PROPUESTA DE UN PROCEDIMIENTO DE MEDICIÓN PARA LA CALIBRACIÓN DE LAS CAMPANAS GASOMÉTRICAS MODELOS LJQ-20 Y LJQ-100.
Saved in:
| Title: | PROPUESTA DE UN PROCEDIMIENTO DE MEDICIÓN PARA LA CALIBRACIÓN DE LAS CAMPANAS GASOMÉTRICAS MODELOS LJQ-20 Y LJQ-100. |
|---|---|
| Alternate Title: | PROPOSAL OF A MEASUREMENT PROCEDURE FOR THE CALIBRATION OF BELL PROVERS MODELS LJQ-20 AND LJQ-100. |
| Authors: | BUZÓN-GONZÁLEZ, F. J.1 franb@inimet.cu, HERNÁNDEZ-LEONARD, A. R.2 alehl@inimet.cu |
| Source: | Revista Cubana de Física. jul2023, Vol. 40 Issue 1, pE15-E19. 5p. |
| Subjects: | GAS analysis, GAS measurement, TEMPERATURE sensors, GAS-meters, CALIBRATION, MATHEMATICAL models, HUMIDITY, IMPORTS |
| Geographic Terms: | CUBA |
| Abstract (English): | The article proposes a measurement procedure for the calibration of the gasometric bells models LJQ-20 and LJQ-100 manufactured by Hangzhou TNMA MEasure Technology Co., Ltd. These bells are used in Cuba for gas meter verification and currently this service is imported. Different calibration methods are presented, but the taping method is recommended as the most suitable due to its low cost and high quality. A mathematical model is provided to calculate the volume of the bell and periodic calibration of pressure, humidity, and temperature sensors is recommended. Uncertainty components are mentioned and the need for the bell to have a cross section close to a true circle is highlighted. It is proposed to implement this procedure in Cuba to achieve economic savings. [Extracted from the article] |
| Abstract (Spanish): | El artículo propone un procedimiento de medición para la calibración de las campanas gasométricas modelos LJQ-20 y LJQ-100 fabricadas por Hangzhou TNMA MEasure Technology Co., Ltd. Estas campanas se utilizan en Cuba para la verificación de contadores de gas y actualmente se importa este servicio. Se presentan diferentes métodos de calibración, pero se recomienda el método de encintado como el más adecuado debido a su bajo costo y alta calidad. Se proporciona un modelo matemático para calcular el volumen de la campana y se recomienda la calibración periódica de los sensores de presión, humedad y temperatura. Se mencionan las componentes de incertidumbre y se destaca la necesidad de que la campana tenga una sección transversal cercana a un círculo verdadero. Se propone implementar este procedimiento en Cuba para lograr un ahorro económico. [Extracted from the article] |
| Copyright of Revista Cubana de Física is the property of Universidad de La Habana and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | MedicLatina |
| FullText | Links: – Type: pdflink Text: Availability: 0 |
|---|---|
| Header | DbId: lth DbLabel: MedicLatina An: 169902557 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: PROPUESTA DE UN PROCEDIMIENTO DE MEDICIÓN PARA LA CALIBRACIÓN DE LAS CAMPANAS GASOMÉTRICAS MODELOS LJQ-20 Y LJQ-100. – Name: TitleAlt Label: Alternate Title Group: TiAlt Data: PROPOSAL OF A MEASUREMENT PROCEDURE FOR THE CALIBRATION OF BELL PROVERS MODELS LJQ-20 AND LJQ-100. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22BUZÓN-GONZÁLEZ%2C+F%2E+J%2E%22">BUZÓN-GONZÁLEZ, F. J.</searchLink><relatesTo>1</relatesTo><i> franb@inimet.cu</i><br /><searchLink fieldCode="AR" term="%22HERNÁNDEZ-LEONARD%2C+A%2E+R%2E%22">HERNÁNDEZ-LEONARD, A. R.</searchLink><relatesTo>2</relatesTo><i> alehl@inimet.cu</i> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Revista+Cubana+de+Física%22">Revista Cubana de Física</searchLink>. jul2023, Vol. 40 Issue 1, pE15-E19. 