Assessment of coronary artery stents using 16-slice MDCT angiography: evaluation of a dedicated reconstruction kernel and a noise reduction filter.
Saved in:
| Title: | Assessment of coronary artery stents using 16-slice MDCT angiography: evaluation of a dedicated reconstruction kernel and a noise reduction filter. |
|---|---|
| Authors: | Seifarth, Harald1 seifarth@uni-muenster.de, Raupach, Rainer2, Schaller, Stefan2, Fallenberg, Eva1, Flohr, Thomas2, Heindel, Walter1, Fischbach, Roman1, Maintz, David1, Fallenberg, Eva Maria (AUTHOR) |
| Source: | European Radiology. Apr2005, Vol. 15 Issue 4, p721-726. 6p. |
| Subjects: | Tomography, Surgical stents, Medical imaging systems, Angiography, Noise, Noise control, Coronary artery surgery, Comparative studies, Computed tomography, Research methodology, Medical cooperation, Research, Pilot projects, Evaluation research, Medical artifacts, Coronary angiography |
| Abstract: | To compare the effect of different reconstruction kernels and a noise-reducing postprocessing filter on the delineation of coronary artery stents in 16-slice CT-angiography. Ten patients with coronary stents (seven LAD, five RCX and three RCA) were examined with a 16-slice MDCT using standard acquisition parameters. Images were reconstructed using a medium soft (B30f) and a dedicated, edge-enhancing kernel (B46f). Additional postprocessing with an edge-preserving filter was performed on B46f images to reduce the image noise. In multiplanar reformations (MPRs) along and perpendicular to the stent axis, intraluminal attenuation values and the visible lumen diameter were measured. Image noise was measured in the subcutaneous fat using a region of interest (ROI) technique. Arterial enhancement in the aorta was 275.1 HU. Attenuation in the stent lumen was 390.4, 340.0 and 346.8 HU in MPRs derived from B30f, original B46 and postprocessed B46f images. The mean noise level was 20.4, 35.0 and 24.9 HU respectively. The visible lumen diameter was significantly greater in B46f and postprocessed B46f images (2.17 and 2.16 mm), compared to 1.93 mm in B30f images (p<0.01). Edge-enhancing reconstruction kernels increase the visible stent lumen, but also increase image noise. Dedicated postprocessing filters can reduce the introduced noise without a loss of spatial resolution. [ABSTRACT FROM AUTHOR] |
| Copyright of European Radiology is the property of Springer Nature and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |