Evaluation of MRI issues for an access port with a Radiofrequency Identification (RFID) tag.

Saved in:
Bibliographic Details
Title: Evaluation of MRI issues for an access port with a Radiofrequency Identification (RFID) tag.
Authors: Titterington, Blake1 batitterington@gmail.com, Shellock, Frank G.2 frank.shellock@mrisafety.com
Source: Magnetic Resonance Imaging (0730725X). Oct2013, Vol. 31 Issue 8, p1439-1444. 6p.
Subjects: Magnetic resonance imaging, Radio frequency identification systems, Tags (Metadata), Magnetic fields, Temperature effect, Medical protocols
Abstract: Objective: A medical implant that contains metal, such as an RFID tag, must undergo proper MRI testing to ensure patient safety and to determine that the function of the RFID tag is not compromised by exposure to MRI conditions. Therefore, the objective of this investigation was to assess MRI issues for a new access port that incorporates an RFID tag. Materials and Methods: Samples of the access port with an RFID tag (Medcomp Power Injectable Port with CertainID; Medcomp, Harleysville, PA) were evaluated using standard protocols to assess magnetic field interactions (translational attraction and torque; 3-T), MRI-related heating (3-T), artifacts (3-T), and functional changes associated with different MRI conditions (nine samples, exposed to different MRI conditions at 1.5-T and 3-T). Results: Magnetic field interactions were not substantial and will pose no hazards to patients. MRI-related heating was minimal (highest temperature change, 1.7°C; background temperature rise, 1.6°C). Artifacts were moderate in size in relation to the device. Exposures to MRI conditions at 1.5-T and 3-T did not alter or damage the functional aspects of the RFID tag. Conclusions: Based on the findings of the test, this new access port with an RFID tag is acceptable (or, MR conditional, using current MRI labeling terminology) for patients undergoing MRI examinations at 1.5-T/64-MHz and 3-T/128-MHz. [ABSTRACT FROM AUTHOR]
Copyright of Magnetic Resonance Imaging (0730725X) is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 90204208
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Evaluation of MRI issues for an access port with a Radiofrequency Identification (RFID) tag.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Titterington%2C+Blake%22">Titterington, Blake</searchLink><relatesTo>1</relatesTo><i> batitterington@gmail.com</i><br /><searchLink fieldCode="AR" term="%22Shellock%2C+Frank+G%2E%22">Shellock, Frank G.</searchLink><relatesTo>2</relatesTo><i> frank.shellock@mrisafety.com</i>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Magnetic+Resonance+Imaging+%280730725X%29%22">Magnetic Resonance Imaging (0730725X)</searchLink>. Oct2013, Vol. 31 Issue 8, p1439-1444. 6p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Magnetic+resonance+imaging%22">Magnetic resonance imaging</searchLink><br /><searchLink fieldCode="DE" term="%22Radio+frequency+identification+systems%22">Radio frequency identification systems</searchLink><br /><searchLink fieldCode="DE" term="%22Tags+%28Metadata%29%22">Tags (Metadata)</searchLink><br /><searchLink fieldCode="DE" term="%22Magnetic+fields%22">Magnetic fields</searchLink><br /><searchLink fieldCode="DE" term="%22Temperature+effect%22">Temperature effect</searchLink><br /><searchLink fieldCode="DE" term="%22Medical+protocols%22">Medical protocols</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: Objective: A medical implant that contains metal, such as an RFID tag, must undergo proper MRI testing to ensure patient safety and to determine that the function of the RFID tag is not compromised by exposure to MRI conditions. Therefore, the objective of this investigation was to assess MRI issues for a new access port that incorporates an RFID tag. Materials and Methods: Samples of the access port with an RFID tag (Medcomp Power Injectable Port with CertainID; Medcomp, Harleysville, PA) were evaluated using standard protocols to assess magnetic field interactions (translational attraction and torque; 3-T), MRI-related heating (3-T), artifacts (3-T), and functional changes associated with different MRI conditions (nine samples, exposed to different MRI conditions at 1.5-T and 3-T). Results: Magnetic field interactions were not substantial and will pose no hazards to patients. MRI-related heating was minimal (highest temperature change, 1.7°C; background temperature rise, 1.6°C). Artifacts were moderate in size in relation to the device. Exposures to MRI conditions at 1.5-T and 3-T did not alter or damage the functional aspects of the RFID tag. Conclusions: Based on the findings of the test, this new access port with an RFID tag is acceptable (or, MR conditional, using current MRI labeling terminology) for patients undergoing MRI examinations at 1.5-T/64-MHz and 3-T/128-MHz. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Magnetic Resonance Imaging (0730725X) is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=90204208
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1016/j.mri.2013.04.005
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 6
        StartPage: 1439
    Subjects:
      – SubjectFull: Magnetic resonance imaging
        Type: general
      – SubjectFull: Radio frequency identification systems
        Type: general
      – SubjectFull: Tags (Metadata)
        Type: general
      – SubjectFull: Magnetic fields
        Type: general
      – SubjectFull: Temperature effect
        Type: general
      – SubjectFull: Medical protocols
        Type: general
    Titles:
      – TitleFull: Evaluation of MRI issues for an access port with a Radiofrequency Identification (RFID) tag.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Titterington, Blake
      – PersonEntity:
          Name:
            NameFull: Shellock, Frank G.
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 10
              Text: Oct2013
              Type: published
              Y: 2013
          Identifiers:
            – Type: issn-print
              Value: 0730725X
          Numbering:
            – Type: volume
              Value: 31
            – Type: issue
              Value: 8
          Titles:
            – TitleFull: Magnetic Resonance Imaging (0730725X)
              Type: main
ResultId 1