Reliability of two calculation codes for efficiency calibrations of HPGe detectors

Saved in:
Bibliographic Details
Title: Reliability of two calculation codes for efficiency calibrations of HPGe detectors
Authors: Abbas, K. kamel.abbas@jrc.it, Simonelli, F.1, D’Alberti, F.1, Forte, M.1, Stroosnijder, M.F.1
Source: Applied Radiation & Isotopes. May2002, Vol. 56 Issue 5, p703. 7p.
Subjects: Calibration, Gamma rays
Abstract: This paper reports the reliability of efficiency calibrations for γ-ray detectors using the calculation codes ANGLE and LabSOCS. For experimental verification, three HPGe detectors under various laboratory geometry configurations were used for this study. An overall comparison between experimental and calculated efficiency calibration curves is presented and comments on the various error sources affecting the final results are given. The deviations are generally below 10%, which could be acceptable for many applications. [Copyright &y& Elsevier]
Copyright of Applied Radiation & Isotopes is the property of Pergamon Press - An Imprint of Elsevier Science and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
Description
Abstract:This paper reports the reliability of efficiency calibrations for <f>γ</f>-ray detectors using the calculation codes ANGLE and LabSOCS. For experimental verification, three HPGe detectors under various laboratory geometry configurations were used for this study. An overall comparison between experimental and calculated efficiency calibration curves is presented and comments on the various error sources affecting the final results are given. The deviations are generally below 10%, which could be acceptable for many applications. [Copyright &y& Elsevier]
ISSN:09698043