Electrophysiological findings in two bilateral cochlear implant cases: does the duration of deafness affect electrically evoked auditory brain stem responses?
Saved in:
| Title: | Electrophysiological findings in two bilateral cochlear implant cases: does the duration of deafness affect electrically evoked auditory brain stem responses? |
|---|---|
| Authors: | Thai-Van H (AUTHOR), Gallego S (AUTHOR), Truy E (AUTHOR), Veuillet E (AUTHOR), Collet L (AUTHOR) |
| Source: | Annals of Otology, Rhinology & Laryngology. Nov2002, Vol. 111 Issue 11, p1008-1014. 7p. |
| Abstract: | Bilateral cochlear implantation provides an interesting model for in vivo study of the effect of long-term profound deafness on neural transmission. We present electrophysiological observations on 2 patients implanted with the MXM Binaural Digisonic Convex system. This uncommon design consists of 2 electrode arrays placed bilaterally into the scala tympani and controlled by a single speech processor. In both patients, the duration of deafness before cochlear implantation clearly differed from one ear to the other. Electrically evoked auditory brain stem responses (EABRs) were measured and the EABRs from the ear with the longer deafness duration showed a lengthening of wave V latency. In I patient, recordings from this ear also showed a lack of reproducibility of wave III. The data suggest that neural responsiveness in the peripheral and intermediate auditory pathways is adversely affected by deafness duration. Poor EABRs on one ear possibly result from the total duration of deafness in this ear and/or compensation by the other ear. [ABSTRACT FROM AUTHOR] |
| Copyright of Annals of Otology, Rhinology & Laryngology is the property of Sage Publications Inc. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Psychology and Behavioral Sciences Collection |
| FullText | Links: – Type: pdflink Text: Availability: 0 |
|---|---|
| Header | DbId: pbh DbLabel: Psychology and Behavioral Sciences Collection An: 106842308 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: Electrophysiological findings in two bilateral cochlear implant cases: does the duration of deafness affect electrically evoked auditory brain stem responses? – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Thai-Van+H%22">Thai-Van H</searchLink> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Gallego+S%22">Gallego S</searchLink> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Truy+E%22">Truy E</searchLink> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Veuillet+E%22">Veuillet E</searchLink> (AUTHOR)<br /><searchLink fieldCode="AR" term="%22Collet+L%22">Collet L</searchLink> (AUTHOR) – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Annals+of+Otology%2C+Rhinology+%26+Laryngology%22">Annals of Otology, Rhinology & Laryngology</searchLink>. Nov2002, Vol. 111 Issue 11, p1008-1014. 7p. – Name: Abstract Label: Abstract Group: Ab Data: Bilateral cochlear implantation provides an interesting model for in vivo study of the effect of long-term profound deafness on neural transmission. We present electrophysiological observations on 2 patients implanted with the MXM Binaural Digisonic Convex system. This uncommon design consists of 2 electrode arrays placed bilaterally into the scala tympani and controlled by a single speech processor. In both patients, the duration of deafness before cochlear implantation clearly differed from one ear to the other. Electrically evoked auditory brain stem responses (EABRs) were measured and the EABRs from the ear with the longer deafness duration showed a lengthening of wave V latency. In I patient, recordings from this ear also showed a lack of reproducibility of wave III. The data suggest that neural responsiveness in the peripheral and intermediate auditory pathways is adversely affected by deafness duration. Poor EABRs on one ear possibly result from the total duration of deafness in this ear and/or compensation by the other ear. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Annals of Otology, Rhinology & Laryngology is the property of Sage Publications Inc. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=pbh&AN=106842308 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.1177/000348940211101111 Languages: – Code: eng Text: English PhysicalDescription: Pagination: PageCount: 7 StartPage: 1008 Titles: – TitleFull: Electrophysiological findings in two bilateral cochlear implant cases: does the duration of deafness affect electrically evoked auditory brain stem responses? Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Thai-Van H – PersonEntity: Name: NameFull: Gallego S – PersonEntity: Name: NameFull: Truy E – PersonEntity: Name: NameFull: Veuillet E – PersonEntity: Name: NameFull: Collet L IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 11 Text: Nov2002 Type: published Y: 2002 Identifiers: – Type: issn-print Value: 00034894 Numbering: – Type: volume Value: 111 – Type: issue Value: 11 Titles: – TitleFull: Annals of Otology, Rhinology & Laryngology Type: main |
| ResultId | 1 |