5p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22GAS+analysis%22">GAS analysis</searchLink><br /><searchLink fieldCode="DE" term="%22GAS+measurement%22">GAS measurement</searchLink><br /><searchLink fieldCode="DE" term="%22TEMPERATURE+sensors%22">TEMPERATURE sensors</searchLink><br /><searchLink fieldCode="DE" term="%22GAS-meters%22">GAS-meters</searchLink><br /><searchLink fieldCode="DE" term="%22CALIBRATION%22">CALIBRATION</searchLink><br /><searchLink fieldCode="DE" term="%22MATHEMATICAL+models%22">MATHEMATICAL models</searchLink><br /><searchLink fieldCode="DE" term="%22HUMIDITY%22">HUMIDITY</searchLink><br /><searchLink fieldCode="DE" term="%22IMPORTS%22">IMPORTS</searchLink> – Name: SubjectGeographic Label: Geographic Terms Group: Su Data: <searchLink fieldCode="DE" term="%22CUBA%22">CUBA</searchLink> – Name: Abstract Label: Abstract (English) Group: Ab Data: The article proposes a measurement procedure for the calibration of the gasometric bells models LJQ-20 and LJQ-100 manufactured by Hangzhou TNMA MEasure Technology Co., Ltd. These bells are used in Cuba for gas meter verification and currently this service is imported. Different calibration methods are presented, but the taping method is recommended as the most suitable due to its low cost and high quality. A mathematical model is provided to calculate the volume of the bell and periodic calibration of pressure, humidity, and temperature sensors is recommended. Uncertainty components are mentioned and the need for the bell to have a cross section close to a true circle is highlighted. It is proposed to implement this procedure in Cuba to achieve economic savings. [Extracted from the article] – Name: Abstract Label: Abstract (Spanish) Group: Ab Data: El artículo propone un procedimiento de medición para la calibración de las campanas gasométricas modelos LJQ-20 y LJQ-100 fabricadas por Hangzhou TNMA MEasure Technology Co., Ltd. Estas campanas se utilizan en Cuba para la verificación de contadores de gas y actualmente se importa este servicio. Se presentan diferentes métodos de calibración, pero se recomienda el método de encintado como el más adecuado debido a su bajo costo y alta calidad. Se proporciona un modelo matemático para calcular el volumen de la campana y se recomienda la calibración periódica de los sensores de presión, humedad y temperatura. Se mencionan las componentes de incertidumbre y se destaca la necesidad de que la campana tenga una sección transversal cercana a un círculo verdadero. Se propone implementar este procedimiento en Cuba para lograr un ahorro económico. [Extracted from the article] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Revista Cubana de Física is the property of Universidad de La Habana and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=lth&AN=169902557 |
| RecordInfo | BibRecord: BibEntity: Languages: – Code: spa Text: Spanish PhysicalDescription: Pagination: PageCount: 5 StartPage: E15 Subjects: – SubjectFull: GAS analysis Type: general – SubjectFull: GAS measurement Type: general – SubjectFull: TEMPERATURE sensors Type: general – SubjectFull: GAS-meters Type: general – SubjectFull: CALIBRATION Type: general – SubjectFull: MATHEMATICAL models Type: general – SubjectFull: HUMIDITY Type: general – SubjectFull: IMPORTS Type: general – SubjectFull: CUBA Type: general Titles: – TitleFull: PROPUESTA DE UN PROCEDIMIENTO DE MEDICIÓN PARA LA CALIBRACIÓN DE LAS CAMPANAS GASOMÉTRICAS MODELOS LJQ-20 Y LJQ-100. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: BUZÓN-GONZÁLEZ, F. J. – PersonEntity: Name: NameFull: HERNÁNDEZ-LEONARD, A. R. IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 07 Text: jul2023 Type: published Y: 2023 Identifiers: – Type: issn-print Value: 02539268 Numbering: – Type: volume Value: 40 – Type: issue Value: 1 Titles: – TitleFull: Revista Cubana de Física Type: main |
| ResultId | 1 |