Semi-empirical model for fluorescence lines evaluation in diagnostic x-ray beams. |
| Authors: |
Bontempi, Marco1,2 m.bontempi@biomec.ior.it, Andreani, Lucia3,4, Labanti, Claudio4,5, Costa, Paulo Roberto6, Rossi, Pier Luca3, Baldazzi, Giuseppe3,4 |
| Source: |
Applied Radiation & Isotopes. Jan2016, Vol. 107, p152-159. 8p. |
| Subjects: |
Tungsten, Diagnostic imaging, Fluorescence, X-rays, Bremsstrahlung, Anodes, Electron impact ionization |
| Abstract: |
Diagnostic x-ray beams are composed of bremsstrahlung and discrete fluorescence lines. The aim of this study is the development of an efficient model for the evaluation of the fluorescence lines. The most important electron ionization models are analyzed and implemented. The model results were compared with experimental data and with other independent spectra presented in the literature. The implemented peak models allow the discrimination between direct and indirect radiation emitted from tungsten anodes. The comparison with the independent literature spectra indicated a good agreement. [ABSTRACT FROM AUTHOR] |
|
Copyright of Applied Radiation & Isotopes is the property of Pergamon Press - An Imprint of Elsevier Science and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) |
| Database: |
Engineering Source |