Bibliographic Details
| Title: |
The LNE-LNHB water calorimeter for primary measurement of absorbed dose at low depth in water: application to medium-energy x-rays. |
| Authors: |
Rapp, B.1 benjamin.rapp@cea.fr, Perichon, N.1, Denoziere, M.1, Daures, J.1, Ostrowsky, A.1, Bordy, J.-M.1 |
| Source: |
Physics in Medicine & Biology. 2013, Vol. 58 Issue 9, p2769-2786. 18p. |
| Subjects: |
Calorimeters, Metrology, Radiation dosimetry, Ionizing radiation, X-rays |
| Abstract: |
Water calorimeters are used to establish absorbed dose standards in several national metrology laboratories involved in ionizing radiation dosimetry. These calorimeters have been first used in high-energy photons of 60Co or accelerator beams, where the depth of measurement in water is large (5 or 10 cm). The LNE-LNHB laboratory has developed a specific calorimeter which makes measurements at low depth in water (down to 0.5 cm) easier, in order to fulfil the reference conditions required by the international dosimetry protocols for medium-energy x-rays. This new calorimeter was first used to measure the absorbed dose rate in water at a depth of 2 cm for six medium-energy x-ray reference beams with a tube potential from 80 to 300 kV. The relative combined standard uncertainty obtained on the absorbed dose rate to water is lower than 0.8%. An overview of the design of the calorimeter is given, followed by a detailed description of the calculation of the correction factors and the calorimetric measurements. [ABSTRACT FROM AUTHOR] |
|