MEDIDAS LINEARES EM RADIOGRAFIAS PANORÂMICAS DIGITALIZADAS, FORNECIDAS POR PROGRAMA DE IMAGEM PARA PLANEJAMENTO NA IMPLANTODONTIA: CORRELAÇÃO E ANÁLISE CRÍTICA.

Saved in:
Bibliographic Details
Title: MEDIDAS LINEARES EM RADIOGRAFIAS PANORÂMICAS DIGITALIZADAS, FORNECIDAS POR PROGRAMA DE IMAGEM PARA PLANEJAMENTO NA IMPLANTODONTIA: CORRELAÇÃO E ANÁLISE CRÍTICA.
Authors: Kiatake Fontão, Flávia Noemy Gasparini1, Montenegro Chinellato, Luiz Eduardo2, Rubira Bullen, Izabel Regina Fisher3, Gião Dezotti, Mariela Siqueira4
Source: Revista Salusvita. 2006, Vol. 25 Issue 2, p165-183. 19p.
Subjects: PANORAMIC radiography, IMAGING system software, EDENTULOUS mouth, CALIPERS, COMPUTER software
Abstract (English): The present study evaluated the linear measurements on panoramic radiographs obtained in image software for presurgical implant planning (Planimp) compared to the manual technique using a digital caliper. The sample comprised 50 panoramic radiographs of edentulous patients, on which reference points were traced for delineation of 14 anatomical dimensions on each radiograph. The radiographs were digitized with a HP Scanjet 4C/T scanner (Hewllett Packard). Linear measurements of the anatomical dimensions were achieved by the manual method (digital caliper) and digital method (Planimp software). The data achieved were submitted to statistical analysis of variance, at a significance level of 5%. This statistical analysis has not demonstrate any significant difference between the manual and digital measurements. The results allowed the conclusion that the Planimp software displayed proper performances for calibration of the radiographic image and for measurement of linear dimensions. [ABSTRACT FROM AUTHOR]
Abstract (Portuguese): O objetivo da presente pesquisa foi avaliar, comparativamente, as medidas lineares em radiografias panorâmicas obtidas no programa Planimp, desenvolvido para planejamento pré-cirúrgico em implantes, com as obtidas pelo método manual, utilizando um paquímetro digital. A amostra constou de 50 radiografias panorâmicas de pacientes edêntulos, nas quais foram marcados pontos de referência para delimitar 14 distâncias anatômicas em cada radiografia. As radiografias foram digitalizadas por meio de um scanner HP Scanjet 4C/T (Hewllett Packard). As mensurações lineares das distâncias anatômicas foram realizadas pelo método manual (paquímetro digital) e pelo método digital (Planimp). Os dados obtidos foram submetidos à análise estatística por meio do teste de variância, com nível de significância de 5%. Esta análise estatística mostrou que não houve diferença significante entre as medidas manuais e digitais. A partir dos resultados concluiu-se que o programa Planimp apresentou desempenho adequado para a calibração da imagem radiográfica, bem como para a mensuração de distâncias lineares. [ABSTRACT FROM AUTHOR]
Copyright of Revista Salusvita is the property of Salusvita Journal and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: MedicLatina
FullText Links:
  – Type: pdflink
Text:
  Availability: 0
Header DbId: lth
DbLabel: MedicLatina
An: 79921302
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: MEDIDAS LINEARES EM RADIOGRAFIAS PANORÂMICAS DIGITALIZADAS, FORNECIDAS POR PROGRAMA DE IMAGEM PARA PLANEJAMENTO NA IMPLANTODONTIA: CORRELAÇÃO E ANÁLISE CRÍTICA.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Kiatake+Fontão%2C+Flávia+Noemy+Gasparini%22">Kiatake Fontão, Flávia Noemy Gasparini</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Montenegro+Chinellato%2C+Luiz+Eduardo%22">Montenegro Chinellato, Luiz Eduardo</searchLink><relatesTo>2</relatesTo><br /><searchLink fieldCode="AR" term="%22Rubira+Bullen%2C+Izabel+Regina+Fisher%22">Rubira Bullen, Izabel Regina Fisher</searchLink><relatesTo>3</relatesTo><br /><searchLink fieldCode="AR" term="%22Gião+Dezotti%2C+Mariela+Siqueira%22">Gião Dezotti, Mariela Siqueira</searchLink><relatesTo>4</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Revista+Salusvita%22">Revista Salusvita</searchLink>. 2006, Vol. 25 Issue 2, p165-183. 19p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22PANORAMIC+radiography%22">PANORAMIC radiography</searchLink><br /><searchLink fieldCode="DE" term="%22IMAGING+system+software%22">IMAGING system software</searchLink><br /><searchLink fieldCode="DE" term="%22EDENTULOUS+mouth%22">EDENTULOUS mouth</searchLink><br /><searchLink fieldCode="DE" term="%22CALIPERS%22">CALIPERS</searchLink><br /><searchLink fieldCode="DE" term="%22COMPUTER+software%22">COMPUTER software</searchLink>
