A single‐camera telecentric imaging stereo‐digital image correlation system and its application to carotid pulse signal detection.

Saved in:
Bibliographic Details
Title: A single‐camera telecentric imaging stereo‐digital image correlation system and its application to carotid pulse signal detection.
Authors: Chen, Chao1,2 (AUTHOR), Tang, Xiaojun3 (AUTHOR) xiaojuntang87@sina.com, Pan, Bing2 (AUTHOR) panb@buaa.edu.cn
Source: Strain. Jun2024, Vol. 60 Issue 3, p1-10. 10p.
Subjects: Signal detection, Stereo vision (Computer science), Carotid intima-media thickness, Imaging systems, Digital images, Digital cameras, Carotid artery, Acceleration (Mechanics), Digital image correlation
Abstract: Carotid pulse signal (CPS) is one of the most important biomarkers for evaluating cardiovascular risk. To realize easy‐to‐operate and accurate CPS detection, a single‐camera telecentric imaging stereo‐digital image correlation (stereo‐DIC) system is established. The system uses a specially designed colour separation device and a single colour camera attached with a telecentric lens to record images of blue and red colours from different optical paths. By processing the recorded two sub‐channel colour images using regular stereo‐DIC algorithm, full‐field 3D displacements of the skin on the neck directly over the carotid artery are obtained. The maximum fluctuations in the out‐of‐plane displacement direction are extracted as CPS, based on which velocity, acceleration and frequency can be further extracted. The performance of the established system was demonstrated by detecting the pulse signal of a volunteer before and after exercise. [ABSTRACT FROM AUTHOR]
Copyright of Strain is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Full text is not displayed to guests.
Description
Abstract:Carotid pulse signal (CPS) is one of the most important biomarkers for evaluating cardiovascular risk. To realize easy‐to‐operate and accurate CPS detection, a single‐camera telecentric imaging stereo‐digital image correlation (stereo‐DIC) system is established. The system uses a specially designed colour separation device and a single colour camera attached with a telecentric lens to record images of blue and red colours from different optical paths. By processing the recorded two sub‐channel colour images using regular stereo‐DIC algorithm, full‐field 3D displacements of the skin on the neck directly over the carotid artery are obtained. The maximum fluctuations in the out‐of‐plane displacement direction are extracted as CPS, based on which velocity, acceleration and frequency can be further extracted. The performance of the established system was demonstrated by detecting the pulse signal of a volunteer before and after exercise. [ABSTRACT FROM AUTHOR]
ISSN:00392103
DOI:10.1111/str.12466