Accuracy, repeatability, and interplatform reproducibility of T1 quantification methods used for DCE‐MRI: Results from a multicenter phantom study.
Saved in:
| Title: | Accuracy, repeatability, and interplatform reproducibility of T |
|---|---|
| Authors: | Bane, Octavia1,2, Hectors, Stefanie J.1,2, Wagner, Mathilde1,2, Arlinghaus, Lori L.3, Aryal, Madhava P.4, Cao, Yue4, Chenevert, Thomas L.5, Fennessy, Fiona6, Huang, Wei7, Hylton, Nola M.8, Kalpathy‐Cramer, Jayashree9, Keenan, Kathryn E.10, Malyarenko, Dariya I.5, Mulkern, Robert V.6, Newitt, David C.8, Russek, Stephen E.10, Stupic, Karl F.10, Tudorica, Alina11, Wilmes, Lisa J.8, Yankeelov, Thomas E.12 |
| Source: | Magnetic Resonance in Medicine. May2018, Vol. 79 Issue 5, p2564-2575. 12p. |
| Abstract: | Purpose: To determine the in vitro accuracy, test‐retest repeatability, and interplatform reproducibility of T1 quantification protocols used for dynamic contrast‐enhanced MRI at 1.5 and 3 T. Methods: A T1 phantom with 14 samples was imaged at eight centers with a common inversion‐recovery spin‐echo (IR‐SE) protocol and a variable flip angle (VFA) protocol using seven flip angles, as well as site‐specific protocols (VFA with different flip angles, variable repetition time, proton density, and Look‐Locker inversion recovery). Factors influencing the accuracy (deviation from reference NMR T1 measurements) and repeatability were assessed using general linear mixed models. Interplatform reproducibility was assessed using coefficients of variation. Results: For the common IR‐SE protocol, accuracy (median error across platforms = 1.4–5.5%) was influenced predominantly by T1 sample ( |
| Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
|
Full text is not displayed to guests.
Login for full access.
|
|
Be the first to leave a comment!