Conditions to obtain precise and true measurements of the intramolecular 13C distribution in organic molecules by isotopic 13C nuclear magnetic resonance spectrometry.
Saved in:
| Title: | Conditions to obtain precise and true measurements of the intramolecular 13C distribution in organic molecules by isotopic 13C nuclear magnetic resonance spectrometry. |
|---|---|
| Authors: | Bayle, Kevin1, Gilbert, Alexis2,3, Julien, Maxime1, Yamada, Keita2, Silvestre, Virginie1, Robins, Richard J.1, Akoka, Serge1, Yoshida, Naohiro2,3, Remaud, Gérald S.1 gerald.remaud@univ-nantes.fr |
| Source: | Analytica Chimica Acta. Oct2014, Vol. 846, p1-7. 7p. |
| Subjects: | Carbon isotopes, Intramolecular isotope effects, NMR spectrometers, Organic compounds, Biosynthesis, Chemical reactions, Chemical shift (Nuclear magnetic resonance) |
| Abstract: | Intramolecular 13 C composition gives access to new information on the (bio) synthetic history of a given molecule. Isotopic 13 C NMR spectrometry provides a general tool for measuring the position-specific 13 C content. As an emerging technique, some aspects of its performance are not yet fully delineated. This paper reports on (i) the conditions required to obtain satisfactory trueness and precision for the determination of the internal 13 C distribution, and (ii) an approach to determining the “absolute” position-specific 13 C content. In relation to (i), a precision of <1% can be obtained whatever the molecule on any spectrometer, once quantitative conditions are met, in particular appropriate proton decoupling efficiency. This performance is a prerequisite to the measurement of isotope fractionation either on the transformed or residual compound when a chemical reaction or process is being studied. The study of the trueness has revealed that the response of the spectrometer depends on the 13 C frequency range of the studied molecule, i.e. the chemical shift range. The “absolute value” and, therefore, the trueness of the 13 C NMR measurements has been assessed on acetic acid and by comparison to the results obtained on the fragments from COOH and CH 3 by isotopic mass spectrometry coupled to a pyrolysis device (GC-Py–irm-MS), this technique being the reference method for acetic acid. Of the two NMR spectrometers used in this work, one gave values that corresponded to those obtained by GC-Py–irm-MS (thus, the “true” value) while the other showed a bias, which was dependent to the range covered by the resonance frequencies of the molecule. Therefore, the former can be used directly for studying isotope affiliations, while the latter can only be used directly for comparative data, for example in authenticity studies, but can also be used to obtain the true values by applying appropriate correction factors. The present study assesses several key protocol steps required to enable the determination of position-specific 13 C content by isotopic 13 C NMR, irrespective of the NMR spectrometer: parameters to be adjusted, performance test using [1,2- 13 C 2 ]acetic acid, generation of correction factors. [ABSTRACT FROM AUTHOR] |
| Copyright of Analytica Chimica Acta is the property of Elsevier B.V. and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |