Development of a standalone delivery sequence model for proton arc therapy.
Saved in:
| Title: | Development of a standalone delivery sequence model for proton arc therapy. |
|---|---|
| Authors: | Liu, Gang1,2,3 (AUTHOR) liugang9910@163.com, Zhao, Lewei4 (AUTHOR), Liu, Peilin5 (AUTHOR), Yan, Di5 (AUTHOR), Deraniyagala, Rohan5 (AUTHOR), Stevens, Craig5 (AUTHOR), Li, Xiaoqiang5 (AUTHOR), Ding, Xuanfeng5 (AUTHOR) xuanfengding@gmail.com |
| Source: | Medical Physics. Apr2024, Vol. 51 Issue 4, p3067-3075. 9p. |
| Subjects: | Proton therapy, Medical research personnel, Electric arc, Proton beams, Inclinometer, Test methods |
| Abstract: | Background: Spot‐scanning proton arc (SPArc) has been drawing significant interests in recent years because of its capability of continuous proton irradiation during the gantry rotation. Previous studies demonstrated SPArc plans were delivered on a prototype of the DynamicARC solution, IBA ProteusONE. Purpose: We built a novel delivery sequence model through an independent experimental approach: the first SPArc delivery sequence model (DSMSPArc). Based on the model, we investigated SPArc treatment efficiency improvement in the routine proton clinical operation. Methods: SPArc test plans were generated and delivered on a prototype of the DynamicARC solution, IBA ProteusONE. An independent gantry inclinometer and the machine logfiles were used to derive the DSMSPArc. Seventeen SPArc plans were used to validate the model's accuracy independently. Two random clinical operation dates (6th January and 22nd March, 2021) from a single‐room proton therapy center (PTC) were selected to quantitatively assess the improvement of treatment efficiency compared to the IMPT. Results: The difference between the logfile and DSMSPArc is about 3.2 ± 4.8%. SPArc reduced 58.1% of the average treatment delivery time per patient compared to IMPT (p < 0.01). Daily treatment throughput could be increased by 30% using SPArc using a single‐room proton therapy system. Conclusions: The first model of dynamic arc therapy is established in this study through an independent experimental approach using logfiles and measurements which allows clinical users and investigators to simulate the dynamic treatment delivery and assess the daily treatment throughput improvement. [ABSTRACT FROM AUTHOR] |
| Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
|
Full text is not displayed to guests.
Login for full access.
|
|
| Abstract: | Background: Spot‐scanning proton arc (SPArc) has been drawing significant interests in recent years because of its capability of continuous proton irradiation during the gantry rotation. Previous studies demonstrated SPArc plans were delivered on a prototype of the DynamicARC solution, IBA ProteusONE. Purpose: We built a novel delivery sequence model through an independent experimental approach: the first SPArc delivery sequence model (DSMSPArc). Based on the model, we investigated SPArc treatment efficiency improvement in the routine proton clinical operation. Methods: SPArc test plans were generated and delivered on a prototype of the DynamicARC solution, IBA ProteusONE. An independent gantry inclinometer and the machine logfiles were used to derive the DSMSPArc. Seventeen SPArc plans were used to validate the model's accuracy independently. Two random clinical operation dates (6th January and 22nd March, 2021) from a single‐room proton therapy center (PTC) were selected to quantitatively assess the improvement of treatment efficiency compared to the IMPT. Results: The difference between the logfile and DSMSPArc is about 3.2 ± 4.8%. SPArc reduced 58.1% of the average treatment delivery time per patient compared to IMPT (p < 0.01). Daily treatment throughput could be increased by 30% using SPArc using a single‐room proton therapy system. Conclusions: The first model of dynamic arc therapy is established in this study through an independent experimental approach using logfiles and measurements which allows clinical users and investigators to simulate the dynamic treatment delivery and assess the daily treatment throughput improvement. [ABSTRACT FROM AUTHOR] |
|---|---|
| ISSN: | 00942405 |
| DOI: | 10.1002/mp.16879 |