Un análisis del efecto del volumen del cuerpo de la jeringa sobre la falla mecánica durante la inyección.
Saved in:
| Title: | Un análisis del efecto del volumen del cuerpo de la jeringa sobre la falla mecánica durante la inyección. |
|---|---|
| Alternate Title: | An analysis of the effect of syringe barrel volume on mechanical failure during injection. |
| Authors: | Krisdiyanto1 krisdiyanto@umy.ac.id, Acknowledgment1 |
| Source: | Gaceta Médica Boliviana. ene-jun2025, Vol. 48 Issue 1, p77-82. 6p. |
| Subjects: | SYRINGES, MECHANICAL failures, STRUCTURAL failures, SAFETY, INJECTIONS, MEDICAL equipment, YIELD stress |
| Abstract (English): | Background: syringes are critical medical devices whose mechanical integrity is essential to ensure safety during injections. The cylinder volume can influence their mechanical performance. Objective: this study investigates how cylinder volume affects the mechanical integrity of syringes by analyzing von Mises stress, buckling risk, and overall safety during injection. Methods: autodesk Inventor 2021 and Ansys 2019 R2 were used to analyze syringes with volumes of 1 ml, 3 ml, 5 ml, and 10 ml. Von Mises stress and buckling susceptibility were evaluated during simulated injections. Results: smaller syringes showed higher stress levels: 14.030 MPa for the 1 ml syringe, 10.532 MPa for the 3 ml, and 7.0150 MPa for the 5 ml and 10 ml. The 5 ml syringe was the most susceptible to buckling, with a critical pressure of -0.1757 MPa. Despite this, all designs are safe under normal operating conditions. Conclusions: cylinder volume affects mechanical performance, but with proper design, even smaller syringes can remain safe, enhancing the reliability of medical treatments. [ABSTRACT FROM AUTHOR] |
| Abstract (Spanish): | Antecedentes: las jeringas son dispositivos médicos críticos cuya integridad mecánica es esencial para garantizar su seguridad en las inyecciones. El volumen del cilindro puede influir en su rendimiento mecánico. Objetivo: este estudio investiga cómo el volumen del cilindro afecta la integridad mecánica de las jeringas, analizando el esfuerzo de von Mises, el riesgo de pandeo y la seguridad general durante la inyección. Métodos: se utilizaron Autodesk Inventor 2021 y Ansys 2019 R2 para analizar jeringas de 1 ml, 3 ml, 5 ml y 10 ml. Se evaluaron los esfuerzos de von Mises y la susceptibilidad al pandeo durante inyecciones simuladas. Resultados: las jeringas más pequeñas presentaron mayores esfuerzos: 14,030 MPa para la de 1 ml, 10,532 MPa para la de 3 ml y 7,0150 MPa para las de 5 ml y 10 ml. La jeringa de 5 ml fue la más susceptible al pandeo, con una presión crítica de -0,1757 MPa. A pesar de ello, todas son seguras bajo condiciones normales. Conclusiones: el volumen del cilindro afecta el rendimiento mecánico, pero con un buen diseño, incluso las jeringas pequeñas pueden ser seguras, mejorando la fiabilidad de los tratamientos médicos. [ABSTRACT FROM AUTHOR] |
| Copyright of Gaceta Médica Boliviana is the property of Universidad Mayor de San Simon, Facultad de Medicina and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | MedicLatina |
| FullText | Links: – Type: pdflink Text: Availability: 0 |
|---|---|
| Header | DbId: lth DbLabel: MedicLatina An: 188788090 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Un análisis del efecto del volumen del cuerpo de la jeringa sobre la falla mecánica durante la inyección. – Name: TitleAlt Label: Alternate Title Group: TiAlt Data: An analysis of the effect of syringe barrel volume on mechanical failure during injection. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Krisdiyanto%22">Krisdiyanto</searchLink><relatesTo>1</relatesTo><i> krisdiyanto@umy.ac.id</i><br /><searchLink fieldCode="AR" term="%22Acknowledgment%22">Acknowledgment</searchLink><relatesTo>1</relatesTo> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Gaceta+Médica+Boliviana%22">Gaceta Médica Boliviana</searchLink>. ene-jun2025, Vol. 48 Issue 1, p77-82. 