22">Ostrowsky, A.</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Bordy%2C+J%2E-M%2E%22">Bordy, J.-M.</searchLink><relatesTo>1</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Physics+in+Medicine+%26+Biology%22">Physics in Medicine & Biology</searchLink>. 2013, Vol. 58 Issue 9, p2787-2806. 20p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Calorimetry%22">Calorimetry</searchLink><br /><searchLink fieldCode="DE" term="%22Palliative+treatment%22">Palliative treatment</searchLink><br /><searchLink fieldCode="DE" term="%22Ionization+%28Atomic+physics%29%22">Ionization (Atomic physics)</searchLink><br /><searchLink fieldCode="DE" term="%22Coefficients+%28Statistics%29%22">Coefficients (Statistics)</searchLink><br /><searchLink fieldCode="DE" term="%22X-rays%22">X-rays</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Nowadays, the absorbed dose to water for kilovoltage x-ray beams is determined from standards in terms of air-kerma by application of international dosimetry protocols. New standards in terms of absorbed dose to water has just been established for these beams at the LNE-LNHB, using water calorimetry, at a depth of 2 cm in water in accordance with protocols. The aim of this study is to compare these new standards in terms of absorbed dose to water, to the dose values calculated from the application of four international protocols based on air-kerma standards (IAEA TRS-277, AAPM TG-61, IPEMB and NCS- 10). The acceleration potentials of the six beams studied are between 80 and 300 kV with half-value layers between 3.01 mm of aluminum and 3.40 mm of copper. A difference between the two methods smaller than 2.1% was reported. The standard uncertainty of water calorimetry being below 0.8%, and the one associated with the values from protocols being around 2.5%, the results are in good agreement. The calibration coefficients of some ionization chambers in terms of absorbed dose to water, established by application of calorimetry and air-kerma based dosimetry protocols, were also compared. The best agreement with the calibration coefficients established by water calorimetry was found for those established with the AAPM TG-61 protocol. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>© 2025 IOP Publishing Ltd. All rights, including for text and data mining, AI training, and similar technologies, are reserved</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=90049793
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1088/0031-9155/58/9/2787
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 20
        StartPage: 2787
    Subjects:
      – SubjectFull: Calorimetry
        Type: general
      – SubjectFull: Palliative treatment
        Type: general
      – SubjectFull: Ionization (Atomic physics)
        Type: general
      – SubjectFull: Coefficients (Statistics)
        Type: general
      – SubjectFull: X-rays
        Type: general
    Titles:
      – TitleFull: Comparison between absorbed dose to water standards established by water calorimetry at the LNE-LNHB and by application of international air-kerma based protocols for kilovoltage medium energy x-rays.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Perichon, N.
      – PersonEntity:
          Name:
            NameFull: Rapp, B.
      – PersonEntity:
          Name:
            NameFull: Denoziere, M.
      – PersonEntity:
          Name:
            NameFull: Daures, J.
      – PersonEntity:
          Name:
            NameFull: Ostrowsky, A.
      – PersonEntity:
          Name:
            NameFull: Bordy, J.-M.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 07
              M: 05
              Text: 2013
              Type: published
              Y: 2013
          Identifiers: