A single-photon CCD-based setup for in situ measurement of the X-ray spectrum of mammographic units

Saved in:
Bibliographic Details
Title: A single-photon CCD-based setup for in situ measurement of the X-ray spectrum of mammographic units
Authors: Labate, L.1 luca.labate@ipcf.cnr.it, Levato, T.2, Galimberti, M.1, Giulietti, A.1, Giulietti, D.2, Sanna, M.1, Traino, C.3, Lazzeri, M.3, Gizzi, L.A.1
Source: Nuclear Instruments & Methods in Physics Research Section A. Sep2008, Vol. 594 Issue 2, p278-282. 5p.
Subjects: X-ray tubes, Mammograms, Photons, CCD cameras
Abstract: Abstract: A technique enabling in situ measurements of the spectrum of X-ray tubes employed in mammographic screenings is described. The technique involves the use of a commercially available CCD camera and a set of metal foils and is particularly useful to perform a fast evaluation of the spectral properties of a Mo anode mammographic system operating at standard flux levels. A description of the detector calibration procedure is first given, followed by a discussion of the study and choice of an appropriate set of X-ray attenuation foils. Finally, the use of the system for a spectroscopic characterization of a mammographic system is reported. [Copyright &y& Elsevier]
Copyright of Nuclear Instruments & Methods in Physics Research Section A is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Description
Abstract:Abstract: A technique enabling in situ measurements of the spectrum of X-ray tubes employed in mammographic screenings is described. The technique involves the use of a commercially available CCD camera and a set of metal foils and is particularly useful to perform a fast evaluation of the spectral properties of a Mo anode mammographic system operating at standard flux levels. A description of the detector calibration procedure is first given, followed by a discussion of the study and choice of an appropriate set of X-ray attenuation foils. Finally, the use of the system for a spectroscopic characterization of a mammographic system is reported. [Copyright &y& Elsevier]
ISSN:01689002
DOI:10.1016/j.nima.2008.06.035