6p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22SYRINGES%22">SYRINGES</searchLink><br /><searchLink fieldCode="DE" term="%22MECHANICAL+failures%22">MECHANICAL failures</searchLink><br /><searchLink fieldCode="DE" term="%22STRUCTURAL+failures%22">STRUCTURAL failures</searchLink><br /><searchLink fieldCode="DE" term="%22SAFETY%22">SAFETY</searchLink><br /><searchLink fieldCode="DE" term="%22INJECTIONS%22">INJECTIONS</searchLink><br /><searchLink fieldCode="DE" term="%22MEDICAL+equipment%22">MEDICAL equipment</searchLink><br /><searchLink fieldCode="DE" term="%22YIELD+stress%22">YIELD stress</searchLink> – Name: Abstract Label: Abstract (English) Group: Ab Data: Background: syringes are critical medical devices whose mechanical integrity is essential to ensure safety during injections. The cylinder volume can influence their mechanical performance. Objective: this study investigates how cylinder volume affects the mechanical integrity of syringes by analyzing von Mises stress, buckling risk, and overall safety during injection. Methods: autodesk Inventor 2021 and Ansys 2019 R2 were used to analyze syringes with volumes of 1 ml, 3 ml, 5 ml, and 10 ml. Von Mises stress and buckling susceptibility were evaluated during simulated injections. Results: smaller syringes showed higher stress levels: 14.030 MPa for the 1 ml syringe, 10.532 MPa for the 3 ml, and 7.0150 MPa for the 5 ml and 10 ml. The 5 ml syringe was the most susceptible to buckling, with a critical pressure of -0.1757 MPa. Despite this, all designs are safe under normal operating conditions. Conclusions: cylinder volume affects mechanical performance, but with proper design, even smaller syringes can remain safe, enhancing the reliability of medical treatments. [ABSTRACT FROM AUTHOR] – Name: Abstract Label: Abstract (Spanish) Group: Ab Data: Antecedentes: las jeringas son dispositivos médicos críticos cuya integridad mecánica es esencial para garantizar su seguridad en las inyecciones. El volumen del cilindro puede influir en su rendimiento mecánico. Objetivo: este estudio investiga cómo el volumen del cilindro afecta la integridad mecánica de las jeringas, analizando el esfuerzo de von Mises, el riesgo de pandeo y la seguridad general durante la inyección. Métodos: se utilizaron Autodesk Inventor 2021 y Ansys 2019 R2 para analizar jeringas de 1 ml, 3 ml, 5 ml y 10 ml. Se evaluaron los esfuerzos de von Mises y la susceptibilidad al pandeo durante inyecciones simuladas. Resultados: las jeringas más pequeñas presentaron mayores esfuerzos: 14,030 MPa para la de 1 ml, 10,532 MPa para la de 3 ml y 7,0150 MPa para las de 5 ml y 10 ml. La jeringa de 5 ml fue la más susceptible al pandeo, con una presión crítica de -0,1757 MPa. A pesar de ello, todas son seguras bajo condiciones normales. Conclusiones: el volumen del cilindro afecta el rendimiento mecánico, pero con un buen diseño, incluso las jeringas pequeñas pueden ser seguras, mejorando la fiabilidad de los tratamientos médicos. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Gaceta Médica Boliviana is the property of Universidad Mayor de San Simon, Facultad de Medicina and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=lth&AN=188788090 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.47993/gmb.v48i1.959 Languages: – Code: spa Text: Spanish PhysicalDescription: Pagination: PageCount: 6 StartPage: 77 Subjects: – SubjectFull: SYRINGES Type: general – SubjectFull: MECHANICAL failures Type: general – SubjectFull: STRUCTURAL failures Type: general – SubjectFull: SAFETY Type: general – SubjectFull: INJECTIONS Type: general – SubjectFull: MEDICAL equipment Type: general – SubjectFull: YIELD stress Type: general Titles: – TitleFull: Un análisis del efecto del volumen del cuerpo de la jeringa sobre la falla mecánica durante la inyección. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Krisdiyanto – PersonEntity: Name: NameFull: Acknowledgment IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 01 Text: ene-jun2025 Type: published Y: 2025 Identifiers: – Type: issn-print Value: 10122966 Numbering: – Type: volume Value: 48 – Type: issue Value: 1 Titles: – TitleFull: Gaceta Médica Boliviana Type: main |
| ResultId | 1 |