Fixation of the electrode array with bone paté in cochlear implant surgery.

Saved in:
Bibliographic Details
Title: Fixation of the electrode array with bone paté in cochlear implant surgery.
Authors: Gstoettner WK (AUTHOR), Hamzavi J (AUTHOR), Baumgartner WD (AUTHOR), Czerny CC (AUTHOR)
Source: Acta Oto-Laryngologica. 08/31/2000, Vol. 120 Issue 3, p369-374. 6p.
Abstract: The aims of this study are to report on our experiences using a bone paté (a mixture of bone dust and fibrin glue) fixation technique of the Combi 40/Combi 40+ electrode array during cochlear implant surgery. A retrospective study involving 57 patients implanted with Combi 40/Combi 40+ cochlear implants over 31 months fixated with bone paté was performed, utilizing our electrical fitting data as well as re-operation and X-ray findings. We found no signs of electrode retraction in any of the 57 patients. Electrical fitting data showed no reduction in the number of active channels between the first and last times patients were evaluated (a period of 4-31 months). Furthermore, a comparison of X-rays performed within the first week following cochlear implant surgery and again 7-27 months later showed no signs of electrode retraction in 10 patients. Three patients who required re-operations after implantation allowed us the opportunity to examine the effectiveness of the bone paté fixation and revealed new bone growth with the electrode fixed in place. Histological examinations of the newly built bone showed laminar and compact bone structures. The use of bone paté was thus found to be a suitable method of electrode fixation in cochlear implant surgery. Compared with other fixation methods, it is inexpensive and uses no foreign materials. [ABSTRACT FROM AUTHOR]
Copyright of Acta Oto-Laryngologica is the property of Taylor & Francis Ltd and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Psychology and Behavioral Sciences Collection
Description
Abstract:The aims of this study are to report on our experiences using a bone paté (a mixture of bone dust and fibrin glue) fixation technique of the Combi 40/Combi 40+ electrode array during cochlear implant surgery. A retrospective study involving 57 patients implanted with Combi 40/Combi 40+ cochlear implants over 31 months fixated with bone paté was performed, utilizing our electrical fitting data as well as re-operation and X-ray findings. We found no signs of electrode retraction in any of the 57 patients. Electrical fitting data showed no reduction in the number of active channels between the first and last times patients were evaluated (a period of 4-31 months). Furthermore, a comparison of X-rays performed within the first week following cochlear implant surgery and again 7-27 months later showed no signs of electrode retraction in 10 patients. Three patients who required re-operations after implantation allowed us the opportunity to examine the effectiveness of the bone paté fixation and revealed new bone growth with the electrode fixed in place. Histological examinations of the newly built bone showed laminar and compact bone structures. The use of bone paté was thus found to be a suitable method of electrode fixation in cochlear implant surgery. Compared with other fixation methods, it is inexpensive and uses no foreign materials. [ABSTRACT FROM AUTHOR]
ISSN:00016489
DOI:10.1080/000164800750000586