All-Systolic Non-ECG-gated Myocardial Perfusion MRI: Feasibility of Multi-Slice Continuous First-Pass Imaging.

Saved in:
Bibliographic Details
Title: All-Systolic Non-ECG-gated Myocardial Perfusion MRI: Feasibility of Multi-Slice Continuous First-Pass Imaging.
Authors: Sharif, Behzad1,2, Arsanjani, Reza1,3,4, Dharmakumar, Rohan1,2,4, Bairey Merz, C. Noel2,3,4,5, Berman, Daniel S.1,3,4, Li, Debiao1,2,4
Source: Magnetic Resonance in Medicine. Dec2015, Vol. 74 Issue 6, p1661-1674. 14p.
Abstract: Purpose: To develop and test the feasibility of a new method for non-ECG-gated first-pass perfusion (FPP) cardiac MR capable of imaging multiple short-axis slices at the same systolic cardiac phase. Methods: A magnetization-driven pulse sequence was developed for non-ECG-gated FPP imaging without saturationrecovery preparation using continuous slice-interleaved radial sampling. The image reconstruction method, dubbed TRACE, used self-gating based on reconstruction of a real-time imagebased navigator combined with reference-constrained compressed sensing. Data from ischemic animal studies (n=5) was used in a simulation framework to evaluate temporal fidelity. Healthy subjects (n=5) were studied using both the proposed approach and the conventional method to compare the myocardial contrast-to-noise ratio (CNR). Patients (n=2) underwent adenosine stress studies using the proposed method. Results: Temporal fidelity of the developed method was shown to be sufficient at high heart-rates. The healthy volunteers studies demonstrated normal perfusion and no dark-rim artifacts. Compared with the conventional scheme, myocardial CNR for the proposed method was slightly higher (8.660.6 versus 8.060.7). Patient studies showed stress-induced perfusion defects consistent with invasive angiography. Conclusion: The presented methods and results demonstrate feasibility of the proposed approach for high-resolution non-ECG-gated FPP imaging of 3 myocardial slices at the same systolic phase, and indicate its potential for achieving desirable image quality (high CNR and no dark-rim artifacts). [ABSTRACT FROM AUTHOR]
Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
Description
Abstract:Purpose: To develop and test the feasibility of a new method for non-ECG-gated first-pass perfusion (FPP) cardiac MR capable of imaging multiple short-axis slices at the same systolic cardiac phase. Methods: A magnetization-driven pulse sequence was developed for non-ECG-gated FPP imaging without saturationrecovery preparation using continuous slice-interleaved radial sampling. The image reconstruction method, dubbed TRACE, used self-gating based on reconstruction of a real-time imagebased navigator combined with reference-constrained compressed sensing. Data from ischemic animal studies (n=5) was used in a simulation framework to evaluate temporal fidelity. Healthy subjects (n=5) were studied using both the proposed approach and the conventional method to compare the myocardial contrast-to-noise ratio (CNR). Patients (n=2) underwent adenosine stress studies using the proposed method. Results: Temporal fidelity of the developed method was shown to be sufficient at high heart-rates. The healthy volunteers studies demonstrated normal perfusion and no dark-rim artifacts. Compared with the conventional scheme, myocardial CNR for the proposed method was slightly higher (8.660.6 versus 8.060.7). Patient studies showed stress-induced perfusion defects consistent with invasive angiography. Conclusion: The presented methods and results demonstrate feasibility of the proposed approach for high-resolution non-ECG-gated FPP imaging of 3 myocardial slices at the same systolic phase, and indicate its potential for achieving desirable image quality (high CNR and no dark-rim artifacts). [ABSTRACT FROM AUTHOR]
ISSN:07403194
DOI:10.1002/mrm.25752