– Name: Abstract
  Label: Abstract (English)
  Group: Ab
  Data: The present study evaluated the linear measurements on panoramic radiographs obtained in image software for presurgical implant planning (Planimp) compared to the manual technique using a digital caliper. The sample comprised 50 panoramic radiographs of edentulous patients, on which reference points were traced for delineation of 14 anatomical dimensions on each radiograph. The radiographs were digitized with a HP Scanjet 4C/T scanner (Hewllett Packard). Linear measurements of the anatomical dimensions were achieved by the manual method (digital caliper) and digital method (Planimp software). The data achieved were submitted to statistical analysis of variance, at a significance level of 5%. This statistical analysis has not demonstrate any significant difference between the manual and digital measurements. The results allowed the conclusion that the Planimp software displayed proper performances for calibration of the radiographic image and for measurement of linear dimensions. [ABSTRACT FROM AUTHOR]
– Name: Abstract
  Label: Abstract (Portuguese)
  Group: Ab
  Data: O objetivo da presente pesquisa foi avaliar, comparativamente, as medidas lineares em radiografias panorâmicas obtidas no programa Planimp, desenvolvido para planejamento pré-cirúrgico em implantes, com as obtidas pelo método manual, utilizando um paquímetro digital. A amostra constou de 50 radiografias panorâmicas de pacientes edêntulos, nas quais foram marcados pontos de referência para delimitar 14 distâncias anatômicas em cada radiografia. As radiografias foram digitalizadas por meio de um scanner HP Scanjet 4C/T (Hewllett Packard). As mensurações lineares das distâncias anatômicas foram realizadas pelo método manual (paquímetro digital) e pelo método digital (Planimp). Os dados obtidos foram submetidos à análise estatística por meio do teste de variância, com nível de significância de 5%. Esta análise estatística mostrou que não houve diferença significante entre as medidas manuais e digitais. A partir dos resultados concluiu-se que o programa Planimp apresentou desempenho adequado para a calibração da imagem radiográfica, bem como para a mensuração de distâncias lineares. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Revista Salusvita is the property of Salusvita Journal and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=lth&AN=79921302
RecordInfo BibRecord:
  BibEntity:
    Languages:
      – Code: por
        Text: Portuguese
    PhysicalDescription:
      Pagination:
        PageCount: 19
        StartPage: 165
    Subjects:
      – SubjectFull: PANORAMIC radiography
        Type: general
      – SubjectFull: IMAGING system software
        Type: general
      – SubjectFull: EDENTULOUS mouth
        Type: general
      – SubjectFull: CALIPERS
        Type: general
      – SubjectFull: COMPUTER software
        Type: general
    Titles:
      – TitleFull: MEDIDAS LINEARES EM RADIOGRAFIAS PANORÂMICAS DIGITALIZADAS, FORNECIDAS POR PROGRAMA DE IMAGEM PARA PLANEJAMENTO NA IMPLANTODONTIA: CORRELAÇÃO E ANÁLISE CRÍTICA.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Kiatake Fontão, Flávia Noemy Gasparini
      – PersonEntity:
          Name:
            NameFull: Montenegro Chinellato, Luiz Eduardo
      – PersonEntity:
          Name:
            NameFull: Rubira Bullen, Izabel Regina Fisher
      – PersonEntity:
          Name:
            NameFull: Gião Dezotti, Mariela Siqueira
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 08
              Text: 2006
              Type: published
              Y: 2006
          Identifiers:
            – Type: issn-print
              Value: 01019910
          Numbering:
            – Type: volume
              Value: 25
            – Type: issue
              Value: 2
          Titles:
            – TitleFull: Revista Salusvita
              Type: main
ResultId 1