Ultraportable Quantitative Ultrasound for Hepatic Steatosis Assessment.

Saved in:
Bibliographic Details
Title: Ultraportable Quantitative Ultrasound for Hepatic Steatosis Assessment.
Authors: Hériard-Dubreuil, Baptiste1,2 (AUTHOR) baptiste.heriard-dubreuil@e-scopics.com, Besson, Adrien1 (AUTHOR), Mamou, Jonathan3 (AUTHOR), Gay, Joël1 (AUTHOR), Foucher, Juliette4 (AUTHOR), De Ledinghen, Victor4 (AUTHOR), Cohen-Bacrie, Claude1 (AUTHOR)
Source: Ultrasound in Medicine & Biology. Dec2024, Vol. 50 Issue 12, p1842-1848. 7p.
Subjects: Fatty liver, Pearson correlation (Statistics), Intraclass correlation, Ultrasonic imaging, Scanning systems
Abstract: This study aimed to evaluate the performances of quantitative ultrasound (QUS) for the detection and assessment of hepatic steatosis when implemented using an ultraportable ultrasound scanner. Seven established QUS parameters were investigated. Ultrasound signals were acquired using a new ultraportable ultrasound device, the Hepatoscope. The feasibility of QUS using the Hepatoscope was first assessed in vitro. Clinical reliability, accuracy and staging capabilities were evaluated in 60 patients referred to a hepatology consultation for known chronic liver disease and enrolled in a prospective clinical investigation using the controlled attenuation parameter (CAP) as ground truth. QUS parameters showed moderate (intra-class correlation coefficient [ICC] >0.50) to excellent (ICC >0.90) reliability. Two parameters, namely Lizzi-Feleppa mid-band fit and attenuation, were both reliable (ICC = 0.89 and 0.86, respectively) and correlated with the CAP (squared Pearson correlation coefficient of R2 = 0.65 and R2 = 0.6, respectively). For steatosis detection (S0 vs. ≥S1), the two parameters yielded an area under the receiving operating characteristic curve of 0.90 and 0.86, respectively (95% confidence interval: [0.81–0.99] and [0.76–0.96], respectively). QUS can be reliably and accurately implemented on ultraportable ultrasound scanners. The combination of ultraportability and quantitative assessment of liver fat is promising for large-scale screening and monitoring of hepatic steatosis. [ABSTRACT FROM AUTHOR]
Copyright of Ultrasound in Medicine & Biology is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Description
Abstract:This study aimed to evaluate the performances of quantitative ultrasound (QUS) for the detection and assessment of hepatic steatosis when implemented using an ultraportable ultrasound scanner. Seven established QUS parameters were investigated. Ultrasound signals were acquired using a new ultraportable ultrasound device, the Hepatoscope. The feasibility of QUS using the Hepatoscope was first assessed in vitro. Clinical reliability, accuracy and staging capabilities were evaluated in 60 patients referred to a hepatology consultation for known chronic liver disease and enrolled in a prospective clinical investigation using the controlled attenuation parameter (CAP) as ground truth. QUS parameters showed moderate (intra-class correlation coefficient [ICC] >0.50) to excellent (ICC >0.90) reliability. Two parameters, namely Lizzi-Feleppa mid-band fit and attenuation, were both reliable (ICC = 0.89 and 0.86, respectively) and correlated with the CAP (squared Pearson correlation coefficient of R2 = 0.65 and R2 = 0.6, respectively). For steatosis detection (S0 vs. ≥S1), the two parameters yielded an area under the receiving operating characteristic curve of 0.90 and 0.86, respectively (95% confidence interval: [0.81–0.99] and [0.76–0.96], respectively). QUS can be reliably and accurately implemented on ultraportable ultrasound scanners. The combination of ultraportability and quantitative assessment of liver fat is promising for large-scale screening and monitoring of hepatic steatosis. [ABSTRACT FROM AUTHOR]